A Monte Carlo evaluation for effects of probable dimensional uncertainties of low dose rate brachytherapy seeds on dose.

Berkay Camgöz1, Mehmet N Kumru1.   

Abstract

The aim of this study is to determine effects of size deviations of brachytherapy seeds on two dimensional dose distributions around the seed. Although many uncertainties are well known, the uncertainties which stem from geometric features of radiation sources are weakly considered and predicted. Neither TG-43 report which is not completely in common consensus, nor individual scientific MC and experimental studies include sufficient data for geometric uncertainties. Sizes of seed and its components can vary in a manufacturing deviation. This causes geometrical uncertainties, too. In this study, three seeds which have different geometrical properties were modeled using EGSnrc-Code Packages. Seeds were designed with all their details using the geometry package. 5% deviations of seed sizes were assumed. Modified seeds were derived from original seed by changing sizes by 5%. Normalizations of doses which were calculated from three kinds of brachytherapy seed and their derivations were found to be about 3%-20%. It was shown that manufacturing differences of brachytherapy seed cause considerable changes in dose distribution.

Keywords:  Brachytherapy; Dose; EGSnrc; Monte Carlo; TG-43 report

Year:  2014        PMID: 25184054      PMCID: PMC4150094          DOI: 10.1016/j.rpor.2014.01.001

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  13 in total

1.  Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version.

Authors:  I Kawrakow
Journal:  Med Phys       Date:  2000-03       Impact factor: 4.071

2.  Monte Carlo modeling of the transverse-axis dose distribution of the model 200 103Pd interstitial brachytherapy source.

Authors:  J F Williamson
Journal:  Med Phys       Date:  2000-04       Impact factor: 4.071

3.  A discretized approach to determining TG-43 brachytherapy dosimetry parameters: case study using Monte Carlo calculations for the MED3633 103Pd source.

Authors:  M J Rivard
Journal:  Appl Radiat Isot       Date:  2001-12       Impact factor: 1.513

4.  Experimental determination of the anisotropy function and anisotropy factor for model 6711 I-125 seeds.

Authors:  R S Sloboda; G V Menon
Journal:  Med Phys       Date:  2000-08       Impact factor: 4.071

5.  An evaluation of the AAPM-TG43 dosimetry protocol for I-125 brachytherapy seed.

Authors:  Paul Mobit; Iulian Badragan
Journal:  Phys Med Biol       Date:  2004-07-21       Impact factor: 3.609

6.  Enhancements to commissioning techniques and quality assurance of brachytherapy treatment planning systems that use model-based dose calculation algorithms.

Authors:  Mark J Rivard; Luc Beaulieu; Firas Mourtada
Journal:  Med Phys       Date:  2010-06       Impact factor: 4.071

7.  Benchmarking brachydose: Voxel based EGSnrc Monte Carlo calculations of TG-43 dosimetry parameters.

Authors:  R E P Taylor; G Yegin; D W O Rogers
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

8.  A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: report of AAPM Task Group No. 138 and GEC-ESTRO.

Authors:  Larry A DeWerd; Geoffrey S Ibbott; Ali S Meigooni; Michael G Mitch; Mark J Rivard; Kurt E Stump; Bruce R Thomadsen; Jack L M Venselaar
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

9.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

10.  Monte Carlo calculations of AAPM Task Group Report No. 43 dosimetry parameters for the MED3631-A/M125I source.

Authors:  M J Rivard
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

View more