INTRODUCTION: In vitro hydrodynamic characterization of prosthetic heart valves provides important information regarding their operation, especially if performed by noninvasive techniques of anemometry. Once velocity profiles for each valve are provided, it is possible to compare them in terms of hydrodynamic performance. In this first experimental study using laser doppler anemometry with mechanical valves, the simulations were performed at a steady flow workbench. OBJECTIVE: To compare unidimensional velocity profiles at the central plane of two bi-leaflet aortic prosthesis from St. Jude (AGN 21 - 751 and 21 AJ - 501 models) exposed to a steady flow regime, on four distinct sections, three downstream and one upstream. METHODS: To provide similar conditions for the flow through each prosthesis by a steady flow workbench (water, flow rate of 17L/min.) and, for the same sections and sweeps, to obtain the velocity profiles of each heart valve by unidimensional measurements. RESULTS: It was found that higher velocities correspond to the prosthesis with smaller inner diameter and instabilities of flow are larger as the section of interest is closer to the valve. Regions of recirculation, stagnation of flow, low pressure, and flow peak velocities were also found. CONCLUSIONS: Considering the hydrodynamic aspect and for every section measured, it could be concluded that the prosthesis model AGN 21 - 751 (RegentTM) is superior to the 21 AJ - 501 model (Master Series). Based on the results, future studies can choose to focus on specific regions of the these valves.
INTRODUCTION: In vitro hydrodynamic characterization of prosthetic heart valves provides important information regarding their operation, especially if performed by noninvasive techniques of anemometry. Once velocity profiles for each valve are provided, it is possible to compare them in terms of hydrodynamic performance. In this first experimental study using laser doppler anemometry with mechanical valves, the simulations were performed at a steady flow workbench. OBJECTIVE: To compare unidimensional velocity profiles at the central plane of two bi-leaflet aortic prosthesis from St. Jude (AGN 21 - 751 and 21 AJ - 501 models) exposed to a steady flow regime, on four distinct sections, three downstream and one upstream. METHODS: To provide similar conditions for the flow through each prosthesis by a steady flow workbench (water, flow rate of 17L/min.) and, for the same sections and sweeps, to obtain the velocity profiles of each heart valve by unidimensional measurements. RESULTS: It was found that higher velocities correspond to the prosthesis with smaller inner diameter and instabilities of flow are larger as the section of interest is closer to the valve. Regions of recirculation, stagnation of flow, low pressure, and flow peak velocities were also found. CONCLUSIONS: Considering the hydrodynamic aspect and for every section measured, it could be concluded that the prosthesis model AGN 21 - 751 (RegentTM) is superior to the 21 AJ - 501 model (Master Series). Based on the results, future studies can choose to focus on specific regions of the these valves.
In vitro hydrodynamic characterization of prosthetic heart valves
provides important information regarding their operation [, especially if
performed by noninvasive techniques of anemometry [. Regarding hydrodynamic
performance, it is possible to compare the velocity profiles for each valve and develop
new designs. Velocity profiles are different for every type of valve and regions of flow
stagnation and separation could occur, inducing formation of thrombosis, tissue
overgrowth and/or calcification as well as blood hemolysis due to shear stress
[. ISO 5840:2005 standard
offers a pattern for the hydrodynamic performance testing of prosthetic valves,
considering a steady or pulsatile flow and allowing for the evaluation of these valves
[.Nevertheless, the comparison between prosthetic heart valves is possible under some
conditions of flow similarity, which is the specific objective of this study: to compare
the velocity profiles of two St. Jude bi-leaflet aortic valves with a nominal diameter
of 21 mm exposed to a steady flow regime. The 1D velocity profiles were obtained at the
central plane of flow, on four distinct sections, three downstream and one upstream of
the valves.This first experimental study is due to an academic agreement established between
UNICAMP (Medical School, Department of Surgery) and EPUSP (Polytechnic School,
Mechanical Engineering Department). The study was developed at the Medicine Center and
Experimental Surgery and the Biomedical and Environmental Engineering Laboratory, from
the Medical School (UNICAMP) and Polytechnic School (EPUSP), respectively.
