Repeatability and reproducibility of eight macular intra-retinal layer thicknesses determined by an automated segmentation algorithm using two SD-OCT instruments.

Xinting Liu1, Meixiao Shen1, Shenghai Huang1, Lin Leng1, Dexi Zhu1, Fan Lu1.   

Abstract

PURPOSE: To evaluate the repeatability, reproducibility, and agreement of thickness profile measurements of eight intra-retinal layers determined by an automated algorithm applied to optical coherence tomography (OCT) images from two different instruments.
METHODS: Twenty normal subjects (12 males, 8 females; 24 to 32 years old) were enrolled. Imaging was performed with a custom built ultra-high resolution OCT instrument (UHR-OCT, ∼3 µm resolution) and a commercial RTVue100 OCT (∼5 µm resolution) instrument. An automated algorithm was developed to segment the macular retina into eight layers and quantitate the thickness of each layer. The right eye of each subject was imaged two times by the first examiner using each instrument to assess intra-observer repeatability and once by the second examiner to assess inter-observer reproducibility. The intraclass correlation coefficient (ICC) and coefficients of repeatability and reproducibility (COR) were analyzed to evaluate the reliability.
RESULTS: The ICCs for the intra-observer repeatability and inter-observer reproducibility of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers, except the photoreceptor inner segments, which ranged from 0.051 to 0.643, and the outer segments, which ranged from 0.709 to 0.959. The CORs were less than 6.73% for the total retina and all intra-retinal layers. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. However, the ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the inner segment and outer segment layers.
CONCLUSIONS: Thickness measurements of the intra-retinal layers determined by the automated algorithm are reliable when applied to images acquired by the UHR-OCT and RTVue100 instruments.

Entities:  

Mesh:

Year:  2014        PMID: 24505345      PMCID: PMC3914879          DOI: 10.1371/journal.pone.0087996

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Evaluation of intra-retinal layer thickness plays an important role in the diagnosis and monitoring of various retinal diseases. For example, thinning of the retinal nerve fiber layer (RNFL) is often noted in glaucoma and myopia [1], [2], and thinning of the ganglion cell complex (GCC) occurs during the development of glaucoma [3]. The thickness of the outer nuclear layer (ONL) in the fovea was reported to correlate with visual acuity in central serous chorioretinopathy eyes [4]. In early age-related macular degeneration when vision field defects are present, the photoreceptor outer segment (OS) layer thins and the retinal pigment epithelium (RPE) thickens [5]. Segmentation of intra-retinal layers is very important not only in ophthalmology but also in neurology. Except for the RNFL, the thicknesses of the deeper retinal layers are reported to change in multiple sclerosis, Parkinsonian syndromes, and less frequently in disorders such as neuromyelitis optica and Wilson’s disease [6], [7], [8], [9], [10]. Optical coherence tomography (OCT) is a noninvasive, noncontact diagnostic tool that can provide in vivo cross-sectional images of the retina with high resolution [11]. It has become an essential tool for diagnosing and monitoring the development of various retinal pathologies [12], [13], [14]. Spectral domain OCT (SD-OCT) is the most readily available OCT system for retinal imaging and has a faster scan speed and higher axial resolution than time domain OCT. Currently, most of the commercially available SD-OCT instruments have a resolution of approximately 5 µm [15], [16]. Ultra-high resolution OCT (UHR-OCT), with an axial resolution of approximately 3 µm or less, has the ability to image retinal ultrastructure [17], [18], [19]. Because the segmentation software of most commercial systems is limited to measuring the thicknesses of only a few layers, such as the total retina and the RNFL, several computer automated algorithms for segmenting intra-retinal layers have been proposed to quantitatively evaluate the thickness of more layers that can be imaged with advanced SD-OCT imaging techniques [16]. The use of automated algorithms has undoubtedly enhanced the quantitative diagnosis of ophthalmic disease. To our best knowledge, however, there are few studies that test the reliability of the intra-retinal layer thickness measurements determined by automated segmentation of OCT images. Knowing the level of reliability of such measurements is very important for clinical applications. Thus, the goal of this study was to investigate the repeatability and reproducibility of thickness measurements determined by an automated segmentation algorithm applied to images of eight intra-retinal layers acquired by a custom-built UHR-OCT instrument and a commercially available RTVue100 OCT (Optovue, Fremont, CA, USA) instrument.

Materials and Methods

Subjects

This study followed the tenets of the Declaration of Helsinki and was approved by Ethics Committee of Wenzhou Medical University. Twenty normal subjects (12 males and 8 females, mean ± standard deviation age: 25.1±2.0 years, range: 24 to 32 years) were included, and each signed an informed consent. The inclusion criteria were as follows: no history of ocular or systemic disease, 20/20 or better visual acuity, range of refractive error between −2.00 diopter (D) and +0.50 D, no history of intraocular pressure higher than 21 mmHg, and a normal appearance of the macula.

