Diffusion-weighted MR imaging of transplanted kidneys: Preliminary report.

Katarzyna Wypych-Klunder1, Andrzej Adamowicz2, Adam Lemanowicz1, Wojciech Szczęsny2, Zbigniew Włodarczyk2, Zbigniew Serafin1.   

Abstract

BACKGROUND: An aim of this study was to assess the feasibility of DWI in the early period after kidney transplantation. We also aimed to compare ADC and eADC values in the cortex and medulla of the kidney, to estimate image noise and variability of measurements, and to verify possible relation between selected labolatory results and diffusion parameters in the transplanted kidney. MATERIAL/
METHODS: Examinations were performed using a 1.5 T MR unit. DWI (SE/EPI) was performed in the axial plane using b-values of 600 and 1000. ADC and eADC measurements were performed in four regions of interest within the renal cortex and in three regions within the medulla. Relative variability of results and signal-to-noise ratio (SNR) were calculated.
RESULTS: The analysis included 15 patients (mean age 52 years). The mean variability of ADC was significantly lower than that of eADC (6.8% vs. 10.8%, respectively; p<0.0001). The mean variability of measurements performed in the cortex was significantly lower than that in the medulla (6.2% vs. 11.5%, respectively; p<0.005). The mean SNR was higher in the measurements using b600 than b1000, it was higher in ADC maps than in the eADC maps, and it was higher in the cortex than in the medulla. ADC and eADC measured at b1000 in the cortex were higher in the group of the patients with eGFR ≤30 ml/min./1.73 m(2) as compared to patients with eGFR >30 ml/min./1.73 m(2) (p<0.05).
CONCLUSIONS: Diffusion-weighted imaging of transplanted kidneys is technically challenging, especially in patients in the early period after transplantation. From a technical point of view, the best quality parameters offer quality ADC measurement in the renal cortex using b1000. ADC and eADC values in the renal cortex measured at b1000 present a relationship with eGFR.

Entities:  

Keywords:  Diffusion Magnetic Resonance Imaging; Kidney Transplantation; Magnetic Resonance Imaging

Year:  2014        PMID: 24826200      PMCID: PMC4018246          DOI: 10.12659/PJR.890502

Source DB:  PubMed          Journal:  Pol J Radiol        ISSN: 1733-134X


Background

Renal transplantation is the most successful method of treating patients with advanced chronic renal disease. It is performed in dialysed patients and as so-called preemptive transplantations in patients due to start dialyses. This procedure significantly improves patients’ quality of life, as it decreases the number of complications and limitations associated with renal disease. Renal transplantation is commonly performed. However, there is no efficient and non-invasive method of assessing graft function and identifying parenchymal abnormalities of the kidney [1]. In practice, biopsy is the only method which allows to identify parenchymal abnormalities of the graft, such as acute and chronic rejection, acute glomerular necrosis, drug-induced injury or delayed graft function after transplantation. A significant problem associated with this method is a substantial number of non-diagnostic bioptats and a certain number of false results. Extended monitoring using biopsy is controversial. Methods of functional magnetic resonance imaging of the kidneys can become a response to these needs. Diffusion-weighted imaging (DWI) provides information on diffusion and perfusion of the kidney, which allows to assess integrity of the renal tubules and condition of the renal microvasculature [2]. Diffusion tensor imaging (DTI) additionally shows diffusion directionality, which may be important due to high anisotropy of the renal ultrastructure [3]. Blood oxygenation level-dependent imaging (BOLD) provides information on metabolism and it seems possible that reduction of the oxygenation seen in BOLD will allow to distinguish acute graft rejection [4]. Finally, perfusion scan using ASL method, which gives insight into the graft parenchymal flow, can help to monitor the chronically decreased renal function [5]. In the literature individual clinical studies on the use of DWI in imaging of transplanted kidneys can be found. This is a newly developed method and thus lacks standard study protocols [6]. Optimal parameters of diffusion gradients (value b), selection of measurement site (cortex and medulla) and method of presentation of diffusion images are discussed, as they are the basis of the measurements: ADC (apparent diffusion coefficient) and eADC (exponential apparent diffusion coefficient). The aim of this study was to assess the feasibility of DWI in the early period after kidney transplantation, to compare ADC and eADC values in the cortex and medulla of the kidney, image noise and variability of measurements, and to verify possible relation between selected laboratory results and diffusion parameters in the transplanted kidney.

