Evaluation of the Delta4 phantom for IMRT and VMAT verification.

James L Bedford1, Young K Lee, Philip Wai, Christopher P South, Alan P Warrington.   

Abstract

The Delta(4) diode array phantom (Scandidos, Uppsala, Sweden) was evaluated for verification of segmental intensity-modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) on an Elekta linear accelerator (Crawley UK). The device was tested for angular sensitivity by irradiating it from 36 different gantry angles, and the responses of the device to various step-and-shoot segment doses and dose rates were evaluated using an ionization chamber as a comparison. The phantom was then compared with ionization chamber and film results for two prostate and pelvic nodes IMRT plans, two head and neck IMRT plans and two lung VMAT plans. These plans were calculated using Pinnacle(3) (Philips Radiation Oncology Systems, Madison, WI). The uniformity of angular response was better than 0.5% over the range of gantry angles. The uniformity of response of the Delta(4) to different segment monitor units and dose rates was better than 0.5%. The assessment of the IMRT and VMAT plans showed that the Delta(4) measured a dose within 2.5% of the ionization chamber, and compared to film recorded a slightly larger region (range -2% to +7%) agreeing with the planned dose to within 3% and 3 mm. The Delta(4) is a complex device and requires careful benchmarking, but following the successful completion of these measurements, the Delta(4) has been introduced into clinical use.

Entities:  

Mesh:

Year:  2009        PMID: 19384007     DOI: 10.1088/0031-9155/54/9/N04

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  64 in total

1.  Dose verification for respiratory-gated volumetric modulated arc therapy.

Authors:  Jianguo Qian; Lei Xing; Wu Liu; Gary Luxton
Journal:  Phys Med Biol       Date:  2011-07-13       Impact factor: 3.609

2.  A comprehensive investigation of the accuracy and reproducibility of a multitarget single isocenter VMAT radiosurgery technique.

Authors:  Andrew Thomas; Michael Niebanck; Titania Juang; Zhiheng Wang; Mark Oldham
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

3.  A study on the correlation between plan complexity and gamma index analysis in patient specific quality assurance of volumetric modulated arc therapy.

Authors:  Dhanabalan Rajasekaran; Prakash Jeevanandam; Prabakar Sukumar; Arulpandiyan Ranganathan; Samdevakumar Johnjothi; Vivekanandan Nagarajan
Journal:  Rep Pract Oncol Radiother       Date:  2014-09-06

4.  Independent calculation of monitor units for VMAT and SPORT.

Authors:  Xin Chen; Karl Bush; Aiping Ding; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

5.  Comparison of deliverable IMRT and VMAT for spine metastases using a simultaneous integrated boost.

Authors:  Y K Lee; J L Bedford; H A McNair; M A Hawkins
Journal:  Br J Radiol       Date:  2013-02       Impact factor: 3.039

6.  A dosimetric analysis of volumetric-modulated arc radiotherapy with jaw width restriction vs 7 field intensity-modulated radiotherapy for definitive treatment of cervical cancer.

Authors:  B Huang; Z Fang; Y Huang; P Lin; Z Chen
Journal:  Br J Radiol       Date:  2014-05-16       Impact factor: 3.039

7.  Dosimetry challenges for implementing emerging technologies.

Authors:  Fang-Fang Yin; Mark Oldham; Jing Cai; Qiuwen Wu
Journal:  J Phys Conf Ser       Date:  2010

8.  Paraspinal volumetric modulated arc therapy.

Authors:  J L Bedford; H M Convery; V N Hansen; F H Saran
Journal:  Br J Radiol       Date:  2012-01-03       Impact factor: 3.039

9.  Analyzing the performance of ArcCHECK diode array detector for VMAT plan.

Authors:  Rajesh Thiyagarajan; Arunai Nambiraj; Sujit Nath Sinha; Girigesh Yadav; Ashok Kumar; Vikraman Subramani
Journal:  Rep Pract Oncol Radiother       Date:  2015-12-02

10.  Advantage of 3D volumetric dosemeter in delivery quality assurance of dynamic arc therapy: comparison of pencil beam and Monte Carlo calculations.

Authors:  H-J Shin; J H Song; J-Y Jung; Y-K Kwak; C S Kay; Y-N Kang; B O Choi; H S Jang; S H Son
Journal:  Br J Radiol       Date:  2013-11-14       Impact factor: 3.039

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.