Instruments and Image Acquisition

Retinal OCT imaging was performed with two SD-OCT instruments configured as shown in Table 1. Briefly, the UHR-OCT used a superluminescent diode (SLD: T840; SuperLum Diodes Ltd., Moscow, Russia) [12]. For imaging the posterior segment of the eye, it was adapted onto a slit-lamp system with the installation of an ocular lens (60 D; Volk Optical, Mentor, OH, USA) on the sample arm. The field of the scan was set to approximately 15° to 20°. The power of the incident light was set to 750 µW, which is well below the safety standard, according to the American National Standards Institute (ANSI Z136.1-2000). The calibrated scan depth was 1.48 mm in air. To calibrate the scan width for the retinal imaging, a model eye (OEMI-7, Ocular Instruments, Bellevue, WA, USA), with a grid implanted on the fundus, was used. Each lattice was 1 mm in width. A horizontal B-scan was performed crossing the grid, and the pixel numbers corresponding to one lattice were acquired.
Table 1

Configuration of the two OCT instruments.

Technical detailsUHR-OCTRTVue 100 OCT
Axial Resolution3 µm5 µm
Scan Speed24,000 Scans/s26,000 Scans/s
Center wavelength840 nm840 nm
Band width100 nm50 nm
Scan width8 mm8 mm
Image Size of each B-scan1,365 pixels (depth)×2,048 pixels (width)640 pixels(depth)×960 pixels (width)

Procedure

Before imaging, all eyes received an ocular examination including visual acuity testing, autorefraction, intraocular pressure, and ophthalmoscopic examination. After enrollment, all eyes were imaged without mydriasis. Two repeated measurements were performed in a short time on the same day by a single examiner (XL) using both SD-OCT instruments to test the intra-observer repeatability. The OCT imaging was also performed one time by another examiner (LL) using each OCT instrument on the same day to test inter-observer reproducibility. The order of OCT examinations was chosen randomly for each patient. During OCT imaging, the subjects were asked to move their head away from the headrest after each image acquisition, and after five minutes they were asked to reposition their head for the following measurement [20]. One author (XL), who did not know which images were taken by which observer or from which subjects, processed the images.

Measurements of Macular Thickness of Intra-retinal Layers

Custom software for automatic segmentation was developed to measure the thicknesses of eight intra-retinal layers on 2-D images produced by each OCT instrument. The image segmentation algorithm mainly employed graph theory and shortest-path search based on an optimization algorithm of dynamic programming technique as described in a previous study [16]. Nine boundaries of the intra-retinal layer structures were detected (Fig. 1) including (1) internal limiting membrane (ILM); (2) nerve fiber layer/ganglion cell layer (NFL/GCL); (3) inner plexiform layer/inner nuclear layer (IPL/INL); (4) inner nuclear layer/outer plexiform layer (INL/OPL); (5) outer plexiform layer/outer nuclear layer (OPL/ONL); (6) external limiting membrane (ELM); (7) inner segment/outer segment of receptors (IS/OS); (8) outer segment/retina pigment epithelium (OS/RPE); (9) retina pigment epithelium/choroid (RPE/choroid). Each of these nine boundaries was detected sequentially by a two- step segmentation procedure. First, a graph based on node cost assignments was built. The node costs are mainly based on the intensity gradient values along the vertical direction and other features, such as the edge direction, which depended upon the boundary of interest. Second, the layer boundary was extracted by a shortest path search applied to the graph using a dynamic programming algorithm. Then, all boundaries detected were overlaid on the OCT images and were verified by visual inspection performed by one of the authors (XL). A semi-automated approach was implemented in the algorithm to correct the segmentation errors that occurred in regions which had extremely low reflectivity or almost no structural information [21], [22].
Figure 1

Boundaries of intra-retinal layers in OCT macular images.

As seen in this image taken by UHR-OCT in the horizontal meridian, nine boundaries of intra-retinal layers were visualized. Images taken in the vertical meridian by UHR-OCT and in both meridians by the RTVue100 were similar to this.

Boundaries of intra-retinal layers in OCT macular images.

As seen in this image taken by UHR-OCT in the horizontal meridian, nine boundaries of intra-retinal layers were visualized. Images taken in the vertical meridian by UHR-OCT and in both meridians by the RTVue100 were similar to this. Figure 2 illustrates the detailed sequence in the boundary segmentation process. Each OCT image was first pre-processed to reduce the background noise using median and Gaussian filtering techniques. This step helped to improve the performance of the segmentation algorithm. The ILM and the boundary between the RPE and choroid layers were first segmented so that other boundaries could be segmented in a limited search region to save computation time. The ILM was defined as the first highly reflective increase from the inner side of the retinal image. It was most often well demarcated, easily detected, and followed by a sector of high reflectivity. Based on these features, the initialized boundary was determined by the first peak on each sampling line from the inner side of the retinal structure. Then the boundary was refined by finding the shortest path in a limited region based on the initialized boundary. The RPE layer was located on the outermost side of the retina and was one of the most hyper-reflective layers within each retinal OCT image. Thus, we searched for the brightest pixel in each A-scan line below the ILM layer on the pre-processed image as an estimated boundary between the RPE and choroid layers. Then the shortest path search was applied on the graph to refine these two boundaries based on their estimated layer locations. Once these two boundaries were segmented, the process for other boundaries was repeated recursively by limiting the search space based on the previous segmentation to detect a new layer boundary.
Figure 2

The detailed sequence in the boundary segmentation process.

