Correlation of Back Optic Zone Radius measurement of rigid contact lenses with radiuscope and keratometer.

Tukaram Kumbar1, T Shyam Sunder, Jawalkar Swati, Asma Fathima, U K Addepally, Harsha L Rao, Varsha M Rathi.   

Abstract

PURPOSE: To investigate the accuracy of Back Optic Zone Radius (BOZR, base curve) measurements with manual keratometer for spherical polymethylmethacrylate (PMMA) lenses.
METHODS: 100 spherical PMMA contact lenses were selected randomly from the trial sets. One investigator measured the BOZR with radiuscope. The second investigator measured the BOZR with keratometer. The two investigators were masked to each other's readings. As the keratometer is designed to measure a convex corneal surface, the BOZR readings are less than actual radii. A correction factor of 0.025 was used to have corrected keratometers readings.
RESULTS: There was a high correlation between the BOZR measurement with radiuscope and the uncorrected (Pearson's correlation coefficient, r=0.99) and corrected (r=0.99) measurements with keratometer. The mean difference between BOZR measurement with radiuscope and the uncorrected and corrected measurements with keratometer were 0.003 mm (p=0.62) and -0.024 mm (p<0.001) respectively. The 95% limits of agreement for the BOZR measurements with radiuscope and the uncorrected and corrected measurements with keratometer were -0.11 to 0.11 mm and -0.14 to 0.09 mm respectively.
CONCLUSION: We found that the uncorrected BOZR measurements with keratometer were similar to the BOZR measurements with radiuscope. With unavailability of radiuscope in majority of contact lens practice, keratometer can be used to measure the BOZR of contact lenses.
Copyright © 2012 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23017211     DOI: 10.1016/j.clae.2012.09.001

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  1 in total

1.  Spectral optical coherence tomography vs. fluorescein pattern for rigid gas-permeable lens fit.

Authors:  Ilona Piotrowiak; Bartłomiej Jan Kaluzny; Beata Danek; Adam Chwiędacz; Bartosz Lukasz Sikorski; Grażyna Malukiewicz
Journal:  Med Sci Monit       Date:  2014-07-04
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.