DICOM gray-scale standard display function: clinical diagnostic accuracy of chest radiography in medical-grade gray-scale and consumer-grade color displays.

Antonio J Salazar1, Diego A Aguirre, Juliana Ocampo, Juan C Camacho, Xavier A Díaz.   

Abstract

OBJECTIVE: The purpose of this study was to compare the diagnostic accuracy achieved with and without the calibration method established by the DICOM standard in both medical-grade gray-scale displays and consumer-grade color displays.
MATERIALS AND METHODS: This study involved 76 cases, six radiologists, three displays, and two display calibrations for a total of 2736 observations in a multireader-multicase factorial design. The evaluated conditions were interstitial opacities, pneumothorax, and nodules. CT was adopted as the reference standard. One medical-grade gray-scale display and two consumer-grade color displays were evaluated. Analyses of ROC curves, diagnostic accuracy (measured as AUC), accuracy of condition classification, and false-positive and false-negative rate comparisons were performed. The degree of agreement between readers was also evaluated.
RESULTS: No significant differences in image quality perception by the readers in the presence or absence of calibration were observed. Similar forms of the ROC curves were observed. No significant differences were detected in the observed variables (diagnostic accuracy, accuracy of condition classification, false-positive rates, false-negative rates, and image-quality perception). Strong agreement between readers was also determined for each display with and without calibration.
CONCLUSION: For the chest conditions and selected observers included in this study, no significant differences were observed between the three evaluated displays with respect to accuracy performance with and without calibration.

Mesh:

Year:  2014        PMID: 24848825     DOI: 10.2214/AJR.13.11509

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  6 in total

1.  Comparison of medical-grade and calibrated consumer-grade displays for diagnosis of subtle bone fissures.

Authors:  Daniel Pinto Dos Santos; Jonas Welter; Tilman Emrich; Florian Jungmann; Evelyn Dappa; Peter Mildenberger; Roman Kloeckner
Journal:  Eur Radiol       Date:  2017-06-28       Impact factor: 5.315

2.  Evaluation of low-cost telemammography screening configurations: a comparison with film-screen readings in vulnerable areas.

Authors:  Antonio J Salazar; Javier Romero; Oscar Bernal; Angela Moreno; Sofía Velasco; Xavier Díaz
Journal:  J Digit Imaging       Date:  2014-10       Impact factor: 4.056

3.  TSR guidelines for the practice of teleradiology: 2021 update.

Authors:  Mustafa N Özmen; Oğuz Dicle; Utku Şenol; Üstün Aydıngöz
Journal:  Diagn Interv Radiol       Date:  2021-07       Impact factor: 2.630

4.  Effect of display type and room illuminance in chest radiographs.

Authors:  Esa Liukkonen; Airi Jartti; Marianne Haapea; Heljä Oikarinen; Lauri Ahvenjärvi; Seija Mattila; Terhi Nevala; Kari Palosaari; Marja Perhomaa; Miika T Nieminen
Journal:  Eur Radiol       Date:  2015-12-10       Impact factor: 5.315

5.  Noninferiority and Equivalence Evaluation of Clinical Performance among Computed Radiography, Film, and Digitized Film for Telemammography Services.

Authors:  Antonio J Salazar; Javier A Romero; Oscar A Bernal; Angela P Moreno; Sofía C Velasco; Xavier A Díaz
Journal:  Int J Telemed Appl       Date:  2016-09-29

6.  Reproducible Color Gamut of Hematoxylin and Eosin Stained Images in Standard Color Spaces.

Authors:  Wei-Chung Cheng
Journal:  J Pathol Inform       Date:  2020-11-06
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.