(a) Original image. (b) Image smoothing. (c) Gradient image. (d) The ILM and the boundary between the RPE and choroid layers were first segmented. (e) Limiting detection area and search the minimum-weighted path. (f) Segmented image.

The detailed sequence in the boundary segmentation process.

(a) Original image. (b) Image smoothing. (c) Gradient image. (d) The ILM and the boundary between the RPE and choroid layers were first segmented. (e) Limiting detection area and search the minimum-weighted path. (f) Segmented image.

Statistical Analysis

All statistical analysis was performed with the Statistical Package for the Social Sciences (SPSS) software (v17.0 for Windows; SPSS Inc., Chicago, IL, USA). Descriptive statistics were determined as means ± standard deviations. The intra-observer repeatability was measured with two OCT images obtained by the same operator, and the inter-observer reproducibility was measured with two OCT images obtained by two different operators. The overall mean thickness, the coefficients of repeatability and reproducibility (COR), and intraclass correlation coefficients (ICC) were calculated to evaluate the repeatability and reproducibility of the thickness measurements. The overall mean thicknesses of the eight intra-retinal layers along the central macular 6-mm scan length were determined as the average of the first and second measurement by the same examiner. The COR was defined as the standard deviation of differences between the two measurements divided by the mean value of two different measurements. The ICC was determined based on a mixed-model analysis of variance proposed by Bartko and Carpender [23]. The paired t-test, ICC, 95% limits of agreement (LoA), and Bland and Altman plots [24] were analyzed to evaluate the agreement of thickness measurements between the two SD-OCT instruments. The 95% LoA was defined as the mean difference ±1.96 standard deviation [25]. P-values <0.05 was considered statistically significant.

Results

The automated algorithm successfully segmented eight intra-retinal layers in the macular images obtained by the UHR-OCT and RTVue100 instruments. The algorithm also determined the thickness profiles of each layer along the 6-mm horizontal (Fig. 3) and vertical (Fig. 4) scans obtained by each instrument. There were errors in the segmentation boundary for a few images of lower quality. For example, the algorithm mistakenly identified the OPL/ONL interface in Figure 5A. A similar failure occurred for the RNFL/GCL boundary in Figure 5C. The semi-automated approach successfully corrected the segmentation errors (Fig. 5B and 5D). Visual inspection confirmed that the boundary detection for the eight intra-retinal layers was valid in all images acquired by the two OCT instruments.
Figure 3.Thickness

profiles of eight intra-retinal layers determined from the UHR-OCT and RTVue100 images in the horizontal meridian.

Thickness profiles of eight intra-retinal layers along the horizontal meridian were averaged for 20 normal healthy eyes. Error bars represent standard deviation.

Figure 4

Thickness profiles of eight intra-retinal layers determined from the UHR-OCT and RTVue100 images in the vertical meridian.

Thickness profiles of eight intra-retinal layers along the vertical meridian were averaged for 20 normal healthy eyes. Error bars represent standard deviation.

Figure 5

Segmentation errors in scans of lower image quality and corresponding corrected segmentation after applying the semi-automated approach.

(A) The algorithm mistakenly identified the OPL/ONL interface. (B) Corrected segmentation corresponding (A) after applying the semi-automated approach. (C) The algorithm mistakenly identified the RNFL/GCL boundary. (D) Corrected segmentation corresponding (C) after applying the semi-automated approach.

profiles of eight intra-retinal layers determined from the UHR-OCT and RTVue100 images in the horizontal meridian.

Thickness profiles of eight intra-retinal layers along the horizontal meridian were averaged for 20 normal healthy eyes. Error bars represent standard deviation.

Thickness profiles of eight intra-retinal layers determined from the UHR-OCT and RTVue100 images in the vertical meridian.

Thickness profiles of eight intra-retinal layers along the vertical meridian were averaged for 20 normal healthy eyes. Error bars represent standard deviation.

Segmentation errors in scans of lower image quality and corresponding corrected segmentation after applying the semi-automated approach.

