| Literature DB >> 24710453 |
Pascal Hauri1, Sarah Verlaan, Shaun Graydon, Linda Ahnen, Stephan Klöck, Stephanie Lang.
Abstract
The Delta(4DVH) Anatomy 3D quality assurance (QA) system (ScandiDos), which converts the measured detector dose into the dose distribution in the patient geometry was evaluated. It allows a direct comparison of the calculated 3D dose with the measured back-projected dose. In total, 16 static and 16 volumetric-modulated arc therapy (VMAT) fields were planned using four different energies. Isocenter dose was measured with a pinpoint chamber in homogeneous phantoms to investigate the dose prediction by the Delta(4DVH) Anatomy algorithm for static fields. Dose distributions of VMAT fields were measured using GAFCHROMIC film. Gravitational gantry errors up to 10° were introduced into all VMAT plans to study the potential of detecting errors. Additionally, 20 clinical treatment plans were verified. For static fields, the Delta(4DVH) Anatomy predicted the isocenter dose accurately, with a deviation to the measured phantom dose of 1.1% ± 0.6%. For VMAT fields the predicted Delta(4DVH) Anatomy dose in the isocenter plane corresponded to the measured dose in the phantom, with an average gamma agreement index (GAI) (3 mm/3%) of 96.9± 0.4%. The Delta(4DVH) Anatomy detected the induced systematic gantry error of 10° with a relative GAI (3 mm/3%) change of 5.8% ± 1.6%. The conventional Delta(4PT) QA system detected a GAI change of 4.2%± 2.0%. The conventional Delta(4PT) GAI (3 mm/3%) was 99.8% ± 0.4% for the clinical treatment plans. The mean body and PTV-GAI (3 mm/5%) for the Delta(4DVH) Anatomy were 96.4% ± 2.0% and 97.7%± 1.8%; however, this dropped to 90.8%± 3.4% and 87.1% ± 4.1% for passing criteria of 3 mm/3%. The anatomy-based patient specific quality assurance system predicts the dose distribution correctly for a homogeneous case. The limiting factor for the error detection is the large variability in the error-free plans. The dose calculation algorithm is inferior to that used in the TPS (Eclipse).Entities:
Mesh:
Year: 2014 PMID: 24710453 PMCID: PMC5875461 DOI: 10.1120/jacmp.v15i2.4647
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1DVHs of the PTV and OAR for a prostate cancer patient. The thin lines are the TPS calculated and the large round points are the measured reprojected Delta4DVH Anatomy DVHs.
Figure 2Evaluation of the Anatomy option using two phantoms as patient surrogates. A plan is calculated on a phantom and a verification plan is done on the Delta4 phantom. The plan is irradiated on the Delta4 phantom and on the patient surrogate phantom with a GAFCHROMIC film inserted. From the Delta4 measurements, the dose is back‐projected to the patient anatomy. The different calculated and measured dose distributions are compared (indicated with yellow arrows in the figure).
Comparison of the mean dose measured in the isocenter using a pinpoint chamber with the dose calculated by the Anatomy option and by the Eclipse treatment planning system (TPS)
|
|
|
|
| |
|---|---|---|---|---|
| Anatomy vs. Pinpoint Chamber |
|
|
|
|
| TPS vs. Pinpoint Chamber |
|
|
|
|
| Anatomy vs. TPS |
|
|
|
|
Figure 3GAI for the four comparisons as a mean value for both phantoms and the four energies used. Error bars are the standard deviation.
Figure 4GAI for the four comparisons, as an average of both phantoms and the four energies. The red × = the Conventional‐GAI, the black circles = the TPS Phantom GAI, the green stars = the Anatomy Phantom GAI, and the pink crosses = the Anatomy GAI for the different DTA and DD. From the center to the right, the big red line is a relaxation of the DD and to the left, the red line is an relaxation of the DTA. Error bars are the standard deviation.
Mean three‐dimensional gamma agreement index (GAI) and standard deviation inside the PTV and the body of the patient using Delat4 Anatomy option. Mean GAI in the horizontal isocenter slice and mean dose deviation in the isocenter and mean GAI in the phantom (Conventional GAI)
|
|
| |
|---|---|---|
| Body GAI |
|
|
| PTV GAI |
|
|
| Body GAI |
|
|
| PTV GAI |
|
|
| Horizontal slice |
|
|
| Dose deviation (%) |
|
|
| Conventional GAI |
|
|
Figure 5The relative GAI change for a 2° and 10° gantry angle error. Error bars are the standard deviation.