METHODS
The objective of this study was to compare, during a steady flow regime, the velocity
profiles at the central plane of two St. Jude bi-leaflet aortic valves. For this reason,
it was necessary to establish a suitable methodology for experimental hydrodynamic
testing. An academic agreement celebrated between EPUSP and UNICAMP made it possible to
use the 1D LDA system, the steady flow hydrodynamic workbench, and the valve
prostheses.The in vitro simulations presented in this paper did not have the
purpose of reproducing the physiological conditions of the test fluid (blood analog
properties of viscosity, density, and temperature), neither mimicking the physiological
pressure and volumetric flow curves. The test fluid used was water at 27ºC and a steady
flow condition was imposed in order to compare the velocity profiles of the two
prostheses under the same range of volumetric discharge. In order to meet the specific
purpose of this study (velocity profiles comparison), the ISO 5840:2005 guidelines
[ were not considered for the
hydrodynamic performance analysis in steady flow regime (i.e. , the imposition of
volumetric flow discharge from 5 to 30 L/min. , varying every 5 L/min. ), although the
unique imposed flow rate was close to the mean value established in those guidelines.
Similarly, the question of the "effective orifice area" of the prostheses, discussed in
the standard as criteria for the hydrodynamic performance analysis, was not addressed
here. These characteristics, as well as pulsatile testing, are convenient and will be
the target of next studies [.The materials and methods used in this study are presented below.
Hydrodynamic workbench for steady flow regime
The experimental workbench for the hydrodynamic testing in steady flow regime was
designed at the Biomedical and Environmental Engineering Laboratory of EPUSP and was
adapted at the Medicine Center and Experimental Surgery of UNICAMP.The workbench (Figure 1) consists of two
reservoirs, an acrylic test chamber, sealing rings, three spherical valves,
connections and pipes in PVC, and a positive displacement hydraulic pump with a
nominal flow discharge of 360 gallons per hour (22.71 L/min). The same flow discharge
was imposed for each experiment, allowing for the comparison of the velocity profiles
obtained from the valves. An indirect method of volumetric flow measurement was used,
by determining the effective flow of the pump once the steady state regime was
performed on the hydrodynamic workbench.
Steady flow hydrodynamic workbench photo Highlighted: A (test chamber), 1 (top
spherical valve), 2 (bottom spherical valve) and 3 (maintenance spherical
valve)The effective pump flow discharge was obtained through variations in observed volume
in a chamber of the experimental workbench during a certain period of time measured
by a chronometer. With the experimental workbench adjusted for steady flow and with
the hydraulic circuit to the testing chamber blocked (changing the valve in Figure 1, n.1, to the closed position), the filling up of
the superior reservoir up to a stipulated height for a registered period of time was
observed. The water volume was determined through the internal area of the base of
the superior reservoir. The effective pump flow discharge was obtained dividing this
volume by the measured time.The steady flow regime was performed when the water level in the reservoirs remained
invariable. The same flow resistance in the hydraulic circuit (equivalent resistance)
was established for each experiment. Considering the differences between the two
cardiac prostheses used, they offer distinct resistances to flow circulation, which
implies different pressure distributions for each one (see the Results section). The
equivalent flow resistance (total head loss) for each experiment was ensured by
regulating the inferior spherical valve of the hydrodynamic workbench (Figure 1, n. 2). This spherical valve implies an
adjustable flow resistance to the hydraulic circuit. To validate each experiment, the
other two spherical valves (Figure 1, n. 1 and
n. 3) remained , fully open and fully closed, respectively. In addition, the same
total water volume was ensured on every testing. The effective flow value can be seen
in the results section.In order to use the LDA system, a test chamber was conceived in a specific region of
the hydraulic circuit (Figure 1, Detail A). The
design of this test chamber is shown in Figure
2. It consists of two pipes and an optical confinement (Figure 2, n. 1), both in acrylic, and seal rings.
The prosthesis (Figure 2, n. 4) is positioned
under pressure, by longitudinal assembling of the two pipes, which in turn slide
within the test chamber orifices. The two pipes downstream and upstream of the
prosthesis (respectively, n. 2 and n. 3 in Figure
2) have internal diameters of 21.20 mm (able to shelter the chosen
prosthesis) and both pipes are 210 mm long (upstream pipe length allows for the
developed flow to be measured; downstream pipe length offers LDA measurements on
different sections). Each pipe has one orifice to acquire pressure (Figure 2, n. 5) based on the height of the water
column so that differential pressure can be obtained. When the pipes are positioned,
they lock the prosthesis inside them. In this condition, the orifices are located 40
mm upstream and 20 mm downstream of the prosthesis.