(A) The algorithm mistakenly identified the OPL/ONL interface. (B) Corrected segmentation corresponding (A) after applying the semi-automated approach. (C) The algorithm mistakenly identified the RNFL/GCL boundary. (D) Corrected segmentation corresponding (C) after applying the semi-automated approach. For the twenty normal eyes, the overall mean total retinal thickness measured with UHR-OCT was 292.17±14.70 µm and 293.83±14.96 µm in horizontal and vertical meridians, respectively. The total retinal thickness measured with RTVue was 305.70±15.73 µm and 309.03±15.07 µm in horizontal and vertical meridians, respectively. The overall mean intra-retinal layers thicknesses ranged from 19.85 to 67.18 µm with UHR-OCT and 20.88 to 72.03 µm with the RTVue100. Table 2 and Table 3 show the repeatability and reproducibility of thickness measurements for intra-retinal layers measured with UHR-OCT and RTVue100, respectively. There were no significant differences between the two thickness measurements for either instrument obtained by the same examiner. The ICCs obtained for the intra-observer repeatability and inter-observer reproducibility tests of both SD-OCT instruments were greater than 0.945 for the total retina and all intra-retinal layers except the IS layer, which ranged from 0.051 to 0.643, and the OS layer, which ranged from 0.709 to 0.959. The coefficients of repeatability and reproducibility were less than 6.73% for the total retina and all intra-retinal layers.
Table 2

Repeatability and reproducibility of thickness measurements for intra-retinal layers measured by UHR-OCT.

T1 (µm)T2 (µm)T3 (µm)T1–T2 (µm)T1–T3 (µm)ICCa ICCb CORa (%)CORb (%)
Horizontal Meridian
RNFL 22.72±2.6822.68±2.7922.82±2.640.04±0.96−0.10±1.071.0001.0002.072.01
GCL+IPL 67.36±5.8467.00±5.8168.24±8.040.36±1.01−0.88±5.301.0000.9991.432.59
INL 31.79±2.8432.11±2.6730.84±5.07−0.32±1.180.95±4.530.9970.9953.203.80
OPL 19.91±1.1319.78±1.1219.90±1.000.13±1.420.01±0.910.9810.9844.874.49
ONL 61.71±9.1760.49±10.2459.56±10.451.23±6.412.15±6.460.9980.9951.981.85
IS 27.57±3.5028.29±5.5829.36±6.08−0.72±5.89−1.79±6.650.3000.1580.940.77
OS 32.05±2.3532.21±2.1132.61±1.78−0.17±2.07−0.57±2.170.9860.9642.052.85
RPE 29.06±2.5029.59±1.9029.28±1.82−0.53±2.25−0.22±1.930.9940.9901.783.43
Total 292.18±14.93292.16±14.56292.62±14.850.02±2.28−0.44±2.041.0001.0000.360.50
Vertical Meridian
RNFL 32.57±4.6132.85±4.4133.10±4.75−0.28±1.25−0.53±1.410.9980.9982.973.21
GCL+IPL 62.56±6.0262.27±6.8962.13±5.670.29±2.980.43±1.570.9980.9993.512.04
INL 31.49±3.0931.43±2.3731.87±3.100.06±2.18−0.38±2.550.9950.9944.224.14
OPL 20.43±1.4520.84±1.1420.49±0.88−0.41±1.33−0.06±0.950.9450.9695.044.09
ONL 52.59±10.6851.04±12.2451.72±11.411.55±6.070.87±5.170.9980.9982.122.90
IS 32.22±6.7733.60±7.6032.46±7.68−1.38±5.70−0.23±4.250.3760.6431.902.08
OS 33.39±2.3733.60±2.7233.61±2.59−0.21±1.28−0.22±1.670.9350.9593.062.25
RPE 28.33±2.2328.44±2.5328.52±1.89−0.11±1.02−0.19±1.470.9880.9893.212.86
Total 293.58±14.92294.08±15.03293.90±14.79−0.50±1.31−0.31±1.231.0001.0000.400.46

T1: mean thickness for the first measurement by examiner 1; T2: mean thickness for the second measurement by examiner 1; T3: mean thickness for the first measurement by examiner 2; ICCa: intraclass correlation coefficients of repeatability; ICCb: intraclass correlation coefficients of reproducibility; CORa: coefficients of repeatability; CORb:coefficients of reproducibility; n = 20 eyes.

Table 3

Repeatability and reproducibility of thickness measurements for eight intra-retinal layers measured by the RTVue100.

