Literature DB >> 24499740

Collimator Interchange System for Adaptive Cardiac Imaging in C-SPECT.

Mike Rozler1, Wei Chang1.   

Abstract

Compared to imaging the heart with conventional cameras, dedicated cardiac SPECT systems can achieve much higher performance through use of a small field of view. To realize this potential, however, the heart must be reliably placed in the appropriate small FOV prior to imaging, thus requiring a separate scout operation to locate the heart and estimate its size. Further-more, to achieve high performance across the general population, a system should provide several imaging configurations optimized for different size and location of the heart and the size of the patient. Because of the critical role the collimator plays in SPECT, it would be ideal if a dedicated collimator could be used for each of the different patient groups, as well as for the scout imaging. The ability to exchange collimators without moving the patient can also enable serial studies with different imaging options while preserving anatomic registration. We developed a slit exchange system for the slit-slat collimator of the C-SPECT cardiac platform. The full-scale prototype, a precision link conveyor following a curved, body contouring path, allows four distinct transaxial collimation options. The collimators can be exchanged in 10 seconds without disturbing the patient, thus allowing adaptive clinical SPECT imaging. The positioning precision for all elements of the system is within 0.1 mm and has shown no degradation over 100,000 complete revolutions of the conveyor-twice the expected usage for a clinical system. We consider the rapid and precise operation allowing optimal collimation for different imaging tasks to be an important technological step for cardiac SPECT.

Entities:  

Keywords:  Adaptive systems; biomedical imaging; collimators; single photon emission computed tomography

Year:  2010        PMID: 24499740      PMCID: PMC3830537          DOI: 10.1109/NSSMIC.2010.5874138

Source DB:  PubMed          Journal:  IEEE Trans Nucl Sci        ISSN: 0018-9499            Impact factor:   1.679


  7 in total

1.  American Society of Nuclear Cardiology and Society of Nuclear Medicine joint position statement: attenuation correction of myocardial perfusion SPECT scintigraphy.

Authors:  Gary V Heller; Jonathan Links; Timothy M Bateman; Jack A Ziffer; Edward Ficaro; Mylan C Cohen; Robert C Hendel
Journal:  J Nucl Cardiol       Date:  2004 Mar-Apr       Impact factor: 5.952

2.  The potential for mixed multiplexed and non-multiplexed data to improve the reconstruction quality of a multi-slit-slat collimator SPECT system.

Authors:  Shelan T Mahmood; Kjell Erlandsson; Ian Cullum; Brian F Hutton
Journal:  Phys Med Biol       Date:  2010-03-30       Impact factor: 3.609

3.  C-SPECT - a Clinical Cardiac SPECT/Tct Platform: Design Concepts and Performance Potential.

Authors:  Wei Chang; Caesar E Ordonez; Haoning Liang; Yusheng Li; Jingai Liu
Journal:  IEEE Trans Nucl Sci       Date:  2009-10-06       Impact factor: 1.679

4.  The importance and implementation of accurate 3D compensation methods for quantitative SPECT.

Authors:  B M Tsui; E C Frey; X Zhao; D S Lalush; R E Johnston; W H McCartney
Journal:  Phys Med Biol       Date:  1994-03       Impact factor: 3.609

5.  A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera.

Authors:  Sanjiv S Gambhir; Daniel S Berman; Jack Ziffer; Michael Nagler; Martin Sandler; Jim Patton; Brian Hutton; Tali Sharir; Shlomo Ben Haim; Simona Ben Haim
Journal:  J Nucl Med       Date:  2009-04       Impact factor: 10.057

6.  Adaptive SPECT.

Authors:  Harrison H Barrett; Lars R Furenlid; Melanie Freed; Jacob Y Hesterman; Matthew A Kupinski; Eric Clarkson; Meredith K Whitaker
Journal:  IEEE Trans Med Imaging       Date:  2008-06       Impact factor: 10.048

7.  Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras.

Authors:  Fabio P Esteves; Paolo Raggi; Russell D Folks; Zohar Keidar; J Wells Askew; Shmuel Rispler; Michael K O'Connor; Liudmilla Verdes; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2009-08-18       Impact factor: 5.952

  7 in total
  2 in total

Review 1.  Review of SPECT collimator selection, optimization, and fabrication for clinical and preclinical imaging.

Authors:  Karen Van Audenhaege; Roel Van Holen; Stefaan Vandenberghe; Christian Vanhove; Scott D Metzler; Stephen C Moore
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

Review 2.  Recent advances in cardiac SPECT instrumentation and system design.

Authors:  Mark F Smith
Journal:  Curr Cardiol Rep       Date:  2013-08       Impact factor: 2.931

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.