Biomonitoring method for bisphenol A in human urine by ultra-high-performance liquid chromatography-tandem mass spectrometry.

David J Anderson1, Eric M Brozek2, Kyley J Cox3, Christina A Porucznik3, Diana G Wilkins4.   

Abstract

An ultra-high-performance liquid chromatography-tandem mass spectrometry method for the measurement of total bisphenol A in human urine was developed and validated. The method utilized liquid/liquid extraction with 1-chlorobutane and a human urine aliquot size of 800μL. Chromatography was performed on an Acquity UPLC(®) system with a Kinetex(®) Phenyl-Hexyl column. Mass spectrometric analysis was with negative electrospray ionization on a Quattro Premier XE™. The surrogate matrix method was used for the preparation of calibration standards in synthetic urine due to the presence of BPA in control human urine. The validated calibration range was 0.75-20ng/mL with a limit of detection of 0.1ng/mL. The internal standard was d16-bisphenol A. Method validation utilized quality control samples at three concentrations in both synthetic urine and human urine. Bisphenol A mono-glucuronide was fortified in synthetic urine in each analytical run to monitor the enzymatic conversion of the glucuronide conjugate to BPA by β-glucuronidase. Validated method parameters included linearity, accuracy, precision, integrity of dilution, selectivity, re-injection reproducibility, recovery/matrix effect, solution stability, and matrix stability in human urine. Acceptance criteria for analytical standards and QCs were ±20% of nominal concentration. Matrix stability in human urine was validated after 24h at ambient temperature, after three freeze/thaw cycles, and after frozen storage at -20°C and -80°C for up to 218 days. The method has been applied to the analysis of over 1750 human urine samples from a biomonitoring study. The median and mean urine BPA concentrations were 2.71ng/mL and 4.75ng/mL, respectively.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Bisphenol A; Human urine; Liquid chromatography; Tandem mass spectrometry

Mesh:

Substances:

Year:  2014        PMID: 24594944      PMCID: PMC4068800          DOI: 10.1016/j.jchromb.2014.01.039

Source DB:  PubMed          Journal:  J Chromatogr B Analyt Technol Biomed Life Sci        ISSN: 1570-0232            Impact factor:   3.205


  38 in total

1.  Automated on-line column-switching HPLC-MS/MS method for measuring environmental phenols and parabens in serum.

Authors:  Xiaoyun Ye; Lily J Tao; Larry L Needham; Antonia M Calafat
Journal:  Talanta       Date:  2008-04-24       Impact factor: 6.057

2.  On-line solid phase extraction fast liquid chromatography-tandem mass spectrometry for the analysis of bisphenol A and its chlorinated derivatives in water samples.

Authors:  H Gallart-Ayala; E Moyano; M T Galceran
Journal:  J Chromatogr A       Date:  2010-03-23       Impact factor: 4.759

3.  Simultaneous quantification of bisphenol A and its glucuronide metabolite (BPA-G) in plasma and urine: applicability to toxicokinetic investigations.

Authors:  M Z Lacroix; S Puel; S H Collet; T Corbel; N Picard-Hagen; P L Toutain; C Viguié; V Gayrard
Journal:  Talanta       Date:  2011-07-19       Impact factor: 6.057

4.  Impact of early-life bisphenol A exposure on behavior and executive function in children.

Authors:  Joe M Braun; Amy E Kalkbrenner; Antonia M Calafat; Kimberly Yolton; Xiaoyun Ye; Kim N Dietrich; Bruce P Lanphear
Journal:  Pediatrics       Date:  2011-10-24       Impact factor: 7.124

5.  Development of a method for the determination of bisphenol A at trace concentrations in human blood and urine and elucidation of factors influencing method accuracy and sensitivity.

Authors:  Dan Alan Markham; John M Waechter; Martina Wimber; Narayana Rao; Paul Connolly; Jane Chen Chuang; Steven Hentges; Ronald N Shiotsuka; Stephen Dimond; Anne H Chappelle
Journal:  J Anal Toxicol       Date:  2010 Jul-Aug       Impact factor: 3.367

6.  Simultaneous quantification of multiple classes of phenolic compounds in blood plasma by liquid chromatography-electrospray tandem mass spectrometry.

Authors:  Hong Chang; Yi Wan; Jonathan Naile; Xiaowei Zhang; Steve Wiseman; Markus Hecker; Michael H W Lam; John P Giesy; Paul D Jones
Journal:  J Chromatogr A       Date:  2009-12-03       Impact factor: 4.759

Review 7.  Human exposure to bisphenol A by biomonitoring: methods, results and assessment of environmental exposures.

Authors:  Wolfgang Dekant; Wolfgang Völkel
Journal:  Toxicol Appl Pharmacol       Date:  2007-12-14       Impact factor: 4.219

Review 8.  Human exposure to bisphenol A (BPA).

Authors:  Laura N Vandenberg; Russ Hauser; Michele Marcus; Nicolas Olea; Wade V Welshons
Journal:  Reprod Toxicol       Date:  2007-07-31       Impact factor: 3.143

9.  Simultaneous determination of multiple phthalate metabolites and bisphenol-A in human urine by liquid chromatography-tandem mass spectrometry.

Authors:  Mei Chen; Lin Tao; Erin M Collins; Christine Austin; Chensheng Lu
Journal:  J Chromatogr B Analyt Technol Biomed Life Sci       Date:  2012-07-31       Impact factor: 3.205

10.  Comparison of degradation between indigenous and spiked bisphenol A and triclosan in a biosolids amended soil.

Authors:  Kate A Langdon; Michael Stj Warne; Ronald J Smernik; Ali Shareef; Rai S Kookana
Journal:  Sci Total Environ       Date:  2013-01-31       Impact factor: 7.963

View more
  4 in total

1.  Association between peri-conceptional bisphenol A exposure in women and men and time to pregnancy-The HOPE study.

Authors:  Dabin Yeum; Shinyoung Ju; Kyley J Cox; Yue Zhang; Joseph B Stanford; Christina A Porucznik
Journal:  Paediatr Perinat Epidemiol       Date:  2019-08-29       Impact factor: 3.980

2.  Male exposure to bisphenol A (BPA) and semen quality in the Home Observation of Periconceptional Exposures (HOPE) cohort.

Authors:  Sarah Hatch Pollard; Kyley J Cox; Brenna E Blackburn; Diana G Wilkins; Douglas T Carrell; Joseph B Stanford; Christina A Porucznik
Journal:  Reprod Toxicol       Date:  2019-08-21       Impact factor: 3.143

3.  A Preliminary Study of Biomonitoring for Bisphenol-A in Human Sweat.

Authors:  Christina A Porucznik; Kyley J Cox; Diana G Wilkins; David J Anderson; Nicole M Bailey; Kathryn M Szczotka; Joseph B Stanford
Journal:  J Anal Toxicol       Date:  2015-05-25       Impact factor: 3.220

4.  Exposure Classification and Temporal Variability in Urinary Bisphenol A Concentrations among Couples in Utah--The HOPE Study.

Authors:  Kyley J Cox; Christina A Porucznik; David J Anderson; Eric M Brozek; Kathryn M Szczotka; Nicole M Bailey; Diana G Wilkins; Joseph B Stanford
Journal:  Environ Health Perspect       Date:  2015-09-15       Impact factor: 9.031

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.