| Literature DB >> 32237203 |
Sadia Aftab1,2,3, Michael P Barnes1,4,5, Marcus Doebrich1,5, Joerg Lehmann1,2,5.
Abstract
Accurate quantification of absorbed radiation dose is important for safe and effective delivery of radiation therapy. An important aspect to this is reference dosimetry, which is performed under reference conditions specified by international codes of practice. Such measurements are usually performed in a water phantom. In the Sun Nuclear Corporation (SNC) three-dimensional (3D) scanner water tank system the detector holder is fixed to a horizontal metallic drive rail (MDR) which is in close proximity to the active volume of the detector. In this project, the dosimetric effects of the MDR on reference dosimetry were investigated for MV photons, MeV electrons, and kV photons by comparing reference dosimetry measurements in the SNC 3D scanner against similar measurements in a Standard Imaging (SI) one-dimensional (1D) tank and against measurements in the SNC 3D scanner with an additional, custom-made spacer placed beneath the chamber holder to increase the chamber - MDR separation. A second experiment investigated the difference in chamber reading dependent on chamber to MDR separation by fixing the chamber in the tank independently of the MDR and successively moving the MDR vertically to alter the separation. The results showed that measurements in the SNC 3D scanner agree with both SI 1D tank and SNC 3D scanner with spacer to within ±0.3% for MV photons, ±0.1% for electrons and ±1.2% for kV photons within the calculated setup uncertainty. The second experiment showed that the contribution of backscatter from the MDR was significant if the distance between MDR and chamber was reduced below the chamber's designed position in the SNC 3D scanner. The exception was for kV photons where the contribution of backscatter from the MDR was measured to be 0.5% at the designed distance. Further investigation would be useful for kV photons, where the experiment showed relatively large measurement uncertainties.Entities:
Keywords: 3D scanner; backscatter; reference dosimetry; water tank
Mesh:
Substances:
Year: 2020 PMID: 32237203 PMCID: PMC7170281 DOI: 10.1002/acm2.12858
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
FIG. 1Farmer‐type and ROOS chambers installed in the Sun Nuclear Corporation (SNC) three‐dimensional scanner using the SNC provided chamber holders, shows the close proximity of chamber to metallic drive rail.
FIG. 2Three‐dimensional printed 14.7 cm long spacer, used in experiment 1, to increase the chamber — metallic drive rail separation.
FIG. 3Experiment 2 setup arrangement. Ion chamber placed at reference depth in the Sun Nuclear Corporation three‐dimensional (3D) scanner using a retort stand and a 3D printed holder. The metallic drive rail (MDR) was moved independently of the camber to vary the chamber — MDR separation.
Setup uncertainties for all energies. Variation in source to surface distance (SSD) modelled using inverse square law and variation in depth modeled using percentage depth dose.
| 6 MV | 18 MV | 6 MeV | 12 MeV | 20 MeV | 100 kV | 200 kV | |
|---|---|---|---|---|---|---|---|
| SNC 3D scanner | |||||||
| SSDInvSq (%) | 0.20 | 0.20 | 0.20 | 0.20 | 0.20 | 0.50 | 0.50 |
| DepthPDD (%) | 0.46 | 0.13 | 0.10 | 0.26 | 0.35 | 2.10 | 0.60 |
| Total Uncer. (%) | 0.50 | 0.24 | 0.22 | 0.29 | 0.40 | 2.13 | 0.74 |
| SNC 3D scanner with spacer | |||||||
| SSDInvSq (%) | 0.20 | 0.20 | 0.20 | 0.20 | 0.20 | 0.50 | 0.50 |
| DepthPDD (%) | 0.46 | 0.13 | 0.10 | 0.21 | 0.35 | 2.10 | 0.60 |
| Total Uncer. (%) | 0.50 | 0.24 | 0.22 | 0.29 | 0.40 | 2.13 | 0.74 |
| SI 1D tank | |||||||
| SSDInvSq (%) | 0.20 | 0.20 | 0.20 | 0.20 | 0.20 | 0.50 | 0.50 |
| DepthPDD (%) | 0.58 | 0.13 | 0.10 | 0.21 | 0.35 | 2.70 | 1.40 |
| Total Uncer. (%) | 0.61 | 0.24 | 0.22 | 0.29 | 0.40 | 2.78 | 1.53 |
| Combined uncertainties added in quadrature (%) | |||||||
| SNC 3D scanner vs SNC 3D scanner with spacer | 0.71 | 0.34 | 0.31 | 0.41 | 0.57 | 3.01 | 1.05 |
| SNC 3D scanner vs 1D tank | 0.79 | 0.34 | 0.31 | 0.41 | 0.57 | 3.50 | 1.70 |
1D, one‐dimensional; 3D, Three‐dimensional; SNC, Sun Nuclear Corporation.
Reference dosimetry difference (%) in two different tanks with reproducibility (%) and associated setup uncertainty (%).
| Beam | SNC 3D scanner vs SNC 3D scanner with spacer | SNC 3D scanner vs 1D tank | ||||
|---|---|---|---|---|---|---|
| Mean difference (%) | Reproducibility (1 SD) (%) | Setup uncertainty (%) | Mean difference (%) | Reproducibility (1 SD) (%) | Setup uncertainty (%) | |
| 6 MV | −0.1 | 0.07 | 0.71 | 0.3 | 0.06 | 0.79 |
| 18 MV | 0.0 | 0.03 | 0.34 | 0.2 | 0.03 | 0.34 |
| 6 MeV | 0.0 | 0.07 | 0.31 | 0.1 | 0.07 | 0.31 |
| 12 MeV | 0.0 | 0.14 | 0.41 | 0.0 | 0.21 | 0.41 |
| 20 MeV | 0.1 | 0.06 | 0.57 | −0.1 | 0.25 | 0.57 |
| 100 kV | −1.2 | 0.64 | 3.01 | 0.0 | 0.37 | 3.50 |
| 200 kV | −0.7 | 0.43 | 1.05 | 1.1 | 0.32 | 1.70 |
1D, one‐dimensional; 3D, Three‐dimensional; SNC, Sun Nuclear Corporation.
FIG. 4Relative output against separation between the effective point of measurement of the chamber and the metallic drive rail (MDR) for MV photons. The vertical mark represents the designed chamber — MDR separation in the three‐dimensional scanner for the Farmer‐type chamber which is the distance when the chamber is placed in the Sun Nuclear Corporation supplied standard holder mounted on the MDR.
FIG. 5Relative output against separation between the effective point of measurement of the chamber and the metallic drive rail (MDR) for MV electrons. The vertical mark represents the designed chamber — MDR separation in the three‐dimensional scanner for the ROOS chamber which is the distance when the chamber is placed in the Sun Nuclear Corporation supplied standard holder mounted on the MDR.
FIG. 6Relative output against separation between the effective point of measurement of the chamber and the metallic drive rail (MDR) for kV photons. The vertical mark represents the designed chamber — MDR separation in the three‐dimensional scanner for the Farmer‐type chamber which is the distance when the chamber is placed in the Sun Nuclear Corporation supplied standard holder mounted on the MDR.