Literature DB >> 24817800

Dose Reduction in Fluoroscopic Interventions Using a Combination of a Region of Interest (ROI) X-Ray Attenuator and Spatially-Different, Temporally-Variable Temporal Filtering.

S N Swetadri Vasan1, Liza Pope2, Ciprian N Ionita2, A H Titus1, D R Bednarek2, S Rudin1.   

Abstract

A novel dose reduction technique for fluoroscopic interventions involving a combination of a material x-ray region of interest (ROI) attenuator and spatially different, temporally variable ROI temporal recursive filter, was used to guide the catheter to the ROI in three live animal studies, two involving rabbits and one involving a sheep. In the two rabbit studies presented, a catheter was guided to the entrance of the carotid artery. With the added ROI attenuator the image under the high attenuation region is very noisy. By using temporal filtering with a filter weight of 0.6 on previous frames, the noise is reduced. In the sheep study the catheter was guided to the descending aorta of the animal. The sheep offered a relatively higher attenuation to the incident x-rays and thus a higher temporal filter weight of 0.8 on previous frames was used during the procedure to reduce the noise to levels acceptable by the interventionalist. The image sequences from both studies show that significant dose reduction of 5-6 times can be achieved with acceptable image quality outside the ROI by using the above mentioned technique. Even though the temporal filter weighting outside the ROI is higher, the consequent lag does not prevent perception of catheter movement.

Entities:  

Year:  2013        PMID: 24817800      PMCID: PMC4013118          DOI: 10.1117/12.2006277

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  4 in total

1.  Optimization of variable temporal averaging in digital fluoroscopy.

Authors:  C J Kotre; E Guibelalde
Journal:  Br J Radiol       Date:  2004-08       Impact factor: 3.039

2.  Dose Reduction Technique Using a Combination of a Region of Interest (ROI) Material X-Ray Attenuator and Spatially Different Temporal Filtering for Fluoroscopic Interventions.

Authors:  S N Swetadri Vasan; A Panse; A Jain; P Sharma; Ciprian N Ionita; A H Titus; A N Cartwright; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2012-02-23

3.  Automatic filter placement device for region of interest (ROI) fluoroscopy.

Authors:  M Kezerashvili; S Rudin; D Bednarek
Journal:  Health Phys       Date:  1997-01       Impact factor: 1.316

4.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

  4 in total
  5 in total

1.  Organ and effective dose reduction for region-of-interest (ROI) CBCT and fluoroscopy.

Authors:  Zhenyu Xiong; Sarath Vijayan; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2017-03-09

2.  A Combination of Spatial and Recursive Temporal Filtering for Noise Reduction when Using Region of Interest (ROI) Fluoroscopy for Patient Dose Reduction in Image Guided Vascular Interventions with Significant Anatomical Motion.

Authors:  S V Setlur Nagesh; P Khobragade; C Ionita; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

3.  Assessment of organ and effective dose when using region-of-interest attenuators in cone-beam CT and interventional fluoroscopy.

Authors:  Zhenyu Xiong; Sarath Vijayan; Stephen Rudin; Daniel R Bednarek
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-22

4.  A Patient Dose-Reduction Technique for Neuroendovascular Image-Guided Interventions: Image-Quality Comparison Study.

Authors:  A Sonig; S V Setlur Nagesh; V S Fennell; S Gandhi; L Rangel-Castilla; C N Ionita; K V Snyder; L N Hopkins; D R Bednarek; S Rudin; A H Siddiqui; E I Levy
Journal:  AJNR Am J Neuroradiol       Date:  2018-02-15       Impact factor: 3.825

5.  A novel Region of Interest (ROI) imaging technique for biplane imaging in interventional suites: high-resolution small field-of-view imaging in the frontal plane and dose-reduced, large field-of-view standard-resolution imaging in the lateral plane.

Authors:  S N Swetadri Vasan; C Ionita; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2014-03-19
  5 in total