T1 (µm)T2 (µm)T3 (µm)T1–T2 (µm)T1–T3 (µm)ICCa ICCb CORa(%)CORb(%)
Horizontal Meridian
RNFL 22.42±1.7222.29±1.7622.14±1.670.13±0.530.27±0.591.0000.9991.842.72
GCL+IPL 71.89±6.4672.16±6.5471.76±6.56−0.27±0.820.14±1.000.9990.9991.742.15
INL 33.88±2.8933.75±3.2434.04±2.990.12±0.93−0.16±1.160.9980.9962.663.49
OPL 21.08±1.6320.68±1.6020.37±1.100.40±1.75−0.70±1.590.9900.9794.286.73
ONL 71.44±7.9172.01±7.3271.67±6.68−0.58±2.73−0.23±2.490.9970.9941.722.53
IS 21.16±1.3921.04±1.4321.46±1.000.12±0.70−0.30±1.030.3940.0861.591.52
OS 35.43±1.9034.43±2.4534.64±2.020.99±1.890.79±1.880.8890.9351.932.24
RPE 28.64±2.0629.12±1.9829.10±1.89−0.48±1.43−0.46±1.340.9870.9712.392.60
Total 305.92±15.93305.48±15.56305.18±15.950.44±1.280.74±1.671.0000.9990.430.52
Vertical Meridian
RNFL 32.75±2.9232.64±2.7632.75±2.920.11±2.080.00±2.020.9990.9993.541.71
GCL+IPL 69.84±5.7069.57±5.4269.59±5.770.27±1.150.25±1.140.9990.9991.671.63
INL 34.55±3.1934.52±2.8834.48±2.830.02±1.070.06±0.900.9980.9982.812.49
OPL 21.65±2.3120.73±1.1921.03±1.500.92±1.980.61±1.840.9790.9895.384.93
ONL 65.19±9.5467.05±6.9566.86±6.36−1.86±4.06−1.67±4.490.9890.9892.962.29
IS 21.66±1.7021.39±1.2721.33±1.500.26±1.560.32±2.040.1020.0511.591.76
OS 35.35±3.3134.11±2.7434.80±2.781.24±2.690.55±2.390.7090.7155.214.78
RPE 28.29±2.8028.78±2.4028.39±2.51−0.49±1.91−0.11±2.290.9570.9715.784.42
Total 309.26±15.47308.79±14.71309.25±15.020.47±1.820.01±1.010.9990.9990.560.65

T1: mean thickness for the first measurement by examiner 1; T2: mean thickness for the second measurement by examiner 1; T3: mean thickness for the first measurement by examiner 2; ICCa: intraclass correlation coefficients of repeatability; ICCb: intraclass correlation coefficients of reproducibility; CORa: coefficients of repeatability; CORb: coefficients of reproducibility; n = 20 eyes.

T1: mean thickness for the first measurement by examiner 1; T2: mean thickness for the second measurement by examiner 1; T3: mean thickness for the first measurement by examiner 2; ICCa: intraclass correlation coefficients of repeatability; ICCb: intraclass correlation coefficients of reproducibility; CORa: coefficients of repeatability; CORb:coefficients of reproducibility; n = 20 eyes. T1: mean thickness for the first measurement by examiner 1; T2: mean thickness for the second measurement by examiner 1; T3: mean thickness for the first measurement by examiner 2; ICCa: intraclass correlation coefficients of repeatability; ICCb: intraclass correlation coefficients of reproducibility; CORa: coefficients of repeatability; CORb: coefficients of reproducibility; n = 20 eyes. The overall mean thicknesses of intra-retinal layers measured by UHR-OCT were compared with RTVue100 (Table 4). There were significant differences between the UHR-OCT and RTVue100 measurements of most macular intra-retinal layer thicknesses except the RNFL and RPE. The total retinal thickness measured by the UHR-OCT was significantly thinner than that measured by the RTVue100. The ICC for agreement of the thickness profiles between UHR-OCT and RTVue OCT were greater than 0.80 except for the IS and OS layers. Bland-Altman plots (Fig. 6) were also used to test the agreement of the thicknesses measured with these two SD-OCT devices. The Bland-Altman plots and the 95% LoA results showed good agreement between the two SD-OCT instruments for all of the intra-retinal layers except the ONL and IS.
Table 4

Comparison of the thickness measurements for eight intra-retinal layers and total retina between UHR-OCT and RTVue100 devices.

UHR (µm)RTVue (µm)P valueICC95% LoA (µm)
Horizontal Meridian
RNFL 22.70±2.7622.35±1.770.260.95−2.18 to 2.88
GCL+IPL 67.18±5.9672.03±6.660.000.94−7.89 to 1.79
INL 31.95±2.7633.81±3.110.000.96−4.96 to 1.23
OPL 19.85±0.9020.88±1.390.010.94−3.74 to 1.68
ONL 61.10±9.4271.73±7.700.000.86−20.25 to −1.00
IS 27.93±3.7021.10±1.400.000.01−0.67 to 14.33
OS 32.13±2.0334.93±2.030.000.57−6.04 to 0.44
RPE 29.33±1.9628.88±1.940.340.95−3.39 to 4.29
Total 292.17±15.08305.70±16.140.000.93−20.67 to −6.39
Vertical Meridian
RNFL 32.71±4.5932.69±2.710.960.98−4.89 to 4.93
GCL+IPL 62.42±6.4669.70±5.680.000.92−13.37 to −1.20
INL 31.46±2.5934.54±3.070.000.94−5.81 to −0.34
OPL 20.63±1.1521.19±1.590.020.93−3.59 to −2.48
ONL 51.82±11.3766.12±8.300.000.80−34.20 to 5.59
IS 32.91±6.7821.52±1.310.000.00−2.17 to 24.95
OS 33.49±2.5334.73±2.790.030.66−7.34 to 4.87
RPE 28.39±2.3928.53±2.490.290.96−4.72 to 4.43
Total 293.83±15.35309.03±15.460.000.90−21.24 to −9.15

P value: tested by paired-t test.

