Mechanical stent failure as a cause of life-threatening left main restenosis.

Piotr Kübler1, Brunon Tomasiewicz1, Madeleine Johansson2, Andrzej Szczepański3, Krzysztof Reczuch1.   

Abstract

Entities:  

Year:  2016        PMID: 27625693      PMCID: PMC5011546          DOI: 10.5114/aic.2016.61652

Source DB:  PubMed          Journal:  Postepy Kardiol Interwencyjnej        ISSN: 1734-9338            Impact factor:   1.426


× No keyword cloud information.
A 67-year-old woman with a history of diabetes, hypertension, severe obesity and hypercholesterolemia was admitted to our center with the diagnosis of non ST-segment elevation myocardial infarction. Electrocardiography showed 3 mm ST-segment depressions in leads I, aVF, and V3–V6 and elevation in lead aVR. Troponins were positive and ejection fraction assessed in echocardiography was 50%, without valve abnormalities. Five months earlier, the patient underwent elective percutaneous coronary intervention (PCI) of the left main coronary artery (LMCA) and left anterior descending artery (LAD) using 2 overlapping everolimus-eluting stents (Promus-Premier, Boston Scientific): 3.5/32 mm to the LMCA/LAD and 2.5/32 mm to the LAD with non-compliant 4.0 mm balloon high pressure postdilatation in the LMCA and with full expansion, an optimal angiographic result, unfortunately without intravascular ultrasound (IVUS) assistance. Qualification for PCI at that time followed the Heart Team meeting and discussion with the patient (low values of both Syntax-Score (21) and EuroSCORE II (0.9%)) – she preferred the option of PCI more than coronary artery bypass grafting (CABG). Coronary angiography revealed critical in-stent restenosis in the calcified LMCA (Figure 1) with diffuse non-significant changes in the LAD, circumflex artery and right coronary artery. During angiography severe chest pain and hemodynamic instability occurred, so immediate in-stent balloon angioplasty with 3.0/15 mm and 3.5/15 mm balloons to 10 atm was performed, enough to improve the flow in the left coronary artery and with clinical stability. However, IVUS examination in the LMCA showed some features of mechanical stent failure – separation of struts on one side of the stent circumference, in place of vessel calcification, which could be the most likely cause of restenosis. Unfortunately, because of the lack of baseline IVUS, we could not clearly diagnose the type of stent deformation. We decided to implant another Biolimus-eluting stent, 3.5/14 mm with 4.0/8 mm non-compliant balloon postdilatation, and the final result was perfect (Figure 1).
Figure 1

Coronary angiography on admission showing critical in-stent restenosis in left main coronary artery (upper left side) and final result after additional stent implantation (upper right side). Beneath intravascular ultrasound (IVUS) study with features of mechanical stent failure: separation of struts (white arrows) on one side of the stent circumference (black lines) at 4 mm from the left main ostium. Lower left: cross-section at 2 mm from the left main ostium with reference area at the level of significant restenosis (minimal lumen area 1.5 mm/2.8 mm (4.3 mm2); left main cross-section area 4.0 mm/4.3 mm (13.6 mm2); 68.1% stenosis). Final result after another stent implantation – lower right

Coronary angiography on admission showing critical in-stent restenosis in left main coronary artery (upper left side) and final result after additional stent implantation (upper right side). Beneath intravascular ultrasound (IVUS) study with features of mechanical stent failure: separation of struts (white arrows) on one side of the stent circumference (black lines) at 4 mm from the left main ostium. Lower left: cross-section at 2 mm from the left main ostium with reference area at the level of significant restenosis (minimal lumen area 1.5 mm/2.8 mm (4.3 mm2); left main cross-section area 4.0 mm/4.3 mm (13.6 mm2); 68.1% stenosis). Final result after another stent implantation – lower right Mechanical stent failure (MSF) is a rare but potentially catastrophic complication in complex PCI procedures. The most common types are longitudinal stent deformation, stent fracture, stent underexpansion and stent recoil. The reported general MSF rate of everolimus-eluting stents is about 9.6%, the longitudinal stent deformation rate is about 5.4%, and the stent fracture rate is about 2.9% [1, 2]. However, in many cases angiography alone does not reveal the presence of MSF. Intravascular imaging techniques (IVUS, optical coherence tomography) appear to be more sensitive methods and are more useful in the diagnosis of stent deformation. Because IVUS is not routinely performed, MSF rates are likely to be underestimated [1, 3]. Mechanical stent failure is associated with more frequent occurrence of major adverse cardiac events (in-stent restenosis, stent thrombosis, target lesion revascularization) [4]. Current data show that longitudinal stent deformation risk factors include everolimus stent, Guideliner use, use of postdilatation balloons and multiple stent implantation [2], and stent fracture risk factors include vessel tortuosity, longer stent length, overlapping stenting, calcified and ostial lesions and high deployment pressure [3, 4]. In our case, the stent was placed ostially, in a calcified LMCA, followed by high pressure postdilatation, which may explain the mechanical failure. Additional percutaneous intervention, stent implantation or optionally CABG is the treatment of choice in such a complication. The described case underscores the importance of IVUS-guided LMCA stenting to obtain an optimal result of the procedure.
  4 in total

1.  Mechanical complications of everolimus-eluting stents associated with adverse events: an intravascular ultrasound study.

Authors:  Shinji Inaba; Gary S Mintz; Kyeong Ho Yun; Tadayuki Yakushiji; Takehisa Shimizu; Soo-Jin Kang; Philippe Généreux; Giora Weisz; Leroy E Rabbani; Jeffrey W Moses; Gregg W Stone; Akiko Maehara
Journal:  EuroIntervention       Date:  2014-03-20       Impact factor: 6.534

2.  Incidence and clinical impact of stent fracture after everolimus-eluting stent implantation.

Authors:  Shoichi Kuramitsu; Masashi Iwabuchi; Takuya Haraguchi; Takenori Domei; Ayumu Nagae; Makoto Hyodo; Kyohei Yamaji; Yoshimitsu Soga; Takeshi Arita; Shinichi Shirai; Katsuhiro Kondo; Kenji Ando; Koyu Sakai; Masahiko Goya; Yoshitaka Takabatake; Shinjo Sonoda; Hiroyoshi Yokoi; Fumitoshi Toyota; Hideyuki Nosaka; Masakiyo Nobuyoshi
Journal:  Circ Cardiovasc Interv       Date:  2012-09-25       Impact factor: 6.546

3.  Incidence and mechanisms of longitudinal stent deformation associated with Biomatrix, Resolute, Element, and Xience stents: Angiographic and case-by-case review of 1,800 PCIs.

Authors:  Samer Arnous; Nizar Shakhshir; Andrew Wiper; Farzin-Farth Ordoubadi; Paul Williams; Bernard Clarke; Vaikom Mahadavan; Magdi El-Omar; Mamas Mamas; Douglas Fraser
Journal:  Catheter Cardiovasc Interv       Date:  2015-04-24       Impact factor: 2.692

4.  Incidence and predictors of drug-eluting stent fracture in human coronary artery a pathologic analysis.

Authors:  Gaku Nakazawa; Aloke V Finn; Marc Vorpahl; Elena Ladich; Robert Kutys; Isidora Balazs; Frank D Kolodgie; Renu Virmani
Journal:  J Am Coll Cardiol       Date:  2009-11-17       Impact factor: 24.094