| Literature DB >> 28691256 |
Masashi Tomida1,2, Takeshi Kamomae3, Junji Suzuki4, Yoichi Ohashi1, Yoshiyuki Itoh3, Hiroshi Oguchi2, Takahito Okuda1.
Abstract
Stereotactic body radiation therapy (SBRT) using recently introduced multileaf collimators (MLC) is preferred over circular collimators in the treatment of localized prostate cancer. The objective of this study was to assess the clinical usefulness of MLCs in prostate SBRT by comparing the effectiveness of treatment plans using fixed collimators, variable collimators, and MLCs and by ensuring delivery quality assurance (DQA) for each. For each patient who underwent conventional radiation therapy for localized prostate cancer, mock SBRT plans were created using a fixed collimator, a variable collimator, and an MLC. The total MUs, treatment times, and dose-volume histograms of the planning target volumes and organs at risk for each treatment plan were compared. For DQA, a phantom with a radiochromic film or an ionization chamber was irradiated in each plan. We performed gamma-index analysis to evaluate the consistency between the measured and calculated doses. The MLC-based plans had an ~27% lower average total MU than the plans involving other collimators. Moreover, the average estimated treatment time for the MLC plan was 31% and 20% shorter than that for the fixed and variable collimator plans respectively. The gamma-index passing rate in the DQA using film measurements was slightly lower for the MLC than for the other collimators. The DQA results acquired using the ionization chamber showed that the discrepancies between the measured and calculated doses were within 3% in all cases. The results reinforce the usefulness of MLCs in robotic radiosurgery for prostrate SBRT treatment planning; most notably, the total MU and treatment time were both reduced compared to the cases using other types of collimators. Moreover, although the DQA results based on film dosimetry yielded a slightly lower gamma-index passing rate for the MLC than for the other collimators, the MLC accuracy was determined to be sufficient for clinical use.Entities:
Keywords: multileaf collimator; patient-specific quality assurance; prostate cancer; robotic radiosurgery system; stereotactic body radiation therapy
Mesh:
Year: 2017 PMID: 28691256 PMCID: PMC5875821 DOI: 10.1002/acm2.12128
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Images of (a) fixed collimator, (b) variable collimator, and (c) MLC.
Figure 2Geometrical arrangement of radiation fields for penumbra width comparison with MLC. The penumbrae were oriented and located as follows: (a) along the X‐axis and centered, (b) along the X‐axis and shifted by 30 mm in the +X direction, (c) along the X‐axis and shifted by 30 mm in the −X direction, (d) along the Y‐axis and centered, (e) along the Y‐axis and shifted by 30 mm in the +Y direction, and (f) along the Y‐axis and shifted by 30 mm in the −Y direction.
Average discrepancy between measured and calculated doses in single beam
| Depth (mm) | Field size | Fixed collimator (%) | Variable collimator (%) | MLC (%) | |
|---|---|---|---|---|---|
| Crossline | Inline | ||||
| (a) Inside/penumbra region | |||||
| 15 | 7.5 (7.6 × 7.5) | 0.03 | 0.03 | 4.33 | 2.38 |
| 12.5 (12.6 × 12.5) | 0.03 | 0.01 | 2.04 | 0.61 | |
| 30.0 (32.6 × 32.5) | 0.05 | 0.05 | 1.01 | 0.53 | |
| 60.0 (62.6 × 62.5) | 0.03 | 0.02 | 0.57 | 0.30 | |
| 100 | 7.5 (7.6 × 7.5) | 0.05 | 0.02 | 3.56 | 2.10 |
| 12.5 (12.6 × 12.5) | 0.03 | 0.02 | 1.66 | 0.76 | |
| 30.0 (32.6 × 32.5) | 0.04 | 0.04 | 0.89 | 0.61 | |
| 60.0 (62.6 × 62.5) | 0.04 | 0.02 | 0.57 | 0.42 | |
| (b) Outside region | |||||
| 15 | 7.5 (7.6 × 7.5) | 0.08 | 0.06 | 0.14 | 0.33 |
| 12.5 (12.6 × 12.5) | 0.04 | 0.07 | 0.11 | 0.32 | |
| 30.0 (32.6 × 32.5) | 0.01 | 0.01 | 0.11 | 0.02 | |
| 60.0 (62.6 × 62.5) | 0.04 | 0.02 | 0.16 | 0.28 | |
| 100 | 7.5 (7.6 × 7.5) | 0.02 | 0.03 | 0.10 | 0.28 |
| 12.5 (12.6 × 12.5) | 0.03 | 0.01 | 0.11 | 0.26 | |
| 30.0 (32.6 × 32.5) | 0.03 | 0.02 | 0.14 | 0.11 | |
| 60.0 (62.6 × 62.5) | 0.06 | 0.00 | 0.17 | 0.18 | |
The field sizes are represented as diameters for the fixed and variable collimators (outside the parentheses) and as dimensions in the X and Y directions for the MLC (inside the parentheses).
