A novel removable shield attached to C-arm units against scattered X-rays from a patient's side.

Hiroshige Mori1, Kichiro Koshida, Osamu Ishigamori, Kosuke Matsubara.   

Abstract

OBJECTIVES: We invented a drape-like shield against scattered X-rays that can safely come into contact with medical equipment or people during fluoroscopically guided procedures. The shield can be easily removed from a C-arm unit using one hand. We evaluated the use of the novel removable shield during the endoscopic retrograde cholangiopancreatography (ERCP) procedure.
METHODS: We measured the dose rate of scattered X-rays around endoscopists with and without this removable shield and surveyed the occupational doses to the ERCP staff. We also examined the endurance of the shield.
RESULTS: The removable shield reduced the dose rate of scattered X-rays to one-tenth and reduced the monthly dose to an endoscopist by at least two-fifths. For 2.5 years, there was no damage to the shield and no loosening of the seam. The bonding of the hook-and-loop fasteners did not weaken, although the powerful double-sided tapes made especially for plastic did.
CONCLUSIONS: The removable shield can reduce radiation exposure to the ERCP staff and may contribute to reducing the exposure to the eye lenses of operators. It would also be possible to expand its use to other fluoroscopically guided procedures besides ERCP because it is a light, simple, and useful device. KEY POINTS: • We invented a shield that can be removed from C-arm units with one hand. • The removable shield reduces the dose rate of X-rays to one-tenth. • The removable shield reduces operator exposure by two-fifths. • The removable shield is durable, lasting for several years. • The drape-like removable shield is light, simple, and useful.

Entities:  

Mesh:

Year:  2014        PMID: 24788230     DOI: 10.1007/s00330-014-3186-x

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  15 in total

1.  Using a sterile disposable protective surgical drape for reduction of radiation exposure to interventionalists.

Authors:  Jerry N King; Anna M Champlin; Charles A Kelsey; David A Tripp
Journal:  AJR Am J Roentgenol       Date:  2002-01       Impact factor: 3.959

2.  Staff radiation doses to the lower extremities in interventional radiology.

Authors:  C P Shortt; H Al-Hashimi; L Malone; M J Lee
Journal:  Cardiovasc Intervent Radiol       Date:  2007-05-29       Impact factor: 2.740

3.  Radiation exposure during ERCP: effect of a protective shield.

Authors:  M Y Chen; F L Van Swearingen; R Mitchell; D J Ott
Journal:  Gastrointest Endosc       Date:  1996-01       Impact factor: 9.427

Review 4.  ICRP publication 118: ICRP statement on tissue reactions and early and late effects of radiation in normal tissues and organs--threshold doses for tissue reactions in a radiation protection context.

Authors:  F A Stewart; A V Akleyev; M Hauer-Jensen; J H Hendry; N J Kleiman; T J Macvittie; B M Aleman; A B Edgar; K Mabuchi; C R Muirhead; R E Shore; W H Wallace
Journal:  Ann ICRP       Date:  2012-02

5.  Patient and staff exposure during endoscopic retrograde cholangiopancreatography.

Authors:  N Buls; J Pages; F Mana; M Osteaux
Journal:  Br J Radiol       Date:  2002-05       Impact factor: 3.039

6.  Radiation exposure to personnel performing endoscopic retrograde cholangiopancreatography.

Authors:  L S Naidu; S Singhal; D E Preece; A Vohrah; D E Loft
Journal:  Postgrad Med J       Date:  2005-10       Impact factor: 2.401

7.  Patient and endoscopist radiation doses during ERCP procedures.

Authors:  V Tsapaki; K D Paraskeva; N Mathou; E Andrikopoulos; P Tentas; C Triantopoulou; J A Karagiannis
Journal:  Radiat Prot Dosimetry       Date:  2011-07-09       Impact factor: 0.972

8.  [Efficacy of a self-designed protective lead shield in reduction of radiation exposure dose during endoscopic retrograde cholangiopancreatography].

Authors:  Yun Jung Kim; Kwang Bum Cho; Eun Soo Kim; Kyung Sik Park; Byoung Kuk Jang; Woo Jin Chung; Jae Seok Hwang
Journal:  Korean J Gastroenterol       Date:  2011-01

9.  Heavy metal pad shielding during fluoroscopic interventions.

Authors:  Sergio Dromi; Bradford J Wood; Jay Oberoi; Ziv Neeman
Journal:  J Vasc Interv Radiol       Date:  2006-07       Impact factor: 3.464

10.  Radiation effects analysis in a group of interventional radiologists using biological and physical dosimetry methods.

Authors:  M Ramos; A Montoro; M Almonacid; S Ferrer; J F Barquinero; R Tortosa; G Verdú; P Rodríguez; L L Barrios; J I Villaescusa
Journal:  Eur J Radiol       Date:  2009-04-19       Impact factor: 3.528

View more
  3 in total

1.  Real-time, ray casting-based scatter dose estimation for c-arm x-ray system.

Authors:  Zaid Alnewaini; Eric Langer; Philipp Schaber; Matthias David; Dominik Kretz; Volker Steil; Jürgen Hesser
Journal:  J Appl Clin Med Phys       Date:  2017-01-24       Impact factor: 2.102

2.  Radiation protection effect of mobile shield barrier for the medical personnel during endoscopic retrograde cholangiopancreatography: a quasi-experimental prospective study.

Authors:  Kwang Hyun Chung; Young Sook Park; Sang Bong Ahn; Byoung Kwan Son
Journal:  BMJ Open       Date:  2019-03-20       Impact factor: 2.692

3.  A multicenter study of radiation doses to the eye lenses of clinical physicians performing radiology procedures in Japan.

Authors:  Keisuke Nagamoto; Takashi Moritake; Koichi Nakagami; Koichi Morota; Satoru Matsuzaki; Naoki Kunugita
Journal:  J Occup Health       Date:  2021-01       Impact factor: 2.708

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.