Literature DB >> 28638169

Organ and effective dose reduction for region-of-interest (ROI) CBCT and fluoroscopy.

Zhenyu Xiong1,2, Sarath Vijayan1,2, Stephen Rudin1,3,2, Daniel R Bednarek1,3,2.   

Abstract

In some medical-imaging procedures using CBCT and fluoroscopy, it may be needed to visualize only the center of the field-of-view with optimal quality. To reduce the dose to the patient as well as enable increased contrast in the region of interest (ROI) during CBCT and fluoroscopy procedures, a 0.7 mm thick Cu ROI attenuator with a circular aperture 12% of the FOV was used. The aim of this study was to quantify the dose-reduction benefit of ROI imaging during a typical CBCT and interventional fluoroscopy procedures in the head and torso. The Toshiba Infinix C-Arm System was modeled in BEAMnrc/EGSnrc with and without the ROI attenuator. Patient organ and effective doses were calculated in DOSXYZnrc/EGSnrc Monte-Carlo software for CBCT and interventional procedures. We first compared the entrance dose with and without the ROI attenuator on a 20 cm thick solid-water block. Then we simulated a CBCT scan and an interventional fluoroscopy procedure on the head and torso with and without an ROI attenuator. The results showed that the entrance-surface dose reduction in the solid water is about 85.7% outside the ROI opening and 10.5% in the ROI opening. The results showed a reduction in most organ doses of 45%-70% and in effective dose of 46%-66% compared to the dose in a CBCT scan and in an interventional procedure without the ROI attenuator. This work provides evidence of substantial reduction of organ and effective doses when using an ROI attenuator during CBCT and fluoroscopic procedures.

Entities:  

Keywords:  CBCT; EGSnrc; Monte Carlo; ROI imaging; dose reduction; interventional fluoroscopy procedure

Year:  2017        PMID: 28638169      PMCID: PMC5476203          DOI: 10.1117/12.2254061

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  15 in total

1.  Accurate assessment of patient effective radiation dose and associated detriment risk from radiofrequency catheter ablation procedures.

Authors:  K Perisinakis; J Damilakis; N Theocharopoulos; E Manios; P Vardas; N Gourtsoyiannis
Journal:  Circulation       Date:  2001-07-03       Impact factor: 29.690

2.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

3.  Clinical considerations for allied professionals: radiation safety and protection in the electrophysiology lab.

Authors:  Esther M Weiss; Omar Thabit
Journal:  Heart Rhythm       Date:  2007-10-04       Impact factor: 6.343

4.  Computerized three-dimensional segmented human anatomy.

Authors:  I G Zubal; C R Harrell; E O Smith; Z Rattner; G Gindi; P B Hoffer
Journal:  Med Phys       Date:  1994-02       Impact factor: 4.071

5.  Dose Reduction in Fluoroscopic Interventions Using a Combination of a Region of Interest (ROI) X-Ray Attenuator and Spatially-Different, Temporally-Variable Temporal Filtering.

Authors:  S N Swetadri Vasan; Liza Pope; Ciprian N Ionita; A H Titus; D R Bednarek; S Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-06

6.  Space radiation and cataracts in astronauts.

Authors:  F A Cucinotta; F K Manuel; J Jones; G Iszard; J Murrey; B Djojonegro; M Wear
Journal:  Radiat Res       Date:  2001-11       Impact factor: 2.841

7.  Risk of cataract after exposure to low doses of ionizing radiation: a 20-year prospective cohort study among US radiologic technologists.

Authors:  Gabriel Chodick; Nural Bekiroglu; Michael Hauptmann; Bruce H Alexander; D Michal Freedman; Michele Morin Doody; Li C Cheung; Steven L Simon; Robert M Weinstock; André Bouville; Alice J Sigurdson
Journal:  Am J Epidemiol       Date:  2008-07-29       Impact factor: 4.897

8.  The European Directive on health protection of individuals against the dangers of ionising radiation in relation to medical exposures (97/43/EURATOM).

Authors:  D Teunen
Journal:  J Radiol Prot       Date:  1998-06       Impact factor: 1.394

9.  Incorporating Corrections for the Head-Holder and Compensation Filter when Calculating Skin Dose during Fluoroscopically-Guided Interventions.

Authors:  Sarath Vijayan; Vijay K Rana; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

10.  Application of region-of-interest imaging techniques to neurointerventional radiology.

Authors:  S Rudin; L R Guterman; W E Granger; D R Bednarek; L N Hopkins
Journal:  Radiology       Date:  1996-06       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.