Measured and calculated penumbra widths for rectangular fields formed by MLC
| Penumbra width (mm) | ||||||
|---|---|---|---|---|---|---|
| (c) Shifted 30 mm in the −X direction | (a) Centered | (b) Shifted 30 mm in the +X direction | ||||
| Measurement | Calculation | Measurement | Calculation | Measurement | Calculation | |
| −X direction | 4.1 | 4.2 | 3.7 | 3.9 | 4.0 | 3.8 |
| +X direction | 4.1 | 3.8 | 3.7 | 4.0 | 4.0 | 4.2 |
Parameters of plans developed using each type of collimator
| Collimator | Number of beams | Total MU (103) | Estimated treatment time (min) |
|---|---|---|---|
| Fixed | 177 ± 31 | 36.2 ± 2.8 | 46 ± 4 |
| Variable | 187 ± 34 | 36.4 ± 2.3 | 40 ± 3 |
| MLC | 127 ± 35 | 26.3 ± 3.0 | 32 ± 4 |
| Statistics |
|
|
|
aIn this table, *indicates that there is a significant difference between the fixed collimator and MLC results. Likewise, **indicates that there is a significant difference between the variable collimator and MLC results.
Parameters of DVHs corresponding to plans developed using each type of collimator
| Collimator | Coverage (%) | Prescribed isodose line (%) | Rectum (%) | Bladder (%) | Femoral head (%) | ||
|---|---|---|---|---|---|---|---|
| V50% | V100% | V50% | V100% | V40% | |||
| Fixed | 95.0 ± 0.2 | 82.5 ± 2.1 | 28.8 ± 10.6 | 1.5 ± 0.8 | 24.9 ± 7.4 | 2.3 ± 1.4 | n/a |
| Variable | 95.1 ± 0.5 | 83.6 ± 3.2 | 31.3 ± 10.9 | 1.2 ± 0.7 | 25.7 ± 7.8 | 2.2 ± 1.4 | n/a |
| MLC | 95.4 ± 0.5 | 82.4 ± 2.5 | 29.2 ± 10.0 | 1.4 ± 0.8 | 17.4 ± 7.6 | 2.3 ± 1.3 | n/a |
| Statistics | n.s. | n.s. | n.s. | n.s. |
| n.s. | n/a |
aIn this table, *indicates that there is a significant difference between fixed collimator and MLC results. Likewise, **indicates that there is a significant difference between the variable collimator and MLC results.
n.s., not significant; n/a, not applicable.
Gamma‐index passing rates for each patient
| Patient # | Gamma‐index passing rate (%) | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Fixed collimator | Variable collimator | MLC | |||||||
| Axial | Sagittal | Coronal | Axial | Sagittal | Coronal | Axial | Sagittal | Coronal | |
| 1 | 98.3 | 99.9 | 99.8 | 99.6 | 100.0 | 98.3 | 94.6 | 99.5 | 99.5 |
| 2 | 99.7 | 99.4 | 99.1 | 99.5 | 99.7 | 99.5 | 96.2 | 95.9 | 93.8 |
| 3 | 99.9 | 99.7 | 99.8 | 98.4 | 99.7 | 99.7 | 93.1 | 95.0 | 96.4 |
| 4 | 99.9 | 99.6 | 99.5 | 99.8 | 99.8 | 99.7 | 92.4 | 93.9 | 93.1 |
| 5 | 99.4 | 99.6 | 98.2 | 98.6 | 99.9 | 99.8 | 98.1 | 99.3 | 94.9 |
| 6 | 95.2 | 99.1 | 97.5 | 98.7 | 99.9 | 99.2 | 98.6 | 98.2 | 98.1 |
| 7 | 93.9 | 99.6 | 98.0 | 96.6 | 99.6 | 99.1 | 99.6 | 95.3 | 92.3 |
| 8 | 94.7 | 98.1 | 92.5 | 98.7 | 99.8 | 96.1 | 98.5 | 98.8 | 91.4 |
| 9 | 95.2 | 99.6 | 97.2 | 97.1 | 100.0 | 98.3 | 99.6 | 98.2 | 98.3 |
| 10 | 97.8 | 99.9 | 98.8 | 96.9 | 98.7 | 97.6 | 94.6 | 96.7 | 93.9 |
Figure 3Boxplots of gamma‐index passing rates in the (a) axial, (b) sagittal, and (c) coronal planes. In this figure, * indicates that there is a significant difference between the fixed collimator and MLC results. Likewise, ** indicates that there is a significant difference between the variable collimator and MLC results.
Figure 4Differences between doses measured using ionization chamber and calculated doses for each collimator type. In this figure, * indicates that there is a significant difference between the fixed collimator and MLC results. Likewise, ** indicates that there is a significant difference between the fixed collimator and variable collimator results.