Transparent thin shield for radio frequency transmit coils.

Debra S Rivera1, Jessica Schulz, Thomas Siegert, Verena Zuber, Robert Turner.   

Abstract

OBJECTIVE: To identify a shielding material compatible with optical head-motion tracking for prospective motion correction and which minimizes radio frequency (RF) radiation losses at 7 T without sacrificing line-of-sight to an imaging target.
MATERIALS AND METHODS: We evaluated a polyamide mesh coated with silver. The thickness of the coating was approximated from the composition ratio provided by the material vendor and validated by an estimate derived from electrical conductivity and light transmission measurements. The performance of the shield is compared to a split-copper shield in the context of a four-channel transmit-only loop array.
RESULTS: The mesh contains less than a skin-depth of silver coating (300 MHz) and attenuates light by 15 %. Elements of the array vary less in the presence of the mesh shield as compared to the split-copper shield indicating that the array behaves more symmetrically with the mesh shield. No degradation of transmit efficiency was observed for the mesh as compared to the split-copper shield.
CONCLUSION: We present a shield compatible with future integration of camera-based motion-tracking systems. Based on transmit performance and eddy-current evaluations the mesh shield is appropriate for use at 7 T.

Entities:  

Mesh:

Year:  2014        PMID: 24862451     DOI: 10.1007/s10334-014-0448-1

Source DB:  PubMed          Journal:  MAGMA        ISSN: 0968-5243            Impact factor:   2.310


  10 in total

1.  Characterization and reduction of gradient-induced eddy currents in the RF shield of a TEM resonator.

Authors:  Marcello Alecci; Peter Jezzard
Journal:  Magn Reson Med       Date:  2002-08       Impact factor: 4.668

2.  A conformal transceive array for 7 T neuroimaging.

Authors:  Kyle M Gilbert; Jean-Guy Belliveau; Andrew T Curtis; Joseph S Gati; L Martyn Klassen; Ravi S Menon
Journal:  Magn Reson Med       Date:  2011-08-16       Impact factor: 4.668

3.  An embedded optical tracking system for motion-corrected magnetic resonance imaging at 7T.

Authors:  Jessica Schulz; Thomas Siegert; Enrico Reimer; Christian Labadie; Julian Maclaren; Michael Herbst; Maxim Zaitsev; Robert Turner
Journal:  MAGMA       Date:  2012-06-13       Impact factor: 2.310

4.  Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field.

Authors:  Vasily L Yarnykh
Journal:  Magn Reson Med       Date:  2007-01       Impact factor: 4.668

5.  Fast MRI coil analysis based on 3-D electromagnetic and RF circuit co-simulation.

Authors:  Mikhail Kozlov; Robert Turner
Journal:  J Magn Reson       Date:  2009-06-09       Impact factor: 2.229

6.  Optimized RF shielding techniques for simultaneous PET/MR.

Authors:  D Truhn; F Kiessling; V Schulz
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

7.  7 Tesla compatible in-bore display for functional magnetic resonance imaging.

Authors:  Jens Groebner; Moritz Cornelius Berger; Reiner Umathum; Michael Bock; Jaane Rauschenberg
Journal:  MAGMA       Date:  2013-01-05       Impact factor: 2.310

8.  On consideration of radiated power in RF field simulations for MRI.

Authors:  Wanzhan Liu; Chien-ping Kao; Christopher M Collins; Michael B Smith; Qing X Yang
Journal:  Magn Reson Med       Date:  2012-04-03       Impact factor: 4.668

9.  7T head volume coils: improvements for rostral brain imaging.

Authors:  Nikolai I Avdievich; Hoby P Hetherington; Andrey M Kuznetsov; Jullie W Pan
Journal:  J Magn Reson Imaging       Date:  2009-02       Impact factor: 4.813

10.  Prospective head-movement correction for high-resolution MRI using an in-bore optical tracking system.

Authors:  Lei Qin; Peter van Gelderen; John Andrew Derbyshire; Fenghua Jin; Jongho Lee; Jacco A de Zwart; Yang Tao; Jeff H Duyn
Journal:  Magn Reson Med       Date:  2009-10       Impact factor: 4.668

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.