Literature DB >> 24474088

The effect of a collar and surface finish on cemented femoral stems: a prospective randomised trial of four stem designs.

Jonathan Hutt1, Alexandra Hazlerigg, Ansari Aneel, Geoffrey Epie, Husam Dabis, Roy Twyman, Andrew Cobb.   

Abstract

PURPOSE: The optimal design for a cemented femoral stem remains a matter of debate. Over time, the shape, surface finish and collar have all been modified in various ways. A clear consensus has not yet emerged regarding the relative merits of even the most basic design features of the stem. We undertook a prospective randomised trial comparing surface finish and the effect of a collar on cemented femoral component subsidence, survivorship and clinical function.
METHODS: One hundred and sixty three primary total hip replacement patients were recruited prospectively and randomised to one of four groups to receive a cemented femoral stem with either a matt or polished finish, and with or without a collar.
RESULTS: At two years, although there was a trend for increased subsidence in the matt collarless group, this was not statistically significant (p = 0.18). At a mean of 10.1 years follow-up, WOMAC scores for the surviving implants were good, (Range of means 89-93) without significant differences. Using revision or radiographic loosening as the endpoint, survivorship of the entire cohort was 93 % at 11 yrs, (CI 87-97 %). There were no significant differences in survivorship between the two groups with polished stems or the two groups with matt stems. A comparison of the two collarless stems demonstrated a statistically significant difference in survivorship between polished (100 %) and matt (88 %) finishes (p = 0.02).
CONCLUSIONS: In the presence of a collar, surface finish did not significantly affect survivorship or function. Between the two collarless groups a polished surface conferred an improved survivorship.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24474088      PMCID: PMC4037524          DOI: 10.1007/s00264-013-2256-z

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  30 in total

1.  A rough surface finish adversely affects the survivorship of a cemented femoral stem.

Authors:  Alejandro González Della Valle; Adriana Zoppi; Margaret G E Peterson; Eduardo A Salvati
Journal:  Clin Orthop Relat Res       Date:  2005-07       Impact factor: 4.176

2.  Mid-term results of Exeter vs Endurance cemented stems.

Authors:  Justin J Sherfey; Richard W McCalden
Journal:  J Arthroplasty       Date:  2006-12       Impact factor: 4.757

Review 3.  The design features of cemented femoral hip implants.

Authors:  T Scheerlinck; P-P Casteleyn
Journal:  J Bone Joint Surg Br       Date:  2006-11

4.  Early loosening of the femoral component at the cement-prosthesis interface after total hip replacement.

Authors:  C G Mohler; J J Callaghan; D K Collis; R C Johnston
Journal:  J Bone Joint Surg Am       Date:  1995-09       Impact factor: 5.284

5.  Loosening of matt and polished cemented femoral stems.

Authors:  D W Howie; R G Middleton; K Costi
Journal:  J Bone Joint Surg Br       Date:  1998-07

6.  High failure rate of a modern, proximally roughened, cemented stem for total hip arthroplasty.

Authors:  A Grose; A González Della Valle; P Bullough; S Lyman; I Tomek; P Pellicci
Journal:  Int Orthop       Date:  2006-04-29       Impact factor: 3.075

7.  Favourable mid-term results of the VerSys CT polished cemented femoral stem for total hip arthroplasty.

Authors:  Alejandro González Della Valle; Fernando Comba; Adriana Zoppi; Eduardo A Salvati
Journal:  Int Orthop       Date:  2006-03-31       Impact factor: 3.075

8.  Ten-year survival of the MS-30 matt-surfaced cemented stem.

Authors:  B J Berli; D Schäfer; E W Morscher
Journal:  J Bone Joint Surg Br       Date:  2005-07

9.  Radiographic failure patterns of polished cemented stems.

Authors:  Brian A Jewett; Dennis K Collis
Journal:  Clin Orthop Relat Res       Date:  2006-12       Impact factor: 4.176

10.  A comparison of collared and collarless femoral components in primary cemented total hip arthroplasty: a randomized clinical trial.

Authors:  J B Meding; M A Ritter; E M Keating; P M Faris; K Edmondson
Journal:  J Arthroplasty       Date:  1999-02       Impact factor: 4.757

View more
  3 in total

1.  Effect of a collar on subsidence and local micromotion of cementless femoral stems: in vitro comparative study based on micro-computerised tomography.

Authors:  Valérie Malfroy Camine; Hannes A Rüdiger; Dominique P Pioletti; Alexandre Terrier
Journal:  Int Orthop       Date:  2017-06-07       Impact factor: 3.075

2.  Increased risk for postoperative periprosthetic fracture in hip fracture patients with the Exeter stem than the anatomic SP2 Lubinus stem.

Authors:  Carl Mellner; Jabbar Mohammed; Magnus Larsson; Sandra Esberg; Maciej Szymanski; Nils Hellström; Cecilia Chang; Hans E Berg; Olof Sköldenberg; Björn Knutsson; Per Morberg; Sebastian Mukka
Journal:  Eur J Trauma Emerg Surg       Date:  2019-11-18       Impact factor: 3.693

3.  Survival and performance of a dual tapered-wedge fully HA-coated press fit femoral stem.

Authors:  Ramakanth R Yakkanti; Dylan N Greif; Dennis J Vanden Berge; Raymond P Robinson
Journal:  Arch Orthop Trauma Surg       Date:  2022-03-03       Impact factor: 3.067

  3 in total