Effect of local heating on postprandial blood glucose excursions using the InsuPad device: results of an outpatient crossover study.

Norbert Hermanns1, Gabriel Bitton2, André Reimer3, Michael Krichbaum3, Bernhard Kulzer3, Thomas Haak3.   

Abstract

The InsuPad is a medical device to accelerate insulin resorption by applying local heat at the insulin injection site. This crossover study examined the impact of the InsuPad use on postprandial glucose excursions under daily life conditions. In 1 study phase, diabetic patients used the InsuPad when injecting bolus insulin before breakfast and dinner and measured their blood glucose 5 times daily (before breakfast, lunch, and dinner and after breakfast and dinner). In the other study phase, blood glucose measurements were maintained without using the InsuPad. The order of the study phases was randomized. Twenty patients with a high insulin demand took part (30% type 1 diabetes, age 53.7 ± 8.9 years, diabetes duration 14.9 ± 7.4 years; HbA1c 8.3 ± 0.8%; total daily insulin demand 0.97 ± 0.32 IU per kg). Postprandial glucose excursion was reduced by 15.4 mg/dl (95% CI 9.7-21.2 mg/dl; P = .011) after breakfast and dinner if InsuPad was used. The mean blood glucose was lower by 8.8 mg/dl (95% CI 0:3-18:0 mg/dl; P = .099) when using the InsuPad. Safety parameters and the percentage of hypoglycemic (< 60 mg/dl) or hyperglycemic (> 300 mg/dl) blood glucose measurements were not negatively affected by InsuPad use (hypoglycemic values 1.4% vs 1.5%, P = .961; hyperglycemic values 2.6% vs 4.0%, P = .098). Local heating of the insulin injection site by use of the InsuPad device is an effective and safe method to reduce postprandial blood glucose excursions under daily life conditions without negative side effects on the occurrence of low or high blood glucose values.
© 2014 Diabetes Technology Society.

Entities:  

Keywords:  InsuPad; blood glucose values; glycemic variability; postprandial glucose excursion

Mesh:

Substances:

Year:  2014        PMID: 25113814      PMCID: PMC4455478          DOI: 10.1177/1932296814546162

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  5 in total

Review 1.  Postprandial blood glucose. American Diabetes Association.

Authors: 
Journal:  Diabetes Care       Date:  2001-04       Impact factor: 19.112

2.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

3.  Assessment of diabetes-related distress.

Authors:  W H Polonsky; B J Anderson; P A Lohrer; G Welch; A M Jacobson; J E Aponte; C E Schwartz
Journal:  Diabetes Care       Date:  1995-06       Impact factor: 19.112

4.  Effect of a local heating device on insulin and glucose pharmacokinetic profiles in an open-label, randomized, two-period, one-way crossover study in patients with type 1 diabetes using continuous subcutaneous insulin infusion.

Authors:  Itamar Raz; Ram Weiss; Yevgeny Yegorchikov; Gabriel Bitton; Ron Nagar; Benny Pesach
Journal:  Clin Ther       Date:  2009-05       Impact factor: 3.393

5.  Acceleration of insulin pharmacodynamic profile by a novel insulin infusion site warming device.

Authors:  Eda Cengiz; Stuart A Weinzimer; Jennifer L Sherr; Eileen Tichy; Melody Martin; Lori Carria; Amy Steffen; William V Tamborlane
Journal:  Pediatr Diabetes       Date:  2012-10-28       Impact factor: 4.866

  5 in total
  4 in total

1.  The effect of passive heating on heat shock protein 70 and interleukin-6: A possible treatment tool for metabolic diseases?

Authors:  S H Faulkner; S Jackson; G Fatania; C A Leicht
Journal:  Temperature (Austin)       Date:  2017-03-09

Review 2.  [New technologies in diabetology. How far are we from a closed loop?].

Authors:  G Freckmann
Journal:  Internist (Berl)       Date:  2015-05       Impact factor: 0.743

3.  Towards a Physiological Prandial Insulin Profile: Enhancement of Subcutaneously Injected Prandial Insulin Using Local Warming Devices.

Authors:  Ahmed H El-Laboudi; Nick Oliver
Journal:  Diabetes Ther       Date:  2015-09-02       Impact factor: 2.945

Review 4.  Body temperature regulation in diabetes.

Authors:  Glen P Kenny; Ronald J Sigal; Ryan McGinn
Journal:  Temperature (Austin)       Date:  2016-01-04
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.