Maja Bohac1, Alma Biscevic, Mateja Koncarevic, Marija Anticic, Nikica Gabric, Sudi Patel.
Abstract
PURPOSE: To compare functional outcomes of Wavelight Allegretto Eye-Q 400Hz and Schwind Amaris 750S excimer laser for astigmatism between 2 and 7 diopters(D).Entities:
Mesh:
Year: 2014 PMID: 25150049 PMCID: PMC4181506 DOI: 10.1007/s00417-014-2776-2
Source DB: PubMed Journal: Graefes Arch Clin Exp Ophthalmol ISSN: 0721-832X Impact factor: 3.117
Comparison of visual and refractive results at 1 year with baseline values in myopic astigmatism showing the main results of the investigation
| Visual and refractive results — myopic astigmatism | ||||||
|---|---|---|---|---|---|---|
| Mean ± standard deviation | ||||||
| Variable | Wavelight Allegretto Eye-Q | Schwind Amaris 750S | ||||
| preop | postop |
| preop | postop |
| |
| UDVA | 0.15 ± 0.15 (0.01 to 0.70) | 0.86 ± 0.16 (0.35 to 1.00) | <0.001 | 0.13 ± 0.11 (0.01 to 0.45) | 0.86 ± 0.19 (0.15 to 1.00) | <0.001 |
| CDVA | 0.81 ± 0.17 (0.30 to 1.00) | 0.89 ± 0.16 (0.40 to 1.00) | <0.001 | 0.81 ± 0.18 (0.10 to 1.00) | 0.89 ± 0.20 (0.20 to 1.00) | 0.001 |
| sphere (D) | −2.80 ± 2.01 (−8.50 to 0.00) | −0.16 ± 0.46 (−1.50 to 1.00) | <0.001 | −2.44 ± 2.17 (−7.50 to 0.00) | −0.16 ± 0.55 (−2.00 to 1.25) | <0.001 |
| cylinder (D) | −3.30 ± 1.00 (−7.50 to −2.00) | −0.55 ± 0.46 (−2.25 to 0.00) | <0.001 | −3.21 ± 0.87 (−6.50 to −2.00) | −0.43 ± 0.36 (−1.50 to 0.00) | <0.001 |
UDVA = uncorrected distance visual acuity (decimal)
CDVA = corrected distance visual acuity (decimal)
D = diopter
*2-sample z-test assuming unequal variances
Comparison of visual and refractive results at 1 year with baseline values in mixed astigmatism showing the main results of the investigation
| Visual and refractive results — mixed astigmatism | ||||||
|---|---|---|---|---|---|---|
| Mean ± standard deviation | ||||||
| Variable | Wavelight Allegretto Eye-Q | Schwind Amaris 750S | ||||
| preop | postop |
| preop | postop |
| |
| UDVA | 0.25 ± 0.14 (0.03 to 0.60) | 0.77 ± 0.20 (0.20 to 1.00) | <0.001 | 0.24 ± 0.12 (0.02 to 0.60) | 0.80 ± 0.21 (0.05 to 1.00) | <0.001 |
| CDVA | 0.74 ± 0.22 (0.15 to 1.00) | 0.82 ± 0.21 (0.20 to 1.00) | 0.04 | 0.77 ± 0.20 (0.04 to 0.95) | 0.85 ± 0.21 (0.10 to 1.00) | 0.004 |
| sphere (D) | 2.72 ± 1.79 (0.25 to 7.00) | 0.19 ± 0.52 (−1.50 to +1.50) | <0.001 | 3.11 ± 1.57 (0.50 to 7.50) | 0.28 ± 0.45 (−0.75 to 1.00) | <0.001 |
| cylinder (D) | −3.84 ± 1.21 (−6.50 to −2.00) | −0.85 ± 0.41 (−2.00 to 0.00) | <0.001 | −3.66 ± 1.16 (−7.00 to −2.00) | −0.58 ± 0.38 (−1.50 to 0.00) | <0.001 |
UDVA = uncorrected distance visual acuity (decimal)
CDVA = corrected distance visual acuity (decimal)
D = diopter