Fig. 2
Workbench test chamber detail Highlighted: 1 (optical confinement), 2 (upstream
pipe), 3 (downstream pipe), 4 (bi-leaflet mechanical heart valve) and 5 (holes
for differential pressure acquiring)
Workbench test chamber detail Highlighted: 1 (optical confinement), 2 (upstream
pipe), 3 (downstream pipe), 4 (bi-leaflet mechanical heart valve) and 5 (holes
for differential pressure acquiring)
Prostheses used
The high frequency of valve replacement in the aortic position gave support to the
choice of these prosthesis models in the research here presented [.Two St. Jude bi-leaflet prosthetic aortic valves, models 21 AGN - 751
(RegentTM) and 21 AJ - 501 (Masters Series), [ were used. Although the two valves
have the same nominal diameter of 21 mm, they have internal diameters of 19.6 mm and
16.7 mm, respectively. These prostheses models are shown in Figure 3.
Fig. 3
Aortic bi-leaflet prosthetic heart valves used Highlighted: from left to right,
models 21 AGN - 751 and 21 AJ - 501
Aortic bi-leaflet prosthetic heart valves used Highlighted: from left to right,
models 21 AGN - 751 and 21 AJ - 501Considering dimensional variations of mechanical valves are negligible, only one
prosthesis of each model was used in this study. So the results were considered
independent of the number of samples. On the other hand, the internal diameter
difference of the two prostheses (two different models) enabled the comparison and
discussion of the results. Further studies are necessary to properly define the
number of samples (prostheses).
LDA used: working principle and description
Before describing the LDA system used, its working principle is briefly
discussed.The laser Doppler anemometry system (or LDA) consists of a first stage of laser beam
transmission so that pairs of laser beams converge into an intersection point,
representing the point of interest to be measured (flow containing seeding
particles). At the same time, another stage occurs, which is characterized by the
detection of scattered light radiation from the small intersection volume (when
particle motion is due to specific flow velocities). Then, these data can be
conditioned and processed for a particular type of information, since the flashing
light frequency (Doppler frequency) is proportional to the flow velocity at the
measurement point.In the transmission phase, the purpose is to have the pairs of laser beams converging
in order to form a measuring volume at a particular intersection point. Since it is
usually possible to separate three laser-beam wavelengths (violet, blue, and green)
generated by the source, each wavelength can be manipulated in pairs of beams in
orthogonal planes and it can provide information on up to three velocity components
simultaneously: in each plane and for a specific point. This is why LDA systems are
suitable for accurate velocity measurements. The most well-known configuration is
called backscattering. In this configuration, the probes perform two simultaneous
functions. Firstly, they are responsible for the convergence of monochromatic pairs
of laser beams into the measurement point (intersection volume) through the outer
lens (focusing lens). Secondly, they receive the scattered light (from seeding
particles) through the inner lens (reception phase).When crossing the measurement volume, the seeding particles (contained in the flow)
induce the scattering of light in varying intensity according to the flow velocity at
that point. This returning light is redirected to the detection, signal conditioning,
and processing phases. Finally, the results of the processing phase are manipulated
and displayed using specific software. Thus, it is possible to know the velocity
components of flow (1D, 2D, or 3D, depending on the system configuration) through
sweeping points comprised of a linear sequence of measurement volumes.In this study, the LDA system of the Laboratory of Surgical Technique and
Experimental Surgery at UNICAMP was used. The equipment is from Dantec Dynamics and
it is actually capable of 1D velocimetry measurements only. This LDA system is based
on Argon ions laser (Innova 70 Coherent, nominal power of 4 W), which is refrigerated
by air with a backscattering configuration. The BSA Flow Software, from the same
company, was also used.The scope of this study was the use of the 1D LDA system in order to obtain
information about the horizontal plane of the flow. In addition, a manual traverse
system was used for positioning the probe and, consequently, scanning the observed
measurement points upstream and downstream of the prostheses.