Material and Methods

Examinations were performed using a 1.5 T MR unit (SignaHDxt GEHealthcare) and a 16-channel body coil. Diffusion-weighted imaging (SE/EPI) was performed in axial plane using b-values: b600 (TR 9231, TE 69.4/FE, trig. 30%, EC 1/1, bandwidth 250 kHz, FOV 37×37, slice 7.0 mm, matrix 128×128, NEX 4) and b1000 (TR 10000, TE 96.2/FE, trig. 30%, bandwidth 250kHz, FOV 37×37, slice 7.0 mm, matrix 128×128, NEX4). ADC and eADC measurements were performed in four random regions of interest (ROI) within the renal cortex and in three random regions of interest within the medulla of transplanted kidneys (Figure 1). The surface area of ROI was 25 mm2. Relative variability of ADC and eADC measurement results was calculated for every patient as a ratio of mean value from all ROI of examined area to its standard deviation. Signal-to-noise ratio (SNR) was calculated as a ratio of mean ADC or eADC value within single ROI to its standard deviation [7]. Mean SNR values of all ROI in the given renal area for every patient were included in the calculations, and then averaged between the patients.
Figure 1

An example of measurement of diffusion coefficients in the transplanted kidney.

Consistence of parametric data distribution with normal distribution was assessed using Kolmogorov-Smirnov test with Lilliefors correction. ANOVA test for systems with repeated measurements was used for combined comparison of variability of measurements and SNR. Detailed analysis of difference significance was conducted with use of dependent t-test or Wilcoxon signed-rank test. Correlations between variables were assessed using Pearson’s or Spearman’s correlation coefficients.

Results

A total of 15 patients (11 men and 4 women) aged 24–71 (mean 52 years) were analyzed. Mean time from transplantation was 15 days. A total of 5 patients were excluded due to inability to complete the study or the presence of non-diagnostic images with significant motion artifacts. Clinical characteristics of the study population was presented in Table 1.
Table 1

Clinical characteristics of the study population.

ParameterMean ±SD
Creatinine [mg/dl]3.15±2.77
eGFR [ml/min./1.73 m2]33.0±17.0
Urine protein [mg/dl]36.0±62.2
Uric acid [mg/dl]6.32±1.28
Glucose [mg/dl]112.1±57.8
WBC [103/μl]9.12±2.35
RBC [106/μl]3.50±0.39
HCT [%]31.71 ±3.33
PLT [103/μl]216.4 ±63.8
ADC and eADC values, mean variability of their measurements and mean SNR were presented in Table 2. Comparison of values measured in the cortex and medulla revealed a significant difference only in ADC using b1000 (p<0.01). A significant strong correlation was shown between the variability of ADC measurements and eADC measurements in all groups of parameters (renal cortex and medulla, b-values b600 and b1000); correlation coefficients were between 0.88 and 0.98. Significant strong correlations between SNR values of ADC and eADC images were also revealed in all groups of parameters (renal cortex and medulla, b-values b600 and b1000); correlation coefficients were between 0.88 and 0.98.
Table 2

ADC and eADC values, signal-to-noise ratio (SNR), and variability of measurements.

ParameterMean ±SDMean SNR*Variability of measurements*
ADC
Cortexb6002.10±0.11 [10−3 mm2/s]46.05%
Medullab6002.23±0.22 [10−3 mm2/s]27.810%
Cortexb10001.90±0.09 [10−3 mm2/s]36.84%
Medullab10002.28±0.16 [10−3 mm2/s]23.37%
eADC
Cortexb6000.29±0.02 [mm2/s]36.67%
Medullab6000.28±0.04 [mm2/s]22.014%
Cortexb10000.17±0.01 [mm2/s]21.18%
Medullab10000.14±0.02 [mm2/s]11.715%

Mean SNR and variability of measurements were significantly different between groups (p<0.0001, ANOVA).