*2-sample z-test assuming unequal variances
Fig. 1Attempted vs achieved cylinder correction (D) in myopic astigmatism with Allegretto Eye-Q. The least-squares regression line, best-fit linear equation and R2 are included for comparison with Figs. 2, 3, & 4
Fig. 2Attempted vs achieved cylinder correction (D) in mixed astigmatism with Allegretto Eye-Q. The least-squares regression line, best-fit linear equation and R2 are included for comparison with Figs. 1, 3, & 4
Fig. 3Attempted vs achieved cylinder correction (D) in myopic astigmatism with Amaris 750S. The least-squares regression line, best-fit linear equation and R2 are included for comparison with Figs. 1, 2 & 4
Fig. 4Attempted vs achieved cylinder correction (D) in mixed astigmatism with Amaris 750S. The least-squares regression line, best-fit linear equation and R2 are included for comparison with Figs. 1, 2 & 3
Comparison of high-order aberrations (μm) on 5 mm pupil at 1 year with baseline values in myopic astigmatism, showing the main results of the investigation
| High-order aberrations — myopic astigmatism | ||||||
|---|---|---|---|---|---|---|
| Variable | Mean ± standard deviation | |||||
| Wavelight Allegretto Eye-Q | Schwind Amaris 750S | |||||
| preop | postop |
| preop | postop |
| |
| coma (μm) | 0.12 ± 0.09 (0.01 to 0.53) | 0.11 ± 0.10 (0.01 to 0.90) | 0.592 | 0.11 ± 0.08 (0.00 to 0.40) | 0.13 ± 0.11 (0.01 to 0.80) | 0.166 |
| trefoil (μm) | 0.11 ± 0.06 (0.01 to 0.32) | 0.11 ± 0.13 (0.01 to 0.90) | 0.999 | 0.10 ± 0.06 (0.01 to 0.29) | 0.09 ± 0.08 (0.01 to 0.49) | 0.211 |
| SA (μm) | −0.02 ± 0.07 (−0.28 to 0.20) | 0.00 ± 0.05 (−0.40 to 0.22) | 0.056 | −0.01 ± 0.05 (−0.23 to 0.10) | −0.01 ± 0.07 (−0.26 to 0.14) | 0.504 |
SA = Spherical aberration (μm – micrometer)
*2-sample z-test assuming unequal variances
Comparison of high-order aberrations (μm) on 5 mm pupil at 1 year with baseline values in mixed astigmatism, showing the main results of the investigation
| High-order aberrations — mixed astigmatism | ||||||
|---|---|---|---|---|---|---|
| Variable | Wavelight Allegretto Eye-Q | Schwind Amaris 750S | ||||
| preop | postop |
| preop | postop |
| |
| coma (μm) | 0.12 ± 0.08 (0.02 to 0.45) | 0.10 ± 0.06 (0.04 to 0.28) | 0.347 | 0.12 ± 0.09 (0.01 to 0.46) | 0.11 ± 0.03 (0.01 to 0.53) | 0.420 |
| trefoil (μm) | 0.13 ± 0.13 (0.01 to 0.90) | 0.10 ± 0.05 (0.02 to 0.26) | 0.116 | 0.10 ± 0.07 (0.01 to 0.44) | 0.09 ± 0.08 (0.01 to 0.76) | 0.404 |
| SA (μm) | 0.02 ± 0.06 (−0.11 to 0.14) | 0.00 ± 0.04 (−0.09 to 0.12) | 0.03 | 0.02 ± 0.05 (−0.07 to 0.20) | 0.00 ± 0.04 (−0.18 to 0.10) | <0.001 |
SA = Spherical aberration (μm – micrometer)
*2-sample z-test assuming unequal variances