Comparison of effective dose and lifetime risk of cancer incidence of CT attenuation correction acquisitions and radiopharmaceutical administration for myocardial perfusion imaging.

A K Tootell1, K Szczepura, P Hogg.   

Abstract

OBJECTIVE: To measure the organ dose and calculate effective dose from CT attenuation correction (CTAC) acquisitions from four commonly used gamma camera single photon emission CT/CT systems.
METHODS: CTAC dosimetry data was collected using thermoluminescent dosemeters on GE Healthcare's Infinia™ Hawkeye™ (GE Healthcare, Buckinghamshire, UK) four- and single-slice systems, Siemens Symbia™ T6 (Siemens Healthcare, Erlangen, Germany) and the Philips Precedence (Philips Healthcare, Amsterdam, Netherlands). Organ and effective dose from the administration of (99m)Tc-tetrofosmin and (99m)Tc-sestamibi were calculated using International Commission of Radiological Protection reports 80 and 106. Using these data, the lifetime biological risk was calculated.
RESULTS: The Siemens Symbia gave the lowest CTAC dose (1.8 mSv) followed by the GE Infinia Hawkeye single-slice (1.9 mSv), GE Infinia Hawkeye four-slice (2.5 mSv) and Philips Precedence v. 3.0. Doses were significantly lower than the calculated doses from radiopharmaceutical administration (11 and 14 mSv for (99m)Tc-tetrofosmin and (99m)Tc-sestamibi, respectively). Overall lifetime biological risks were lower, which suggests that using CTAC data posed minimal risk to the patient. Comparison of data for breast tissue demonstrated a higher risk than that from the radiopharmaceutical administration.
CONCLUSION: CTAC doses were confirmed to be much lower than those from radiopharmaceutical administration. The localized nature of the CTAC exposure compared to the radiopharmaceutical biological distribution indicated dose and risk to the breast to be higher. ADVANCES IN KNOWLEDGE: This research proved that CTAC is a comparatively low-dose acquisition. However, it has been shown that there is increased risk for breast tissue especially in the younger patients. As per legislation, justification is required and CTAC should only be used in situations that demonstrate sufficient net benefit.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24998249      PMCID: PMC4453143          DOI: 10.1259/bjr.20140110

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  21 in total

1.  The value and practice of attenuation correction for myocardial perfusion SPECT imaging: a joint position statement from the American Society of Nuclear Cardiology and the Society of Nuclear Medicine.

Authors:  Robert C Hendel; James R Corbett; S James Cullom; E Gordon DePuey; Ernest V Garcia; Timothy M Bateman
Journal:  J Nucl Cardiol       Date:  2002 Jan-Feb       Impact factor: 5.952

2.  16-detector multislice CT: dosimetry estimation by TLD measurement compared with Monte Carlo simulation.

Authors:  A M Groves; K E Owen; H M Courtney; S J Yates; K E Goldstone; G M Blake; A K Dixon
Journal:  Br J Radiol       Date:  2004-08       Impact factor: 3.039

3.  American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: attenuation correction of myocardial perfusion SPECT scintigraphy.

Authors:  Gary V Heller; Jonathan Links; Timothy M Bateman; Jack A Ziffer; Edward Ficaro; Mylan C Cohen; Robert C Hendel
Journal:  J Nucl Cardiol       Date:  2004 Mar-Apr       Impact factor: 5.952

4.  1990 Recommendations of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  1991

5.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

6.  EANM/ESC guidelines for radionuclide imaging of cardiac function.

Authors:  B Hesse; T B Lindhardt; W Acampa; C Anagnostopoulos; J Ballinger; J J Bax; L Edenbrandt; A Flotats; G Germano; T Gmeiner Stopar; P Franken; A Kelion; A Kjaer; D Le Guludec; M Ljungberg; A F Maenhout; C Marcassa; J Marving; F McKiddie; W M Schaefer; L Stegger; R Underwood
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-04       Impact factor: 9.236

7.  Effective dose: a flawed concept that could and should be replaced.

Authors:  D J Brenner
Journal:  Br J Radiol       Date:  2008-04-28       Impact factor: 3.039

8.  Active bone marrow distribution as a function of age in humans.

Authors:  M Cristy
Journal:  Phys Med Biol       Date:  1981-05       Impact factor: 3.609

9.  Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction.

Authors:  Lynne M Hurwitz; Robert E Reiman; Terry T Yoshizumi; Philip C Goodman; Greta Toncheva; Giao Nguyen; Carolyn Lowry
Journal:  Radiology       Date:  2007-10-08       Impact factor: 11.105

10.  A free-response evaluation determining value in the computed tomography attenuation correction image for revealing pulmonary incidental findings: a phantom study.

Authors:  John D Thompson; Peter Hogg; David J Manning; Katy Szczepura; Dev P Chakraborty
Journal:  Acad Radiol       Date:  2014-04       Impact factor: 3.173

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.