RESULTS
The 1D measurement results (one section upstream and three sections downstream of the
valves) obtained for the two St. Jude bi-leaflet prosthetic aortic valves through a
steady flow are presented. Introductorily, the flow conditions, which were derived from
the operating workbench, as well as the proper positioning of the LDA probe, are
presented below.
Preliminary preparation
The prostheses were arranged in the test chamber as described in the methodology. The
test fluid used was water at 27°C, with the addition of seeding particles of 20 µm in
diameter (Dantec Dynamics, Polyamide Seeding Particles).Plastic hoses were conveniently connected with a manometer in order to acquire
differential pressure from the pipes' orifices upstream and downstream of the
prosthesis (Figure 2, n. 5).As mentioned in the methods section, the flow used in the test chamber was the pump's
effective flow, once both the same volume of water on the workbench and the steady
flow regime on each experiment was established. The volumetric flow discharge was
obtained indirectly: volume variation over measured time. This procedure was repeated
three times and an arithmetic average was obtained. It was found that the flow rate
imposed on the prosthesis was approximately 17 L/min. and this value represents the
only possible flow discharge for the test chamber.With the LDA system in operation, the manual traverse system was referenced so that
the laser beams always reached the mean horizontal plane of the pipes. A controlled
routine established, via software, for the operation of the LDA system allowed some
variables of the spectral analyzer to be controlled during the experiments, such as:
the acquisition rate (up to 10 KHz), the photomultiplier voltage (up to 1,000 V), the
amplifier signal gain (35 dB), and the operating power (170 mW). As described in the
methods section, the LDA system operated only as 1D, through a probe with a pair of
laser beams for the green spectrum and with the following characteristics: wavelength
(ƛ) of 514.5 nm, diameter of 1.35 mm, focal length of 160 mm, spacing of the laser
beam pair in the focal lens (frontals) of 38 mm, and fringes spacing (at the
intersection volume of the beans) of 2.182 µm, with a total of 35 fringes.After establishing the horizontal plane for the reference position, the probe was
positioned in the sections and points of interest, as shown in Figure 4.
Fig. 4
Measurement sections referenced at the prosthesis positioning Highlighted: in
the flow direction, upstream at 30 mm, downstream 1 at 8 mm, downstream 2 at 20
mm and downstream 3 at 32 mm from the valve
Measurement sections referenced at the prosthesis positioning Highlighted: in
the flow direction, upstream at 30 mm, downstream 1 at 8 mm, downstream 2 at 20
mm and downstream 3 at 32 mm from the valveFollowing the flow direction, the first point of interest was at 30 mm before the
prosthesis (Figure 4, upstream). Three other
points of interest were located after the prosthesis at 8 mm (Figure 4, downstream 1), 20 mm (Figure 4, downstream 2), and 32 mm (Figure
4, downstream 3). Therefore, the 1D LDA measurements consisted of scans of
four sections for each prosthesis: one upstream and three downstream of the
valve.
Measurements in the hydrodynamic workbench
After each prosthesis was assembled in the test chamber (Figure 2), the steady flow was established, and the probe was
conveniently positioned facing the measurement point (Figure 4), the LDA system was used to obtain the 1D velocity profiles at
the central plane along the inner diameter of the pipes (21.20 mm) and for four
sections along flow direction. Every new measurement at the point of interest was
obtained by displacing the LDA probe longitudinally through 0.50 mm over the
horizontal diameter plane.Under 17L/min. of flow discharge, differential pressure was 42 mm of water column (or
3.09 mmHg) for assembling the prosthesis 21 AGN - 751 (internal diameter of 19.6 mm)
and 63 mm of water column (or 4.63 mmHg) for the prosthesis 21 AJ - 501 (internal
diameter of 16.7 mm).Figures 5 to 8 show the measured velocity profiles for the two prosthetic valve models,
under steady flow conditions with a volumetric discharge of 17 L/min, for each point
of interest. The discussion of the results is presented in the next section.
