Evaluation of a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography of scaphoid fixation screws.

Lukas Filli1, Magda Marcon, Bernhard Scholz, Maurizio Calcagni, Tim Finkenstädt, Gustav Andreisek, Roman Guggenberger.   

Abstract

OBJECTIVES: The aim of this study was to evaluate a prototype correction algorithm to reduce metal artefacts in flat detector computed tomography (FDCT) of scaphoid fixation screws. FDCT has gained interest in imaging small anatomic structures of the appendicular skeleton. Angiographic C-arm systems with flat detectors allow fluoroscopy and FDCT imaging in a one-stop procedure emphasizing their role as an ideal intraoperative imaging tool. However, FDCT imaging can be significantly impaired by artefacts induced by fixation screws.
MATERIALS AND METHODS: Following ethical board approval, commercially available scaphoid fixation screws were inserted into six cadaveric specimens in order to fix artificially induced scaphoid fractures. FDCT images corrected with the algorithm were compared to uncorrected images both quantitatively and qualitatively by two independent radiologists in terms of artefacts, screw contour, fracture line visibility, bone visibility, and soft tissue definition. Normal distribution of variables was evaluated using the Kolmogorov-Smirnov test. In case of normal distribution, quantitative variables were compared using paired Student's t tests. The Wilcoxon signed-rank test was used for quantitative variables without normal distribution and all qualitative variables. A p value of < 0.05 was considered to indicate statistically significant differences.
RESULTS: Metal artefacts were significantly reduced by the correction algorithm (p < 0.001), and the fracture line was more clearly defined (p < 0.01). The inter-observer reliability was "almost perfect" (intra-class correlation coefficient 0.85, p < 0.001).
CONCLUSIONS: The prototype correction algorithm in FDCT for metal artefacts induced by scaphoid fixation screws may facilitate intra- and postoperative follow-up imaging. KEY POINTS: Flat detector computed tomography (FDCT) is a helpful imaging tool for scaphoid fixation. The correction algorithm significantly reduces artefacts in FDCT induced by scaphoid fixation screws. This may facilitate intra- and postoperative follow-up imaging.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25164608     DOI: 10.1007/s00256-014-1986-3

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  26 in total

1.  Fast iterative algorithm for metal artifact reduction in X-ray CT.

Authors:  G Wang; T Frei; M W Vannier
Journal:  Acad Radiol       Date:  2000-08       Impact factor: 3.173

2.  Frequency split metal artifact reduction (FSMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelrieß
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

Review 3.  Flat-detector computed tomography (FD-CT).

Authors:  Willi A Kalender; Yiannis Kyriakou
Journal:  Eur Radiol       Date:  2007-06-23       Impact factor: 5.315

4.  Iterative deblurring for CT metal artifact reduction.

Authors:  G Wang; D L Snyder; J A O'Sullivan; M W Vannier
Journal:  IEEE Trans Med Imaging       Date:  1996       Impact factor: 10.048

5.  Application of a novel metal artifact correction algorithm in flat-panel CT after coil embolization of brain aneurysms: intraindividual comparison.

Authors:  J-H Buhk; M Groth; S Sehner; J Fiehler; N O Schmidt; U Grzyska
Journal:  Rofo       Date:  2013-07-15

6.  Radiation dose of interventional radiology system using a flat-panel detector.

Authors:  Koichi Chida; Yohei Inaba; Haruo Saito; Tadashi Ishibashi; Shoki Takahashi; Masahiro Kohzuki; Masayuki Zuguchi
Journal:  AJR Am J Roentgenol       Date:  2009-12       Impact factor: 3.959

7.  A novel forward projection-based metal artifact reduction method for flat-detector computed tomography.

Authors:  Daniel Prell; Yiannis Kyriakou; Marcel Beister; Willi A Kalender
Journal:  Phys Med Biol       Date:  2009-10-14       Impact factor: 3.609

Review 8.  Review of imaging of scaphoid fractures.

Authors:  Michael Smith; Gregory I Bain; Perry C Turner; Adam C Watts
Journal:  ANZ J Surg       Date:  2010-01       Impact factor: 1.872

Review 9.  The scaphoid.

Authors:  Rosie Sendher; Amy L Ladd
Journal:  Orthop Clin North Am       Date:  2013-01       Impact factor: 2.472

10.  Entrance surface dose during three-dimensional imaging with a flat-panel detector angiography system.

Authors:  Shigeru Suzuki; Shigeru Furui; Ichiro Yamaguchi; Masafumi Yamagishi; Toshi Abe; Ikuo Kobayashi; Takahiro Haruyama
Journal:  J Vasc Interv Radiol       Date:  2008-07-11       Impact factor: 3.464

View more
  2 in total

1.  Prototype metal artefact reduction algorithm in flat panel computed tomography - evaluation in patients undergoing transarterial hepatic radioembolisation.

Authors:  Qeumars Mustafa Hamie; Adrian Raoul Kobe; Leif Mietzsch; Michael Manhart; Gilbert Dominique Puippe; Thomas Pfammatter; Roman Guggenberger
Journal:  Eur Radiol       Date:  2017-07-14       Impact factor: 5.315

2.  Evaluation of a metal artifact reduction algorithm applied to post-interventional flat detector CT in comparison to pre-treatment CT in patients with acute subarachnoid haemorrhage.

Authors:  Angelika Mennecke; Stanislav Svergun; Bernhard Scholz; Kevin Royalty; Arnd Dörfler; Tobias Struffert
Journal:  Eur Radiol       Date:  2016-04-16       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.