Figure 6

Bland-Altman plots of thickness measurements determined with the automated segmentation algorithm on UHR-OCT and RTVue100 images.

Only the images along the horizontal meridian were analyzed. The horizontal full lines represent the mean of thickness differences, and the horizontal dashed lines represent the mean differences ±1.96 standard deviation.

Bland-Altman plots of thickness measurements determined with the automated segmentation algorithm on UHR-OCT and RTVue100 images.

Only the images along the horizontal meridian were analyzed. The horizontal full lines represent the mean of thickness differences, and the horizontal dashed lines represent the mean differences ±1.96 standard deviation. P value: tested by paired-t test.

Discussion

It is very important in the clinical routine to know the repeatability and reproducibility of measurements. This information enables the clinician to evaluate if observed changes are due to fluctuations in the methods or if they are valid changes in the structures. This is especially true for the measurements of the intra-retinal layers, which are important morphometric parameters in the diagnosis of retinal and neurological diseases and the monitoring of the progression of these disorders. In this current study, we applied our automated algorithm to images from a custom-built UHR-OCT instrument and commercially available RTVue100 OCT instrument to yield the thickness profiles of eight intra-retinal layers. The intra-observer and inter-observer test results indicated that both instruments produced highly repeatable and reproducible measurements for most of the intra-retinal layers. The results of inter-instrument comparisons of the thickness measurements suggested that the thicknesses of intra-retinal layers obtained by the methods were not interchangeable between the two different SD-OCT instruments. There are many studies showing the repeatability and reproducibility of the retinal measurements for normal subjects. However due to limitations of the image processing software on the commercial OCT instruments, most of them only focus on the RNFL or the total retinal thickness [15], [20], [26], [27]. The aim of our study was to report the repeatability and reproducibility of thickness measurements for eight intra-retinal layers determined by an automated algorithm applied to images from two different SD-OCT instruments. For both intra- and inter-observer comparisons, the ICCs were high for all layers except the IS. This indicates that, as in previous findings, the thickness measurements for most of the intra-retinal layers were repeatable either in different visits or by different examiners [26], [27], [28]. The repeatability of measurements in the present study is consistent with those in previous reports. Debuc et al. reported the repeatability and reproducibility of thickness measurements for six intra-retinal layers using custom-developed automated software on time domain OCT images [29]. They found ICCs to be greater than 0.75 except for the OPL and OS/RPE. Their results were consistent with ours, and the fuzzy boundaries of the ELM and OS/RPE contributed to the poor repeatability and reproducibility of IS and OS layers. The ELM was not seen clearly in most OCT images and the RPE cells of the villi were embedded into the receptor outer segments so that the OS/RPE boundary was not clear. Garas et al. reported the average and quadrant thicknesses of the RNFL and GCC based on the RTVue100 [20]. The intra-session ICC varied between 93.9% and 99.0%, intra-session CV between 1.95% and 5.69%, and intra-test variability was between 3.11 and 9.13 µm. Their results are comparable to ours. Wang et al reported the repeatability of thickness measurements for nine intra-retinal layers determined by manual segmentation of UHR-OCT images with an axial resolution of about 2 µm [30]. The ICC was >0.90 in most of the intra-retinal layers. In the current study, we found that the repeatability and reproducibility of the UHR-OCT instrument was better than that of the RTVue100. Additionally, the repeatability of the UHR-OCT with an axial resolution below 2 µm [30] was better than that of our UHR-OCT with axial resolution of 3 µm. These results indicate that the axial resolution of OCT may contribute to the repeatability of retinal thickness measurements. This is consistent with Ge et al. [25], who reported that the higher axial resolution OCT instruments have a better repeatability in measurements of central corneal thickness and epithelial thickness. The repeatability of the IS thickness measurements in the current study was not as good as that reported by Wang et al. [30], even though they used manual segmentation in their study. Besides the axial resolution, there are some other factors, such as the image quality and image size that may also contribute to the repeatability. The thickness measurements of most of the intra-retinal layers and the total retina were different between the UHR-OCT and RTVue100. This result was consistent with previously reported studies in which the retinal thickness measurements differed significantly depending upon the OCT systems used. For example, Seibold et al. [31] compared RNFL thickness measurements taken with three different SD-OCT instruments and a time-domain OCT instrument. RNFL thicknesses were significantly different among the four instruments, and they could not be used interchangeably. Similarly, Grover et al. found that the central subfield thickness measured by two different SD-OCT instruments differed by almost 70 µm [32]. Moreover, Wolf-Schnurrbusch et al. [33] compared central retinal thickness measurements in healthy eyes taken by six different commercially available OCT instruments. The six OCT systems each provided different results. Their results imply that the different OCT systems cannot be used interchangeably for the measurement of macular thickness. There are some limitations to the present study. One is the unequal ratio of males to females. Setaro Oeto et al. [34] found the mean thicknesses of the INL and the OPL+ONL were significantly greater in men, and the mean RNFL thickness was greater in women. However, the purpose of our study was to evaluate the reliability of the newly developed segmentation algorithm to measure the thickness profiles of eight intra-retinal layers. Thus, gender is unlikely to influence the outcome of our study. In future studies we will pay attention to gender differences. Another limitation is that it was conducted on normal subjects only. Diseased retinal structures may vary substantially among patients, and this is likely to increase the frequency of segmentation errors. Thus the repeatability and reproducibility values may be reduced in diseased retinas [35], [36]. In future studies, we will apply our new method to a variety of retinal diseases to evaluate the clinical significance of any changes in the repeatability and reproducibility of the intra-retinal measurements. In conclusion, thickness measurements of the intra-retinal layers have good repeatability and reproducibility when determined by the automated algorithm applied to images acquired by the UHR-OCT and RTVue100 instruments. Use of this algorithm will be helpful in the diagnosis of retinal diseases and the evaluation of disease progression.
  36 in total