Fig. 5
Velocity profiles at 30 mm upstream of the valves
Fig. 8
Velocity profiles at 32 mm downstream of the valves
Velocity profiles at 30 mm upstream of the valvesVelocity profiles at 32 mm downstream of the valves
DISCUSSION
As stated in the methods section, the scope of this study does not address the
guidelines of ISO 5840:2005 for hydrodynamic performance analysis. It only consists of
accomplishing a comparison of the velocity profiles obtained for two models of valves
with different internal diameters, under a common condition (volumetric discharge of 17
L/min.). Therefore, the prostheses provide different flow restrictions in the passage of
the flow, which implies different localized head loss. According to the literature, a
prosthesis with larger diameter offers a smaller head loss. To ensure that in each of
the experiments the same equivalent resistance was imposed on the hydraulic circuit, we
used a spherical valve (Figure 1, n. 2). In fact,
when the prosthesis with smaller diameter was used, this specific spherical valve was
kept more open.Each velocity profile obtained at 30 mm upstream the valve allowed for the flow
discharge to be estimated by integrating the velocity profile in the referenced area.
This calculation confirms the previously measured value of 17 L/min. with an error
margin of approximately 5%. However, the testing by means of this single flow discharge
represents an intrinsic limitation of this study [.The results obtained using the 1D LDA correspond to those expected from the literature:
for all of the downstream measurement sections, greater velocities correspond to the
prosthesis with a smaller internal diameter (16.7 mm, St. Jude model 21 AJ - 501), with
higher transversal gradients near the pipe wall. In terms of pressure measurements, the
prosthesis with a larger internal diameter (19.6 mm, St. Jude model 21 AGN - 751)
presented smaller values of differential pressure and, consequently, smaller local head
loss (42 mm of water column, or 3.09 mmHg). This implies smaller resistance to the
passage of flow, compared with the prosthesis with a smaller diameter. In terms of flow
instabilities downstream the prosthesis, it was observed that they are greater in the
section near the prosthesis (section 1, downstream). It was found that although certain
symmetry of the velocity profiles occurs, this symmetry is not significant.Some small negative values of velocity were measured with the LDA system. Recirculation
zones were observed for both prosthesis models, St. Jude 21 AGN - 751 and St. Jude 21 AJ
- 501, particularly in the downstream sections. Although negative values of velocity
were expected in the prostheses surroundings, they were not expected at the farthest
sections (downstream 2 and 3). The transversal gradients of velocity are much more
pronounced in the case of the prosthetic valve with a larger diameter (St. Jude 21 AJ -
501), which is possible to observe by analyzing Figures
6 and 7. Similar future studies with
measurements in more than one direction, i. e. , 2D LDA measurements, will be used to
validate the results obtained in this study.
Fig. 6
Velocity profiles at 8 mm downstream of the valves
Fig. 7
Velocity profiles at 20 mm downstream of the valves
Velocity profiles at 8 mm downstream of the valvesVelocity profiles at 20 mm downstream of the valvesIn terms of hydrodynamics, the prosthesis with a larger internal diameter should be
adopted, considering the smaller peak velocities in the aortic root and the smaller
transversal velocity gradients in this case, with less probability of recirculation.
This is in accordance with the adequate sizing of the effective orifice area criteria so
that residual stenosis after valve implantation can be avoided, thereby minimizing the
occurrence of elevated pressure gradients through the valve [. On the other hand, the occurrence of unsatisfactory
pressure gradient across the prosthesis should also be avoided, as it would result in an
incomplete reduction of the left ventricular hypertrophy [.