Variability of measurements was significant in ANOVA test between examined specific categories (p<0.0001). Further analysis revealed that mean variability of ADC measurement was significantly lower than of eADC measurement (6.8% vs. 10.8%, respectively; p<0.0001). It was also shown that mean variability of measurements performed in the renal cortex was significantly lower than in the medulla (6.2% vs. 11.5%; p<0.005) – Figure 2.
Figure 2

Comparison of measurement variability at b600 and b1000 (A, p>0.9), between ADC and eADC (B, p<0.0001), and between the cortex and the medulla of the kidney (C, p<0.001). Graphs include mean values and their 95% confidence intervals.

Similarly, mean SNR values in respective categories were significantly different in ANOVA test (p<0.0001). It was shown that mean SNR value was higher in the measurements conducted using b600 than b1000 (33.1±17.4 vs. 23.2±12; p<0.0001), it was higher in ADC maps than in eADC maps (33.5±15.6 vs. 22.8±14.66; p<0.0001), and it was higher in the cortex than in the medulla of the transplanted kidney (35.1±14.7 vs. 21.2±14.1; p<0.0001) – Figure 3.
Figure 3

Comparison of SNR at b600 and b1000 (A, p>0.0001), between ADC and eADC (B, p<0.0001), and between the cortex and the medulla of the kidney (C, p<0.001). Graphs include mean values and their 95% confidence intervals.

In the examined material, ADC and eADC values did not correlate with clinical parameters. On the other hand, significant difference was observed in diffusion parameters measured in the renal cortex using b1000 in the patients divided into 2 groups: with eGFR ≤30 ml/min/1.73 m2 and eGFR >30 ml/min/1.73 m2 (p<0.05) – Table 3.
Table 3

Comparison of ADC and eADC values in the kidney cortex measured at b1000 in patients stratified according to estimated glomerular filtration rate (eGFR). In parentheses, 95% confidence intervals of means are given.

ParametereGFR ≤30 ml/min./1.73 m2eGFR >30 ml/min./1.73 m2p
ADCb1000 [10−3 mm2/s]1.81 (1.71–1.92)2.02 (1.87–2.16)0.0206
eADCb1000 [mm2/s]0.16 (0.14–0.18)0.18 (0.16–0.21)0.0497

Discussion

Diffusion-weighted imaging is a standard method of central nervous system imaging. However, DWI of a transplanted kidney encounters various difficulties due to its location on the border of abdomen and lesser pelvis. First of them is breathing motions. In order to minimize this effect, various methods are used, such as: examination during breath hold, respiratory-triggered or navigator-controlled acquisition [8]; immobilization of patient’s lower abdomen using straps is also recommended. In the early period after transplantation a large amount of fluid can be present is the abdominal cavity, resulting in deterioration of image quality, and cooperation is often difficult due to patient’s serious condition. Patients can also have special drains or dressings giving artifacts in the MR study. In our material as many as 25% of patients were excluded from the final analysis due to inability to complete the study or the presence of non-diagnostic images with significant motion artifacts. It appears that despite the aforementioned difficulties, DWI of a transplanted kidney is a promising technique. In spite of the fact that a large part of published studies are experimental ones and their results are not consistent, it seems that diffusion parameters correlate well with renal function expressed as eGFR and occurrence of acute rejection [2,6,9,10]. The aim of DWI is to distinguish early between acute graft rejection, acute tubular necrosis, delayed graft function after transplantation and to identify causes of long-term renal function impairment [8]. In the first publication on diffusion-weighted imaging in the patients after renal transplantation by Thoeny et al., it was shown that in contradistinction to native kidneys, a transplanted kidney is characterized by a lack of significant ADC difference between the cortex and medulla [9]. The authors explained the lack of corticomedullary differentiation as a result of renal denervation or activity of immunosuppressive agents. In our material a significant difference was observed only in ADC using b1000, but it is worth mentioning that Thoeny et al. measured mean ADC from 10 various b-values, which could have eliminated possible differences for single b. Variability of measurements in a single patient in our material ranged from 4% to 15%, which is consistent with data from the literature [6,8,9]. It suggests that clinical significance should be considered only in differences in ADC and eADC ranged 10–20%. On the other hand, significantly lower variability of measurements conducted in the cortex suggests that further research should focus on this area. There are suggestions in the literature that standard DWI is significantly limited by imaging in only one direction, which can result in a loss of a great amount of data. It is associated with histological renal structure, characterized by marked anisotropy due to radial composition of connecting tubules. Therefore some of the authors suggest that multi-directional diffusion tensor imaging (DTI) would better imitate the nature of diffusion and thereby structural integrity of tubules. Additionally, one of the advantages of DTI is the capacity of measuring fractional anisotropy of the motion of water, which can be of a specific use in the assessment of the renal medulla [3,11]. It appears that parenchymal abnormalities of a transplanted kidney affect significantly renal cortical perfusion and its function, expressed as dynamic processes of concentration and dilution of urine [12]. Hence the attempts of separating DWI diffusion and perfusion fraction, similarly to central nervous system imaging [2,9,13]. It is achieved by sequential scanning of the same region using many b-values (usually 8–10 values ranged from 0 to 800 s/mm2) and two-exponential analysis. Preliminary results suggest good correlation between perfusion measured using ASL method and DWI perfusion fraction, and decrease in the perfusion fraction in the grafts characterized by long cold ischemia time, and in the patients with acute rejection and acute glomerular necrosis [6,13]. Clinical meaning of DWI diffusion fraction (so-called pure diffusion) requires conducting further research [6,8].