1.  Broadband superluminescent diode-based ultrahigh resolution optical coherence tomography for ophthalmic imaging.

Authors:  Dexi Zhu; Meixiao Shen; Hong Jiang; Ming Li; Michael R Wang; Yuhong Wang; Lili Ge; Jia Qu; Jianhua Wang
Journal:  J Biomed Opt       Date:  2011-12       Impact factor: 3.170

2.  Quantitative analysis of the intraretinal layers and optic nerve head using ultra-high resolution optical coherence tomography.

Authors:  Yuhong Wang; Hong Jiang; Meixiao Shen; Byron L Lam; Delia Cabrera DeBuc; Yufeng Ye; Ming Li; Aizhu Tao; Yilei Shao; Jianhua Wang
Journal:  J Biomed Opt       Date:  2012-06       Impact factor: 3.170

3.  Comparison of retinal nerve fiber layer thickness in normal eyes using time-domain and spectral-domain optical coherence tomography.

Authors:  Leonard K Seibold; Naresh Mandava; Malik Y Kahook
Journal:  Am J Ophthalmol       Date:  2010-12       Impact factor: 5.258

4.  Macular and retinal nerve fiber layer analysis of normal and glaucomatous eyes in children using optical coherence tomography.

Authors:  Derek B Hess; Sanjay G Asrani; Manisha G Bhide; Laura B Enyedi; Sandra S Stinnett; Sharon F Freedman
Journal:  Am J Ophthalmol       Date:  2005-03       Impact factor: 5.258

5.  Optical coherence tomography grading reproducibility during the Comparison of Age-related Macular Degeneration Treatments Trials.

Authors:  Francis Char DeCroos; Cynthia A Toth; Sandra S Stinnett; Cynthia S Heydary; Russell Burns; Glenn J Jaffe
Journal:  Ophthalmology       Date:  2012-08-28       Impact factor: 12.079

6.  Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments.

Authors:  Ute E K Wolf-Schnurrbusch; Lala Ceklic; Christian K Brinkmann; Milko E Iliev; Manuel Frey; Simon P Rothenbuehler; Volker Enzmann; Sebastian Wolf
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-02-21       Impact factor: 4.799

7.  Comparison of retinal thickness in normal eyes using Stratus and Spectralis optical coherence tomography.

Authors:  Sandeep Grover; Ravi K Murthy; Vikram S Brar; Kakarla V Chalam
Journal:  Invest Ophthalmol Vis Sci       Date:  2009-12-10       Impact factor: 4.799

8.  Relationship between retinal layer thickness and the visual field in early age-related macular degeneration.

Authors:  Jennifer H Acton; R Theodore Smith; Donald C Hood; Vivienne C Greenstein
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-11-09       Impact factor: 4.799

9.  Microcystic macular oedema, thickness of the inner nuclear layer of the retina, and disease characteristics in multiple sclerosis: a retrospective study.

