| Literature DB >> 25200865 |
Jørn Bersvendsen1, Jan O Beitnes, Stig Urheim, Svend Aakhus, Eigil Samset.
Abstract
BACKGROUND: Transcatheter aortic valve implantation involves percutaneously implanting a biomechanical aortic valve to treat severe aortic stenosis. In order to select a proper device, precise sizing of the aortic valve annulus must be completed.Entities:
Mesh:
Year: 2014 PMID: 25200865 PMCID: PMC4274687 DOI: 10.1186/1471-2342-14-31
Source DB: PubMed Journal: BMC Med Imaging ISSN: 1471-2342 Impact factor: 1.930
Figure 1State estimation KF processing chain. Figure adapted from [6]. and denote the state and covariance estimates respectively, at time index k using measurements up to and including time index j.
Figure 2Example measurements. Manual (a-c) and automatic (d-f) measurements of the aortic valve annulus in a 3D TEE recording. Initial (g) and deformed (h-i) subdivision surface.
Patient characteristics
| Age ∗, yr | 69±16 |
| Sex ∗, male/female | 7/6 |
| Disease, n | |
| Aortic stenosis | 12 |
| Aortic insufficiency | 2 |
| Normal | 2 |
| Aortic morphology, n | |
| Tricuspid | 14 |
| Bicuspid | 2 |
| LV EF < 50 % ∗, n | 3 |
∗Age, sex and LV EF were unknown for 3 patients.
Figure 3Sensitivity analysis results. The plot shows the mean and standard deviation of the output diameter distributions as a function of translational and rotational perturbations.
Automatic and manual measurements of arotic annulus diameters
| Minimum | 24.3 ± 3.0 | | |
| Maximum | 26.8 ± 3.5 | | |
| Sagittal | | 23.9 ± 2.7 | 23.9 ± 3.1 |
| Coronal | | 25.4 ± 3.3 | 25.3 ± 3.0 |
| Area-derived | 25.5 ± 3.2 | 24.9 ± 3.0 | 25.0 ± 3.1 |
| Perimeter-derived | 25.6 ± 3.2 | 25.9 ± 3.2 | 25.8 ± 3.1 |
Values are mean ±SD [mm]. Dauto, D1 and D2 denote automatic and manual measurements from the the first and second observer respectively.
Comparison of automatic and manual measurements of arotic annulus diameter
| Sagittal | | |
| 0.063 ± 1.5 | 0.88 | |
| Coronal | | |
| 0.063 ± 2.4 | 0.74 | |
| Perimeter-derived | | |
| -0.35 ± 1.6 | 0.87 | |
| -0.23 ± 2.3 | 0.75 | |
| 0.12 ± 2.1 | 0.78 | |
| Area-derived | | |
| 0.62 ± 1.7 | 0.85 | |
| 0.46 ± 2.3 | 0.74 | |
| -0.16 ± 2.1 | 0.77 |
Biases are mean ±SD [mm]. Dauto, D1 and D2 denote automatic and manual measurements from the the first and second observer respectively.
Figure 4Comparison of automatic and manual perimeter-derived measurements. All diameters are [mm]. D, D1 and D2 denote automatic and manual measurements from the the first and second observer respectively.