Evaluation of low-cost telemammography screening configurations: a comparison with film-screen readings in vulnerable areas.

Antonio J Salazar1, Javier Romero, Oscar Bernal, Angela Moreno, Sofía Velasco, Xavier Díaz.   

Abstract

The aim of this study was to evaluate the diagnostic accuracy for detecting breast cancer using different telemammography configurations, including combinations of both low-cost capture devices and consumer-grade color displays. At the same time, we compared each of these configurations to film-screen readings. This study used a treatment-by-reader-by-case factorial design. The sample included 70 mammograms with 34 malignant cases. The readers consisted of four radiologists who classified all of the cases according to the categories defined by the Breast Imaging Reporting and Data System (BI-RADS). The evaluated capture devices included a specialized film digitizer and a digital camera, and the evaluated displays included liquid crystal display (LCD) and light-emitting diode (LED) consumer-grade color displays. Receiver operating characteristic curves, diagnostic accuracy (measured as the area under these curves), accuracy of the composition classification, sensitivity, specificity, and the degree of agreement between readers in the detection of malignant cases were also evaluated. Comparisons of diagnostic accuracy between film-screen and the different combinations of digital configurations showed no significant differences for nodules, calcifications, and asymmetries. In addition, no differences were observed in terms of sensibility or specificity when the degree of malignancy using the film-screen method was compared to that provided with digital configurations. Similar results were observed for the classification of breast composition. Furthermore, all observed reader agreements of malignant detection between film-screen and digital configurations were substantial. These findings indicate that the evaluated digital devices showed comparable diagnostic accuracy to the reference treatment (film-screen).

Entities:  

Mesh:

Year:  2014        PMID: 24802372      PMCID: PMC4171425          DOI: 10.1007/s10278-014-9695-y

Source DB:  PubMed          Journal:  J Digit Imaging        ISSN: 0897-1889            Impact factor:   4.056


  28 in total

1.  Interpretation of digital mammograms: comparison of speed and accuracy of soft-copy versus printed-film display.

Authors:  Etta D Pisano; Elodia B Cole; Emily O Kistner; Keith E Muller; Bradley M Hemminger; Mary L Brown; R Eugene Johnston; Cherie M Kuzmiak; M Patricia Braeuning; Rita I Freimanis; Mary Scott Soo; J A Baker; Ruth Walsh
Journal:  Radiology       Date:  2002-05       Impact factor: 11.105

2.  Sample size tables for receiver operating characteristic studies.

Authors:  N A Obuchowski
Journal:  AJR Am J Roentgenol       Date:  2000-09       Impact factor: 3.959

3.  Agreement and reading time for differently-priced devices for the digital capture of X-ray films.

Authors:  Antonio José Salazar; Juan Camilo Camacho; Diego Andrés Aguirre
Journal:  J Telemed Telecare       Date:  2011-12-14       Impact factor: 6.184

4.  Introduction to grayscale calibration and related aspects of medical imaging grade liquid crystal displays.

Authors:  Kenneth A Fetterly; Hartwig R Blume; Michael J Flynn; Ehsan Samei
Journal:  J Digit Imaging       Date:  2008-06       Impact factor: 4.056

5.  Evaluation of three pneumothorax size quantification methods on digitized chest X-ray films using medical-grade grayscale and consumer-grade color displays.

Authors:  Antonio J Salazar; Diego A Aguirre; Juliana Ocampo; Juan C Camacho; Xavier A Díaz
Journal:  J Digit Imaging       Date:  2014-04       Impact factor: 4.056

6.  Comparison between differently priced devices for digital capture of X-ray films using computed tomography as a gold standard: a multireader-multicase receiver operating characteristic curve study.

Authors:  Antonio J Salazar; Juan Camilo Camacho; Diego Andrés Aguirre
Journal:  Telemed J E Health       Date:  2011-04-01       Impact factor: 3.536

7.  Diagnostic accuracy of digitized chest X-rays using consumer-grade color displays for low-cost teleradiology services: a multireader-multicase comparison.

Authors:  Antonio J Salazar; Diego A Aguirre; Juliana Ocampo; Xavier A Diaz; Juan C Camacho
Journal:  Telemed J E Health       Date:  2014-02-07       Impact factor: 3.536

8.  DICOM gray-scale standard display function: clinical diagnostic accuracy of chest radiography in medical-grade gray-scale and consumer-grade color displays.

Authors:  Antonio J Salazar; Diego A Aguirre; Juliana Ocampo; Juan C Camacho; Xavier A Díaz
Journal:  AJR Am J Roentgenol       Date:  2014-06       Impact factor: 3.959

9.  Breast cancer screening: a summary of the evidence for the U.S. Preventive Services Task Force.

Authors:  Linda L Humphrey; Mark Helfand; Benjamin K S Chan; Steven H Woolf
Journal:  Ann Intern Med       Date:  2002-09-03       Impact factor: 25.391

10.  Clinical comparison of full-field digital mammography and screen-film mammography for detection of breast cancer.

Authors:  John M Lewin; Carl J D'Orsi; R Edward Hendrick; Lawrence J Moss; Pamela K Isaacs; Andrew Karellas; Gary R Cutter
Journal:  AJR Am J Roentgenol       Date:  2002-09       Impact factor: 3.959

View more
  2 in total

1.  Optimization of Image Quality and Dose in Digital Mammography.

Authors:  Agnes M F Fausto; M C Lopes; M C de Sousa; Tânia A C Furquim; Anderson W Mol; Fermin G Velasco
Journal:  J Digit Imaging       Date:  2017-04       Impact factor: 4.056

2.  Noninferiority and Equivalence Evaluation of Clinical Performance among Computed Radiography, Film, and Digitized Film for Telemammography Services.

Authors:  Antonio J Salazar; Javier A Romero; Oscar A Bernal; Angela P Moreno; Sofía C Velasco; Xavier A Díaz
Journal:  Int J Telemed Appl       Date:  2016-09-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.