Conclusions

Diffusion-weighted imaging of transplanted kidneys, especially in the patients in early period after transplantation, is a technically challenging study. However, its potential use encourages further improvements of acquisition and conducting research in larger populations. From a technical point of view, the best quality is provided by ADC measurement in the renal cortex using b1000. The ADC and eADC measurements in the renal cortex using b1000 are eGFR-dependent.
  13 in total

1.  IVIM-DWI of transplanted kidneys: reduced diffusion and perfusion dependent on cold ischemia time.

Authors:  S Rheinheimer; F Schneider; B Stieltjes; C Morath; M Zeier; H U Kauczor; P Hallscheidt
Journal:  Eur J Radiol       Date:  2012-07-10       Impact factor: 3.528

2.  Functional evaluation of transplanted kidneys with diffusion-weighted and BOLD MR imaging: initial experience.

Authors:  Harriet C Thoeny; Dominik Zumstein; Sonja Simon-Zoula; Ute Eisenberger; Frederik De Keyzer; Lucie Hofmann; Peter Vock; Chris Boesch; Felix J Frey; Peter Vermathen
Journal:  Radiology       Date:  2006-12       Impact factor: 11.105

3.  New motion correction models for automatic identification of renal transplant rejection.

Authors:  Ayman El-Baz; Georgy Gimel'farb; Mohamed A El-Ghar
Journal:  Med Image Comput Comput Assist Interv       Date:  2007

Review 4.  Renal relevant radiology: imaging in kidney transplantation.

Authors:  Asif Sharfuddin
Journal:  Clin J Am Soc Nephrol       Date:  2013-11-07       Impact factor: 8.237

5.  Functional evaluation of transplanted kidneys using arterial spin labeling MRI.

Authors:  Philipp Heusch; Hans-Jörg Wittsack; Dirk Blondin; Alexandra Ljimani; Michael Nguyen-Quang; Petros Martirosian; Hakan Zenginli; Philip Bilk; Patric Kröpil; Till A Heusner; Gerald Antoch; Rotem S Lanzman
Journal:  J Magn Reson Imaging       Date:  2013-10-07       Impact factor: 4.813

6.  Diffusion-weighted MR imaging of native and transplanted kidneys.

Authors:  Harriet C Thoeny; Frederik De Keyzer
Journal:  Radiology       Date:  2011-04       Impact factor: 11.105

7.  Diffusion tensor imaging and tractography for assessment of renal allograft dysfunction-initial results.

Authors:  Katja Hueper; M Gutberlet; T Rodt; W Gwinner; F Lehner; F Wacker; M Galanski; D Hartung
Journal:  Eur Radiol       Date:  2011-06-28       Impact factor: 5.315

8.  Diffusion tensor imaging of renal ischemia reperfusion injury in an experimental model.

Authors:  Jerry S Cheung; Shu Juan Fan; April M Chow; Jingbo Zhang; Kwan Man; Ed X Wu
Journal:  NMR Biomed       Date:  2010-06       Impact factor: 4.044

9.  Evaluation of renal allograft function early after transplantation with diffusion-weighted MR imaging.

Authors:  Ute Eisenberger; Harriet C Thoeny; Tobias Binser; Mathias Gugger; Felix J Frey; Chris Boesch; Peter Vermathen
Journal:  Eur Radiol       Date:  2009-12-16       Impact factor: 5.315

10.  Blood oxygen level-dependent and perfusion magnetic resonance imaging: detecting differences in oxygen bioavailability and blood flow in transplanted kidneys.

