| Literature DB >> 24650143 |
Claus Rath1, Martin Andreas, Caesar Khazen, Dominik Wiedemann, Andreas Habertheuer, Alfred Kocher.
Abstract
Pacemaker lead malpositioning with subsequent cardiac tamponade is a rare, but serious adverse event. We herein report a case of pacemaker lead malpositioning in a 76-year old female caucasian patient. The lead was malpositioned into the roof of the left atrium after perforation of the superior vena cava, resulting in cardiac tamponade. After fast surgical revision and an uneventful post-operative period, the patient was discharged in excellent condition.Entities:
Mesh:
Year: 2014 PMID: 24650143 PMCID: PMC3994474 DOI: 10.1186/1749-8090-9-53
Source DB: PubMed Journal: J Cardiothorac Surg ISSN: 1749-8090 Impact factor: 1.637
Pacing parameters before revision, after revision and normal ranges[3]; mV Millivolt; Ω Ohm; V Volt; ms Milliseconds
| Sensing | P-wave | 6.4 - 6.9 mV | 2.1 mV | ≥1.5 mV |
| R-wave | 9.1 - 14.7 mV | 14.6 mV | ≥5.0 mV | |
| Impedance (bipolar) | Atrial | 760 Ω | 461 Ω | 400 - 1000 Ω |
| Ventricular | 526 Ω | 861 Ω | 400 - 1000 Ω | |
| Threshold | Atrial | 0.4 V at 0.4 ms | 0.6 V at 0.5 ms | <1 V at 0.5 ms |
| Ventricular | 0.5 V at 0.4 ms | 0.5 V at 0.5 ms | <1 V at 0.5 ms |
Figure 1Chest X-ray. Atypical position of the atrial lead.
Figure 2Computed tomography scan. The atrial lead connects to the roof of the left atrium; 2a-2d Four coronal CT slices, from ventral to dorsal.
Figure 3Post-operative chest X-ray. Both leads were re-positioned.