Literature DB >> 24650772

Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study.

Shinji Inaba1, Gary S Mintz, Kyeong Ho Yun, Tadayuki Yakushiji, Takehisa Shimizu, Soo-Jin Kang, Philippe Généreux, Giora Weisz, Leroy E Rabbani, Jeffrey W Moses, Gregg W Stone, Akiko Maehara.   

Abstract

AIMS: Mechanical complications contribute to bare metal and first-generation drug-eluting stent (DES) failure. However, the importance of the mechanical complications of second-generation DES remains unclear. We report mechanical complications associated with everolimus-eluting stent (EES) failures. METHODS AND
RESULTS: We retrospectively analysed 177 consecutive EES-treated lesions in 136 patients who underwent intravascular ultrasound (IVUS) at follow-up. Mechanical complications were identified in 17 patients (five stable angina, 10 unstable angina, two non-ST-elevation myocardial infarction [NSTEMI] without angiographic thrombus). Fifteen (88.2%) were treated with repeat revascularisation. By IVUS, there were 16 focal (94.1%) and one diffuse (5.9%) in-stent restenoses. Complete stent fracture with separation was seen in only one, partial stent fracture with separation was seen in three, and in 13 there was longitudinal deformation (n=2) or stent strut fracture (n=11) with overlapping of the proximal and distal stent fragments. In 13 EES with evidence of overlapping in the setting of either fracture or deformation, there was a 35.5±12.2% smaller stent area compared to the adjacent proximal and distal stent fragments, and >50% neointimal hyperplasia in 12 (92.3%).
CONCLUSIONS: We found EES mechanical complications, often followed by longitudinal deformation or fracture leading to excessive neointimal hyperplasia, in-stent restenosis, and repeat revascularisation.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24650772     DOI: 10.4244/EIJV9I11A220

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  9 in total

1.  Neoatherosclerosis assessed with optical coherence tomography in restenotic bare metal and first- and second-generation drug-eluting stents.

Authors:  Lei Song; Gary S Mintz; Dong Yin; Myong Hwa Yamamoto; Chee Yang Chin; Mitsuaki Matsumura; Khady Fall; Ajay J Kirtane; Manish A Parikh; Jeffrey W Moses; Ziad A Ali; Richard A Shlofmitz; Akiko Maehara
Journal:  Int J Cardiovasc Imaging       Date:  2017-03-09       Impact factor: 2.357

2.  Needles in Our Technology Haystacks: Defining Efficacy Is Easy, Characterizing Complications Is the Challenge.

Authors:  Elazer R Edelman; Pei-Jiang Wang
Journal:  Circ Cardiovasc Interv       Date:  2017-11       Impact factor: 6.546

3.  Impact of Intravascular Ultrasound in Clinical Practice.

Authors:  Andres Vasquez; Neville Mistry; Jasvindar Singh
Journal:  Interv Cardiol       Date:  2014-08

4.  Clinical Impact of Stent Design.

Authors:  Rebecca L Noad; Colm G Hanratty; Simon J Walsh
Journal:  Interv Cardiol       Date:  2014-04

Review 5.  Influences of Stent Design on In-Stent Restenosis and Major Cardiac Outcomes: A Scoping Review and Meta-Analysis.

Authors:  Omer Burak Istanbullu; Gulsen Akdogan
Journal:  Cardiovasc Eng Technol       Date:  2021-08-18       Impact factor: 2.495

6.  Heterogeneous Intravascular Ultrasound Findings of Stent Thrombosis.

Authors:  Toru Morofuji; Shinji Inaba; Hiroe Aisu; Kayo Takahashi; Makoto Saito; Haruhiko Higashi; Toyofumi Yoshii; Takumi Sumimoto
Journal:  Intern Med       Date:  2017-02-01       Impact factor: 1.271

Review 7.  Intravascular imaging in coronary stent restenosis: Prevention, characterization, and management.

Authors:  Amr Abouelnour; Tommaso Gori
Journal:  Front Cardiovasc Med       Date:  2022-08-09

8.  Incidence of coronary drug-eluting stent fracture: A systematic review and meta-analysis.

Authors:  Yang Chen; Dandan Li; Yanhui Liao; Xiongda Yao; Yuehua Ruan; Kai Zou; Hanhui Liao; Jingwen Ding; Hao Qin; Zuozhong Yu; Yuanbin Zhao; Longlong Hu; Renqiang Yang
Journal:  Front Cardiovasc Med       Date:  2022-08-23

9.  Mechanical stent failure as a cause of life-threatening left main restenosis.

Authors:  Piotr Kübler; Brunon Tomasiewicz; Madeleine Johansson; Andrzej Szczepański; Krzysztof Reczuch
Journal:  Postepy Kardiol Interwencyjnej       Date:  2016-08-19       Impact factor: 1.426

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.