Accuracy and precision of the ultrasound cardiac output monitor (USCOM 1A) in pregnancy: comparison with three-dimensional transthoracic echocardiography.

H McNamara1, P Barclay2, V Sharma3.   

Abstract

BACKGROUND: Cardiac output (CO) monitoring is helpful in the assessment of critically ill pregnant women, but invasive monitors are often unsuitable for use. We aimed to measure agreement between the non-invasive ultrasound cardiac output monitor (USCOM) and three-dimensional transthoracic echocardiography (3D-TTE) in pregnant women.
METHODS: Healthy pregnant women from 25 weeks gestation onwards participated. In the left lateral position at rest, CO was measured with the USCOM and 3D-TTE. A single operator performed all USCOM measurements, with a different operator performing all echocardiography. Both were blinded to results from the other device. Each USCOM trace was analysed using two modes: flowtrace (FT) and touchpoint (TP). A second, blinded USCOM reading was taken to assess reproducibility.
RESULTS: USCOM readings were obtained in 92, and 3D-TTE images in 85 participants. The mean CO was 5.7, 7.7, and 6.2 litre min(-1) measured by 3D-TTE, USCOM FT, and USCOM TP, respectively. USCOM bias was +2.0 litre min(-1) (FT) and +0.4 litre min(-1) (TP). Limits of agreement were -0.2 to +4.2 litre min(-1) (FT) and -1.4 to +2.3 litre min(-1) (TP). The mean percentage difference was 32.6% (FT) and 31.4% (TP) for CO and 27.0% (FT) and 27.5% (TP) for stroke volume. Intraclass correlation between repeated USCOM readings was 0.9 (FT) and 0.86 (TP).
CONCLUSIONS: USCOM has acceptable agreement with 3D-TTE for the measurement of CO in pregnancy. The positive bias of the USCOM, particularly in the FT mode, may be due to the hyperdynamic cardiovascular state in pregnancy. We suggest using the TP mode in this patient population.
© The Author 2014. Published by Oxford University Press on behalf of the British Journal of Anaesthesia. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Entities:  

Keywords:  heart, cardiac output; measurement techniques, cardiac output; monitoring, echocardiography; monitoring, ultrasound; pregnancy

Mesh:

Year:  2014        PMID: 25005720     DOI: 10.1093/bja/aeu162

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  7 in total

1.  Accuracy of the ClearSight™ system in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Miho Sumiyoshi; Takuma Maeda; Erika Miyazaki; Naoshi Hotta; Hitoshi Sato; Eisuke Hamaguchi; Hiroko Kanazawa; Yoshihiko Ohnishi; Masataka Kamei
Journal:  J Anesth       Date:  2019-03-23       Impact factor: 2.078

2.  Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery.

Authors:  Takuma Maeda; Kohshi Hattori; Miho Sumiyoshi; Hiroko Kanazawa; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2018-04-03       Impact factor: 2.078

3.  Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology.

Authors:  Wonshill Koh; Kristin A Schneider; Huaiyu Zang; Sarosh P Batlivala; Matthew P Monteleone; Alexis L Benscoter; Meghan M Chlebowski; Ilias D Iliopoulos; David S Cooper
Journal:  Pediatr Cardiol       Date:  2022-02-06       Impact factor: 1.838

4.  Validation of an Ultrasound Cardiac Output Monitor as a Bedside Tool for Pediatric Patients.

Authors:  Fernando Beltramo; Jondavid Menteer; Asma Razavi; Robinder G Khemani; Jacqueline Szmuszkovicz; Christopher J L Newth; Patrick A Ross
Journal:  Pediatr Cardiol       Date:  2015-09-12       Impact factor: 1.655

Review 5.  Understanding the Haemodynamics of Hypertension.

Authors:  B E Smith; V M Madigan
Journal:  Curr Hypertens Rep       Date:  2018-04-10       Impact factor: 5.369

6.  The normal ranges of cardiovascular parameters measured using the ultrasonic cardiac output monitor.

Authors:  Giles N Cattermole; P Y Mia Leung; Grace Y L Ho; Peach W S Lau; Cangel P Y Chan; Stewart S W Chan; Brendan E Smith; Colin A Graham; Timothy H Rainer
Journal:  Physiol Rep       Date:  2017-03

7.  Temporal haemodynamic changes after bisoprolol treatment in patients with uncontrolled hypertension.

Authors:  Weiwei Zeng; Brian Tomlinson
Journal:  Ann Transl Med       Date:  2021-06
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.