Relationship among corneal biomechanics, refractive error, and axial length.

Inmaculada Bueno-Gimeno1, Enrique España-Gregori, Andres Gene-Sampedro, Aitor Lanzagorta-Aresti, David P Piñero-Llorens.   

Abstract

PURPOSE: To evaluate the relationship between different ocular and corneal biomechanical parameters in emmetropic and ametropic healthy white children.
METHODS: This study included 293 eyes of 293 healthy Spanish children (135 boys and 158 girls), ranging in age from 6 to 17 years. Subjects were divided according to the refractive error: control (emmetropia, 99 children), myopia (100 children), and hyperopia (94 children) groups. In all cases, corneal hysteresis (CH) and corneal resistance factor (CRF) were evaluated with the Ocular Response Analyzer system. Axial length (AL) and mean corneal power were also measured by partial coherence interferometry (IOLMaster), and central corneal thickness (CCT) and anterior chamber depth were measured by anterior segment optical coherence tomography (Visante).
RESULTS: Mean (±SD) CH and CRF were 12.12 (±1.71) and 12.30 (±1.89) mm Hg, respectively. Mean (±SD) CCT was 542.68 (±37.20) μm and mean (±SD) spherical equivalent was +0.14 (±3.41) diopters. A positive correlation was found between CH and CRF (p < 0.001), and both correlated as well with CCT (p < 0.0001). Corneal resistance factor was found to decrease with increasing age (p = 0.01). Lower levels of CH were associated with longer AL and more myopia (p < 0.001 and p = 0.001, respectively). Higher values of CH were associated with increasing hyperopia. Significant differences in CH were found between emmetropic and myopic groups (p < 0.001) and between myopic and hyperopic groups (p = 0.011). There were also significant differences in CRF between emmetropic and myopic groups (p = 0.02). Multiple linear regression analysis showed that lower CH and CRF significantly associated with thinner CCT, longer AL, and flatter corneal curvature.
CONCLUSIONS: The Ocular Response Analyzer corneal biomechanical properties seem to be compromised in myopia from an early age, especially in high myopia.

Entities:  

Mesh:

Year:  2014        PMID: 24705484     DOI: 10.1097/OPX.0000000000000231

Source DB:  PubMed          Journal:  Optom Vis Sci        ISSN: 1040-5488            Impact factor:   1.973


  23 in total

1.  Effect of refractive status on Valsalva-induced anterior segment changes.

Authors:  Necip Kara; Serhat Kenan
Journal:  Int Ophthalmol       Date:  2017-06-09       Impact factor: 2.031

2.  Corneal densitometry after photorefractive keratectomy, laser-assisted in situ keratomileusis, and small-incision lenticule extraction.

Authors:  F Poyales; N Garzón; J Mendicute; I Illarramendi; P Caro; O Jáñez; F Argüeso; A López
Journal:  Eye (Lond)       Date:  2017-06-16       Impact factor: 3.775

3.  Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.

Authors:  Xiao-Xiao Guo; Ran You; Shan-Shan Li; Xiu-Fen Yang; Lu Zhao; Fan Zhang; Yan-Ling Wang; Xi Chen
Journal:  Int J Ophthalmol       Date:  2019-10-18       Impact factor: 1.779

4.  Investigation of corneal biomechanics at moderate to high refractive errors.

Authors:  Nehir İnceoğlu; Sinan Emre; Mahmut Oğuz Ulusoy
Journal:  Int Ophthalmol       Date:  2017-05-24       Impact factor: 2.031

5.  Corneal-compensated intraocular pressure, Goldmann-correlated intraocular pressure and their associated factors in the geriatric population, a population-based study.

Authors:  Fereshteh Shokrollahzadeh; Hassan Hashemi; Abbasali Yekta; Hadi Ostadimoghaddam; Ebrahim Jafarzadehpour; Mehdi Khabazkhoob
Journal:  Int Ophthalmol       Date:  2022-01-03       Impact factor: 2.029

6.  Influence of peripheral corneal relaxing incisions during cataract surgery for corneal astigmatism up to 2.5 dioptres on corneal densitometry.

Authors:  Ieva Zukaite; Kaveeta K Bedi; Shahnaz Ali; Mayank A Nanavaty
Journal:  Eye (Lond)       Date:  2019-01-02       Impact factor: 3.775

7.  Comparison among Ocular Response Analyzer, Corvis ST and Goldmann applanation tonometry in healthy children.

Authors:  Ramin Salouti; Ali Agha Alishiri; Reza Gharebaghi; Mostafa Naderi; Khosrow Jadidi; Ahmad Shojaei-Baghini; Mohammadreza Talebnejad; Zahra Nasiri; Seyedmorteza Hosseini; Fatemeh Heidary
Journal:  Int J Ophthalmol       Date:  2018-08-18       Impact factor: 1.779

8.  Correlation between central corneal thickness and myopia.

Authors:  Michael Mimouni; Victor Flores; Yinon Shapira; Shmuel Graffi; Shmuel Levartovsky; Tzahi Sela; Gur Munzer; Igor Kaiserman
Journal:  Int Ophthalmol       Date:  2017-10-26       Impact factor: 2.031

9.  Corneal Biomechanics Determination in Healthy Myopic Subjects.

Authors:  Kunliang Qiu; Xuehui Lu; Riping Zhang; Geng Wang; Mingzhi Zhang
Journal:  J Ophthalmol       Date:  2016-07-21       Impact factor: 1.909

10.  Corneal Biomechanical Properties in Myopic Eyes Measured by a Dynamic Scheimpflug Analyzer.

Authors:  Jingyi Wang; Ying Li; Yumei Jin; Xue Yang; Chan Zhao; Qin Long
Journal:  J Ophthalmol       Date:  2015-10-20       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.