Design conception and experimental setup for in vitro evaluation of mitral prosthetic valves.

Ovandir Bazan1, Jayme Pinto Ortiz.   

Abstract

BACKGROUND: Since most complications related to the operation of prosthetic heart valves is due to disturbances of flow, its hydrodynamic characterization is a useful aid in the design of new prostheses. Simulations of pulsatile flow in cardiac prostheses began nearly 40 years ago, through the development of different mock human circulatory systems, improving the clinical results interpretation. A new design of a pulse duplicator system was developed at Polytechnic School of USP to study prosthetic heart valves.
OBJECTIVE: To present the conception of a new mock circulatory system for hydrodynamic simulations of cardiac prosthetic valves and the assembly plan of an experiment whose focus is the test of mitral prosthesis.
METHODS: Its conception is based on the state-of-art's review of these studies and the experience got with the previous mock circulatory systems, particularly the one used in the Instituto Dante Pazzanese de Cardiologia, São Paulo, SP, Brazil.
RESULTS: In this design, an electric servomotor controlled by computer emits, through a hydraulic piston, a pulse to the left ventricular chamber model, where the heart valves are accomodated. To characterize, in the future, the dynamic operation of mitral prosthetic valves, an experimental setup was mounted to provide measurements of volumetric flow, instantaneous pressure and velocity fields on these valves. Optical access is conveniently provided on the design, making possible the use, in the future, of a LDA system.
CONCLUSIONS: In order to improve the analysis of hydrodynamic shear stress and prediction of haemolysis, the experimental results may be used to regulate a numerical model using 'Computational Fluid Dynamics' (CFD).

Entities:  

Mesh:

Year:  2011        PMID: 21894409

Source DB:  PubMed          Journal:  Rev Bras Cir Cardiovasc


  2 in total

1.  Experimental Validation of a Cardiac Simulator for in vitro Evaluation of Prosthetic Heart Valves.

Authors:  Ovandir Bazan; Jayme Pinto Ortiz
Journal:  Braz J Cardiovasc Surg       Date:  2016-04

2.  Laser Doppler anemometry measurements of steady flow through two bi-leaflet prosthetic heart valves.

Authors:  Ovandir Bazan; Jayme Pinto Ortiz; Francisco Ubaldo Vieira; Reinaldo Wilson Vieira; Nilson Antunes; Fabio Bittencourt Dutra Tabacow; Eduardo Tavares Costa; Orlando Petrucci
Journal:  Rev Bras Cir Cardiovasc       Date:  2013 Oct-Dec
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.