Comparison of glucose concentrations in canine whole blood, plasma, and serum measured with a veterinary point-of-care glucometer.

Natalie D Suchowersky1, Elizabeth A Carlson1, Hollie P Lee1, Ellen N Behrend1.   

Abstract

Previous studies have determined that, compared to whole blood, serum or plasma used in a portable blood glucometer (PBG) may provide more accurate results. We investigated the accuracy of a veterinary PBG (AlphaTRAK 2; Zoetis) for the measurement of glucose concentrations in serum, plasma, and whole blood compared to plasma glucose concentration measured by a biochemical analyzer. Blood samples from 53 client-owned dogs were collected. Lin concordance correlation coefficient (ρc) and Bland-Altman plots were used to determine correlation and agreement between the results obtained for the different sample types. Glucose concentration in whole blood measured by the veterinary PBG was more strongly correlated with the glucose concentration measured by the biochemical analyzer (ρc = 0.92) compared to plasma and serum glucose concentrations (ρc = 0.59 and 0.57, respectively). The mean differences between the glucose concentrations in whole blood, plasma, and serum measured by the veterinary PBG and the glucose concentration determined by the biochemical analyzer were 1.0, 6.3, and 6.7 mmol/L (18, 113, and 121 mg/dL), respectively. Our findings suggest that, when using this veterinary PBG, the accuracy of a glucose measurement obtained is higher when using whole blood compared to plasma or serum. Use of whole blood allows for more correct assessment and diagnosis, which are necessary for appropriate therapeutic intervention.

Entities:  

Keywords:  diabetes mellitus; dogs; error grid; portable blood glucometer

Mesh:

Substances:

Year:  2021        PMID: 34075827      PMCID: PMC8229823          DOI: 10.1177/10406387211019755

Source DB:  PubMed          Journal:  J Vet Diagn Invest        ISSN: 1040-6387            Impact factor:   1.569


  17 in total

1.  Evaluation of five portable blood glucose meters for use in dogs.

Authors:  G Wess; C Reusch
Journal:  J Am Vet Med Assoc       Date:  2000-01-15       Impact factor: 1.936

2.  Comparison of glucose concentrations in serum, plasma, and blood measured by a point-of-care glucometer with serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples.

Authors:  Matthew J Lechner; Rebecka S Hess
Journal:  Am J Vet Res       Date:  2019-12       Impact factor: 1.156

3.  Correlation between glucose concentrations in serum, plasma, and whole blood measured by a point-of-care glucometer and serum glucose concentration measured by an automated biochemical analyzer for canine and feline blood samples.

Authors:  Barbara S Tauk; Kenneth J Drobatz; Koranda A Wallace; Rebecka S Hess
Journal:  J Am Vet Med Assoc       Date:  2015-06-15       Impact factor: 1.936

4.  Coefficient of accuracy and concordance correlation coefficient: new statistics for methods comparison.

Authors:  L Lin; L D Torbeck
Journal:  PDA J Pharm Sci Technol       Date:  1998 Mar-Apr

5.  Formulation and validation of a predictive model to correct blood glucose concentrations obtained with a veterinary point-of-care glucometer in hemodiluted and hemoconcentrated canine blood samples.

Authors:  Selena L Lane; Amie Koenig; Benjamin M Brainard
Journal:  J Am Vet Med Assoc       Date:  2015-02-01       Impact factor: 1.936

6.  The clinical utility of two human portable blood glucose meters in canine and feline practice.

Authors:  Asuka Domori; Ayano Sunahara; Morihiro Tateno; Takako Shimokawa Miyama; Asuka Setoguchi; Yasuyuki Endo
Journal:  Vet Clin Pathol       Date:  2014-01-21       Impact factor: 1.180

7.  The distribution of sodium, potassium and glucose in the blood of some mammals.

Authors:  M F Coldman; W Good
Journal:  Comp Biochem Physiol       Date:  1967-04

8.  Reliability of point-of-care capillary blood glucose measurements in the critical value range.

Authors:  Ron B Schifman; Tan T Nguyen; Susan T Page
Journal:  Arch Pathol Lab Med       Date:  2014-07       Impact factor: 5.534

9.  Evaluation of four portable blood glucose meters in diabetic and non-diabetic dogs and cats.

Authors:  Min-Hee Kang; Do-Hyung Kim; In-Seong Jeong; Gab-Chol Choi; Hee-Myung Park
Journal:  Vet Q       Date:  2015-10-07       Impact factor: 3.320

10.  ISO-based assessment of accuracy and precision of glucose meters in dogs.

Authors:  Y Brito-Casillas; P Figueirinhas; J C Wiebe; L López-Ríos; D Pérez-Barreto; C Melián; A M Wägner
Journal:  J Vet Intern Med       Date:  2014-07-02       Impact factor: 3.333

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.