Literature DB >> 24506636

Dosimetric characterization and output verification for conical brachytherapy surface applicators. Part II. High dose rate 192Ir sources.

Regina K Fulkerson1, John A Micka1, Larry A DeWerd1.   

Abstract

PURPOSE: Historically, treatment of malignant surface lesions has been achieved with linear accelerator based electron beams or superficial x-ray beams. Recent developments in the field of brachytherapy now allow for the treatment of surface lesions with specialized conical applicators placed directly on the lesion. Applicators are available for use with high dose rate (HDR)(192)Ir sources, as well as electronic brachytherapy sources. Part I of this paper discussed the applicators used with electronic brachytherapy sources. Part II will discuss those used with HDR (192)Ir sources. Although the use of these applicators has gained in popularity, the dosimetric characteristics have not been independently verified. Additionally, there is no recognized method of output verification for quality assurance procedures with applicators like these.
METHODS: This work aims to create a cohesive method of output verification that can be used to determine the dose at the treatment surface as part of a quality assurance/commissioning process for surface applicators used with HDR electronic brachytherapy sources (Part I) and(192)Ir sources (Part II). Air-kerma rate measurements for the (192)Ir sources were completed with several models of small-volume ionization chambers to obtain an air-kerma rate at the treatment surface for each applicator. Correction factors were calculated using MCNP5 and EGSnrc Monte Carlo codes in order to determine an applicator-specific absorbed dose to water at the treatment surface from the measured air-kerma rate. Additionally, relative dose measurements of the surface dose distributions and characteristic depth dose curves were completed in-phantom.
RESULTS: Theoretical dose distributions and depth dose curves were generated for each applicator and agreed well with the measured values. A method of output verification was created that allows users to determine the applicator-specific dose to water at the treatment surface based on a measured air-kerma rate.
CONCLUSIONS: The novel output verification methods described in this work will reduce uncertainties in dose delivery for treatments with these kinds of surface applicators, ultimately improving patient care.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24506636      PMCID: PMC3987768          DOI: 10.1118/1.4862506

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  16 in total

1.  AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology.

Authors:  C M Ma; C W Coffey; L A DeWerd; C Liu; R Nath; S M Seltzer; J P Seuntjens
Journal:  Med Phys       Date:  2001-06       Impact factor: 4.071

2.  Calibration of new high dose rate 192Ir sources.

Authors:  K E Stump; L A Dewerd; J A Micka; D R Anderson
Journal:  Med Phys       Date:  2002-07       Impact factor: 4.071

3.  Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations.

Authors:  Mark J Rivard; Bert M Coursey; Larry A DeWerd; William F Hanson; M Saiful Huq; Geoffrey S Ibbott; Michael G Mitch; Ravinder Nath; Jeffrey F Williamson
Journal:  Med Phys       Date:  2004-03       Impact factor: 4.071

4.  Comparison of air-kerma strength determinations for HDR (192)Ir sources.

Authors:  Brian E Rasmussen; Stephen D Davis; Cal R Schmidt; John A Micka; Larry A Dewerd
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

5.  Experimental and Monte Carlo determination of the TG-43 dosimetric parameters for the model 9011 THINSeed brachytherapy source.

Authors:  R M Kennedy; S D Davis; J A Micka; L A DeWerd
Journal:  Med Phys       Date:  2010-04       Impact factor: 4.071

6.  EGSnrc Monte Carlo calculated dosimetry parameters for 192Ir and 169Yb brachytherapy sources.

Authors:  R E P Taylor; D W O Rogers
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

7.  Calibration of 192Ir high-dose-rate afterloading systems.

Authors:  S J Goetsch; F H Attix; D W Pearson; B R Thomadsen
Journal:  Med Phys       Date:  1991 May-Jun       Impact factor: 4.071

8.  Comment on "correspondence factor for nucletron surface applicators".

Authors:  Harvey Stephen Gotts
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

9.  Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No. 43. American Association of Physicists in Medicine.

Authors:  R Nath; L L Anderson; G Luxton; K A Weaver; J F Williamson; A S Meigooni
Journal:  Med Phys       Date:  1995-02       Impact factor: 4.071

10.  Design and evaluation of a HDR skin applicator with flattening filter.

Authors:  D Granero; J Pérez-Calatayud; J Gimeno; F Ballester; E Casal; V Crispín; R van der Laarse
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

View more
  5 in total

1.  Recommendations of the Spanish brachytherapy group (GEB) of Spanish Society of Radiation Oncology (SEOR) and the Spanish Society of Medical Physics (SEFM) for high-dose rate (HDR) non melanoma skin cancer brachytherapy.

Authors:  S Rodríguez; M Arenas; C Gutierrez; J Richart; J Perez-Calatayud; F Celada; M Santos; A Rovirosa
Journal:  Clin Transl Oncol       Date:  2017-08-14       Impact factor: 3.405

2.  Surface dose characterisation of the Varian Ir-192 HDR conical surface applicator set with a vertically orientated source.

Authors:  Konrad Buchauer; Guido Henke; Hans Schiefer; Ludwig Plasswilm
Journal:  Strahlenther Onkol       Date:  2014-08-01       Impact factor: 3.621

3.  Impact of source position on high-dose-rate skin surface applicator dosimetry.

Authors:  Jeho Jeong; Christopher A Barker; Marco Zaider; Gil'ad N Cohen
Journal:  Brachytherapy       Date:  2016-06-01       Impact factor: 2.362

4.  Dosimetric characteristics of a new unit for electronic skin brachytherapy.

Authors:  Teresa Garcia-Martinez; Jan-Pieter Chan; Jose Perez-Calatayud; Facundo Ballester
Journal:  J Contemp Brachytherapy       Date:  2014-02-19

5.  Varian HDR surface applicators - commissioning and clinical implementation.

Authors:  Ileana Iftimia; Andrea B McKee; Per H Halvorsen
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

  5 in total