Claire Dempsey1, Jameen Arm2, Leah Best2, Geetha Govindarajulu3, Anne Capp3, Peter O'Brien3.
Abstract
PURPOSE: The superior image quality of 3 tesla (3T) magnetic resonance (MR) imaging in cervical cancer offers the potential to use a single image set for brachytherapy. This study aimed to determine a suitable single sequence for contouring tumour and organs at risk, applicator reconstruction, and treatment planning.Entities:
Keywords: 3 Tesla; cervical cancer; cervix cancer; magnetic resonance imaging; treatment planning
Year: 2014 PMID: 24790615 PMCID: PMC4003427 DOI: 10.5114/jcb.2014.41528
Source DB: PubMed Journal: J Contemp Brachytherapy ISSN: 2081-2841
Fig. 1Axial-oblique plane, aligned parallel to the plane of the ring in the sagittal scan
Fig. 2Axial double-oblique additional plane, aligned parallel to the plane of the ring in the coronal localizer scan
Magnetic resonance scanning parameters used for image acquisition
| MR sequence | TR (ms) | TE (ms) | FOV (mm) | Slice/Gap(mm) | Matrix (F/P) (mm) | Averages (NEX) | Flip angle (degrees) | Scan Time (mins:sec) | ||
|---|---|---|---|---|---|---|---|---|---|---|
| 2D | Sagittal | TSE T2 | 5000-6000 | 90 | 240 × 240 | 2.5/0.25 | 320 × 320 | 1 | 160 | 2:00 |
| Axial | TSE T2 | 4000-6000 | 85 | 200 × 200 | 1.5/0 | 320 × 256 | 1 | 143 | 6:10 | |
| Oblique TSE T2 | 5000-6000 | 85 | 210 × 210 | 1.8/0.09 | 320 × 256 | 1 | 143 | 5:33 | ||
| Double oblique TSE T2 | 5000-6000 | 84 | 210 × 210 | 1.8/0.09 | 256 × 256 | 1 | 143 | 4:44 | ||
| 3D | Sagittal | TSE SPACE T2 | 1700 | 80 | 240 × 240 | 1/0 | 320 × 288 | 1 | 160 | 8:11 |
| Axial | TSE SPACE T2 | 1700 | 83 | 230 × 230 | 1/0 | 320 × 288 | 1 | 160 | 6:11 | |
| Oblique TSE SPACE T2 | 1400 | 83 | 230 × 230 | 1/0 | 256 × 256 | 1 | 160 | 7:09 |
TSE – turbo spin echo, SPACE – sampling perfection with application-optimised contrast and different flip-angle evolutions, TR – relaxation time, TE – echo time, FOV – field of view, NEX – number of excitations
Intra- and inter-observer differences for different 3T T2 MR scanning sequences
| 2D | Sagittal | TSE T2 | 35.7 | 9.3 | 14.5 | 19.8 |
| Axial | TSE T2 | 6.4 | 0.3 | 11.7 | 6.1 | |
| Oblique TSE T2 | 5.4 | 1.2 | 6.1 | 4.2 | ||
| Double oblique TSE T2 | 1.2 | 0.3 | 0.3 | 0.6 | ||
| 3D | Sagittal | TSE SPACE T2 | 22.8 | 9.5 | 12.9 | 15.1 |
| Axial | TSE SPACE T2 | 1.0 | 1.7 | 0.1 | 0.9 | |
| Oblique TSE SPACE T2 | 6.5 | 1.0 | 4.1 | 3.9 | ||
| 2D | Sagittal | TSE T2 | 37.0 | 28.7-48.9 | 0.59 | 0.40-0.68 |
| Axial | TSE T2 | 32.5 | 17.7-47.3 | 0.55 | 0.45-0.75 | |
| Oblique TSE T2 | 30.6 | 14.1-68.3 | 0.63 | 0.58-0.74 | ||
| Double oblique TSE T2 | 9.2 | 2.9-12.4 | 0.80 | 0.72-0.91 | ||
| 3D | Sagittal | TSE SPACE T2 | 19.8 | 9.8-31.3 | 0.71 | 0.60-0.79 |
| Axial | TSE SPACE T2 | 28.1 | 16.6-46.8 | 0.60 | 0.61-0.72 | |
| Oblique TSE SPACE T2 | 12.3 | 4.1-13.5 | 0.74 | 0.70-0.78 | ||