Manufacturers have developed models of valves for different sutures without addressing
controversies about the proper diameter of the prosthesis according to the patient's
aortic annulus [ and relying on the experience of clinical practice
favoring hemodynamic performance.This can be seen for the two St. Jude prostheses tested here. The AGN 21 - 751
(RegentTM) model is designed for supra-annular suture and the AJ 21 - 501
(Masters Series) model for intra-annular suture. Currently, due to hemodynamic
advantages, most surgeons employ the supra-annular positioning, even if there is a
discrepancy among manufacturers regarding different internal diameter of the prostheses
based on valves with the same nominal diameter [. For the prostheses
studied here, which are from the same manufacturer, the choice of supra-annular
prosthesis implies a valve with a larger inner diameter, although the two models have
the same nominal diameter (21 mm). In fact, smaller peak velocities and a better
hydrodynamic behavior were observed for all measurements of the supra-annular prosthesis
21 AGN - 751 (RegentTM) points of interest (Figures
6 to 8). However, it is not possible to
disregard some surgical priorities that can be more relevant in some cases than the
hydrodynamic aspects for a certain group of patients [.Clearly, the present study has some limitations because it does not include other flow
ranges besides 17 L / min. [ and the LDA system available is
restricted to 1D measurements. It would be feasible to extend this study by using
another type of pump and including additional spherical valves and a flowmeter in the
hydrodynamic circuit, so that it is possible to adjust other values of volumetric
discharge in steady flow regime. As explained in the methodology, this will be the focus
of the next study, using a cardiac simulator (USP), where 2D laser anemometry will be
used for velocity measurements [.
CONCLUSION
For the two prosthetic valves tested according the hydrodynamic aspect considered, it
was possible to verify the superiority of the prosthesis model AGN 21 - 751
(RegentTM) comparing with model 21 AJ - 501 (Master Series). This choice
implies the supra-annular positioning. The results permit to focus, in next work, the
observations and measurements in some specific regions nearby the prosthesis were the
flow recirculation and peak velocities occurs. According with was exposed in methods,
for further testing will be possible include a statistical analysis based on a batch of
valves and regarding some guidelines suggested by the ISO 5840:2005 standard.
Abbreviations, acronyms & symbols
1D
Unidimensional
2D
Bidimensional
3D
Tridimensional
LDA
Laser Doppler anemometer
Author's roles & responsibilities
OB
LDA equipment maintenance estimating; steady workbench design and
manufacturing (except the test section of the prosthesis), test section CAD
re-drafting; transfering and assembling of hydrodynamic workbench between USP
and Unicamp; LDA training; LDA testing with the prostheses; paper writing,
figures, graphs and resion; revisor replicas.
JPO:
Academic partnership coordinator and linked jobs between USP and UNICAMP, LDA
trials monitoring; line of research guiding at USP; paper revision and revisor
replicas.
FUVJ:
Workbench test section CAD drawings; test section manufacturing; LDA training;
LDA testing with the prostheses; paper revision.
RWV:
Academic partnership coordinator and linked jobs between USP and UNICAMP, LDA
trials monitoring; line of research guiding at Unicamp; paper review.
NA:
Responsible for resources finding, maintenance and restructuring of Medicine
Center and Experimental Surgery Laboratory (LDA system); responsible for the
heart valves obtainning from St. Jude Medical Brazil.
FDBT:
Transfering and assembling of hydrodynamic workbench between USP and Unicamp;
LDA training; LDA testing with the prostheses; testing photos.
ETC:
Responsible for the workbench test section building (by the Center for
Biomedical Engineering, CEB, Unicamp ) and enable the maintaining resources of
the LDA equipment via CEB.
OPJ:
Responsible for the Cardiac Surgery discipline at Unicamp and agreementing
assignatures allowing the LDA maintenance.
Authors: Ruggero De Paulis; Christoph Schmitz; Raffaele Scaffa; Paolo Nardi; Luigi Chiariello; Helmut Reul Journal: J Thorac Cardiovasc Surg Date: 2005-10 Impact factor: 5.209
Authors: Mauro Grigioni; Carla Daniele; Giuseppe D'Avenio; Umberto Morbiducci; Costantino Del Gaudio; Mara Abbate; David Di Meo Journal: Expert Rev Med Devices Date: 2004-09 Impact factor: 3.166