Authors:  Elizabeth A Sadowski; Arjang Djamali; Andrew L Wentland; Rebecca Muehrer; Bryan N Becker; Thomas M Grist; Sean B Fain
Journal:  Magn Reson Imaging       Date:  2009-07-03       Impact factor: 2.546

View more
  8 in total

1.  Statistical analysis of ADCs and clinical biomarkers in detecting acute renal transplant rejection.

Authors:  Elizabeth Hollis; Mohamed Shehata; Mohamed Abou El-Ghar; Mohammed Ghazal; Tarek El-Diasty; Michael Merchant; Andrew E Switala; Ayman El-Baz
Journal:  Br J Radiol       Date:  2017-09-13       Impact factor: 3.039

2.  A multimodal computer-aided diagnostic system for precise identification of renal allograft rejection: Preliminary results.

Authors:  Mohamed Shehata; Ahmed Shalaby; Andrew E Switala; Maryam El-Baz; Mohammed Ghazal; Luay Fraiwan; Ashraf Khalil; Mohamed Abou El-Ghar; Mohamed Badawy; Ashraf M Bakr; Amy Dwyer; Adel Elmaghraby; Guruprasad Giridharan; Robert Keynton; Ayman El-Baz
Journal:  Med Phys       Date:  2020-04-03       Impact factor: 4.071

3.  Thoracic splenosis after a gunshot: diffusion-weighted MRI findings.

Authors:  Onur Tutar; Selim Bakan; Cesur Samanci; Fuat Nurili; Haluk Burcak Sayman; Canan Akman
Journal:  Pol J Radiol       Date:  2015-02-20

4.  Usefulness of Apparent Diffusion Coefficient of Diffusion-Weighted Imaging for Differential Diagnosis of Primary Solid and Cystic Renal Masses.

Authors:  Mehmet Öztürk; Afra Ekinci; Şenol Fatih Elbir; Aylin Okur; Serap Doğan; Ökkeş Ibrahim Karahan
Journal:  Pol J Radiol       Date:  2017-04-13

5.  Noninvasive evaluation of renal tissue oxygenation with blood oxygen level-dependent magnetic resonance imaging early after transplantation has a limited predictive value for the delayed graft function.

Authors:  Agata Sławińska; Zbigniew Serafin; Elżbieta Zawada; Marcin Białecki; Katarzyna Wypych; Aleksandra Woderska; Maciej Słupski; Zbigniew Włodarczyk
Journal:  Pol J Radiol       Date:  2018-08-05

6.  A Novel CNN-Based CAD System for Early Assessment of Transplanted Kidney Dysfunction.

Authors:  Hisham Abdeltawab; Mohamed Shehata; Ahmed Shalaby; Fahmi Khalifa; Ali Mahmoud; Mohamed Abou El-Ghar; Amy C Dwyer; Mohammed Ghazal; Hassan Hajjdiab; Robert Keynton; Ayman El-Baz
Journal:  Sci Rep       Date:  2019-04-11       Impact factor: 4.379

7.  Evaluation of Free Breathing Versus Breath Hold Diffusion Weighted Imaging in Terms Apparent Diffusion Coefficient (ADC) and Signal-to-Noise Ratio (SNR) Values for Solid Abdominal Organs.

Authors:  Duygu Herek; Nevzat Karabulut; Ali Kocyıgıt; Ahmet Baki Yagcı
Journal:  Pol J Radiol       Date:  2016-10-25

8.  3D kidney segmentation from abdominal diffusion MRI using an appearance-guided deformable boundary.

Authors:  Mohamed Shehata; Ali Mahmoud; Ahmed Soliman; Fahmi Khalifa; Mohammed Ghazal; Mohamed Abou El-Ghar; Moumen El-Melegy; Ayman El-Baz
Journal:  PLoS One       Date:  2018-07-13       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.