Authors:  Shiv Saidha; Elias S Sotirchos; Mohamed A Ibrahim; Ciprian M Crainiceanu; Jeffrey M Gelfand; Yasir J Sepah; John N Ratchford; Jiwon Oh; Michaela A Seigo; Scott D Newsome; Laura J Balcer; Elliot M Frohman; Ari J Green; Quan D Nguyen; Peter A Calabresi
Journal:  Lancet Neurol       Date:  2012-10-04       Impact factor: 44.182

10.  Retinal neurodegeneration in Wilson's disease revealed by spectral domain optical coherence tomography.

Authors:  Philipp Albrecht; Ann-Kristin Müller; Marius Ringelstein; David Finis; Gerd Geerling; Eva Cohn; Orhan Aktas; Hans-Peter Hartung; Harald Hefter; Axel Methner
Journal:  PLoS One       Date:  2012-11-16       Impact factor: 3.240

View more
  21 in total

1.  RefMoB, a Reflectivity Feature Model-Based Automated Method for Measuring Four Outer Retinal Hyperreflective Bands in Optical Coherence Tomography.

Authors:  Douglas H Ross; Mark E Clark; Pooja Godara; Carrie Huisingh; Gerald McGwin; Cynthia Owsley; Katie M Litts; Richard F Spaide; Kenneth R Sloan; Christine A Curcio
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-07       Impact factor: 4.799

2.  Short term effect of choroid thickness in the horizontal meridian detected by spectral domain optical coherence tomography in myopic children after orthokeratology.

Authors:  Wan-Qing Jin; Sheng-Hai Huang; Jun Jiang; Xin-Jie Mao; Mei-Xiao Shen; Yan Lian
Journal:  Int J Ophthalmol       Date:  2018-06-18       Impact factor: 1.779

3.  Associations of macular thickness in spectral-domain OCT with ocular and systemic cardiovascular parameters - the MIPH Eye & Health Study.

Authors:  Alexander Karl-Georg Schuster; Joachim Ernst Fischer; Urs Vossmerbaeumer
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-09-26       Impact factor: 3.117

4.  In vivo imaging of retinal hemodynamics with OCT angiography and Doppler OCT.

Authors:  Shenghai Huang; Meixiao Shen; Dexi Zhu; Qi Chen; Ce Shi; Zhongping Chen; Fan Lu
Journal:  Biomed Opt Express       Date:  2016-01-25       Impact factor: 3.732

5.  Longitudinal Macular Structure-Function Relationships in Glaucoma and Their Sources of Variability.

Authors:  Kouros Nouri-Mahdavi; Nima Fatehi; Joseph Caprioli
Journal:  Am J Ophthalmol       Date:  2019-05-10       Impact factor: 5.258

6.  Computer-aided analyses of mouse retinal OCT images - an actual application report.

Authors:  Dekuang Yu; Jin Zheng; Ruilin Zhu; Nan Wu; Alex Guan; Kin-Sang Cho; Dong Feng Chen; Gang Luo
Journal:  Ophthalmic Physiol Opt       Date:  2015-05-10       Impact factor: 3.117

Review 7.  Macular imaging with optical coherence tomography in glaucoma.

Authors:  Vahid Mohammadzadeh; Nima Fatehi; Adeleh Yarmohammadi; Ji Woong Lee; Farideh Sharifipour; Ramin Daneshvar; Joseph Caprioli; Kouros Nouri-Mahdavi
Journal:  Surv Ophthalmol       Date:  2020-03-19       Impact factor: 6.048

8.  Analysis of Agreement of Retinal-Layer Thickness Measures Derived from the Segmentation of Horizontal and Vertical Spectralis OCT Macular Scans.

Authors:  Natalia Gonzalez Caldito; Bhavna Antony; Yufan He; Andrew Lang; James Nguyen; Alissa Rothman; Esther Ogbuokiri; Ama Avornu; Laura Balcer; Elliot Frohman; Teresa C Frohman; Pavan Bhargava; Jerry Prince; Peter A Calabresi; Shiv Saidha
Journal:  Curr Eye Res       Date:  2017-12-14       Impact factor: 2.424

9.  Validation of Optical Coherence Tomography Retinal Segmentation in Neurodegenerative Disease.

Authors:  Bryan M Wong; Richard W Cheng; Efrem D Mandelcorn; Edward Margolin; Sherif El-Defrawy; Peng Yan; Anna T Santiago; Elena Leontieva; Wendy Lou; Wendy Hatch; Christopher Hudson
Journal:  Transl Vis Sci Technol       Date:  2019-09-11       Impact factor: 3.283

10.  The use of handheld spectral domain optical coherence tomography in pediatric ophthalmology practice: Our experience of 975 infants and children.

Authors:  Ashwin Mallipatna; Anand Vinekar; Chaitra Jayadev; Supriya Dabir; Munsusamy Sivakumar; Narasimha Krishnan; Pooja Mehta; Tos Berendschot; Naresh Kumar Yadav
Journal:  Indian J Ophthalmol       Date:  2015-07       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.