Katarzyna Wypych-Klunder1, Andrzej Adamowicz2, Adam Lemanowicz1, Wojciech Szczęsny2, Zbigniew Włodarczyk2, Zbigniew Serafin1.
Abstract
BACKGROUND: An aim of this study was to assess the feasibility of DWI in the early period after kidney transplantation. We also aimed to compare ADC and eADC values in the cortex and medulla of the kidney, to estimate image noise and variability of measurements, and to verify possible relation between selected labolatory results and diffusion parameters in the transplanted kidney. MATERIAL/Entities:
Keywords: Diffusion Magnetic Resonance Imaging; Kidney Transplantation; Magnetic Resonance Imaging
Year: 2014 PMID: 24826200 PMCID: PMC4018246 DOI: 10.12659/PJR.890502
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Figure 1An example of measurement of diffusion coefficients in the transplanted kidney.
Clinical characteristics of the study population.
| Parameter | Mean ±SD |
|---|---|
| Creatinine [mg/dl] | 3.15±2.77 |
| eGFR [ml/min./1.73 m2] | 33.0±17.0 |
| Urine protein [mg/dl] | 36.0±62.2 |
| Uric acid [mg/dl] | 6.32±1.28 |
| Glucose [mg/dl] | 112.1±57.8 |
| WBC [103/μl] | 9.12±2.35 |
| RBC [106/μl] | 3.50±0.39 |
| HCT [%] | 31.71 ±3.33 |
| PLT [103/μl] | 216.4 ±63.8 |
ADC and eADC values, signal-to-noise ratio (SNR), and variability of measurements.
| Parameter | Mean ±SD | Mean SNR | Variability of measurements |
|---|---|---|---|
| Cortexb600 | 2.10±0.11 [10−3 mm2/s] | 46.0 | 5% |
| Medullab600 | 2.23±0.22 [10−3 mm2/s] | 27.8 | 10% |
| Cortexb1000 | 1.90±0.09 [10−3 mm2/s] | 36.8 | 4% |
| Medullab1000 | 2.28±0.16 [10−3 mm2/s] | 23.3 | 7% |
| Cortexb600 | 0.29±0.02 [mm2/s] | 36.6 | 7% |
| Medullab600 | 0.28±0.04 [mm2/s] | 22.0 | 14% |
| Cortexb1000 | 0.17±0.01 [mm2/s] | 21.1 | 8% |
| Medullab1000 | 0.14±0.02 [mm2/s] | 11.7 | 15% |
Mean SNR and variability of measurements were significantly different between groups (p<0.0001, ANOVA).
Figure 2Comparison of measurement variability at b600 and b1000 (A, p>0.9), between ADC and eADC (B, p<0.0001), and between the cortex and the medulla of the kidney (C, p<0.001). Graphs include mean values and their 95% confidence intervals.
Figure 3Comparison of SNR at b600 and b1000 (A, p>0.0001), between ADC and eADC (B, p<0.0001), and between the cortex and the medulla of the kidney (C, p<0.001). Graphs include mean values and their 95% confidence intervals.
Comparison of ADC and eADC values in the kidney cortex measured at b1000 in patients stratified according to estimated glomerular filtration rate (eGFR). In parentheses, 95% confidence intervals of means are given.
| Parameter | eGFR ≤30 ml/min./1.73 m2 | eGFR >30 ml/min./1.73 m2 | p |
|---|---|---|---|
| ADCb1000 [10−3 mm2/s] | 1.81 (1.71–1.92) | 2.02 (1.87–2.16) | 0.0206 |
| eADCb1000 [mm2/s] | 0.16 (0.14–0.18) | 0.18 (0.16–0.21) | 0.0497 |