| Literature DB >> 24454522 |
T Juang1, J Newton1, S Das1, J Adamovics2, M Oldham1.
Abstract
Deformable 3D dosimeters have potential applications in validating deformable dose mapping algorithms. This study evaluates a novel deformable PRESAGE® dosimeter and its application toward validating the deformable algorithm employed by VelocityAI. The deformable PRESAGE® dosimeter exhibited a linear dose response with a sensitivity of 0.0032 ΔOD/(Gy/cm). Comparison of an experimental dosimeter irradiated with an MLC pencilbeam checkerboard pattern under lateral compression up to 27% to a non-deformed control dosimeter irradiated with the same pattern verified dose tracking under deformation. CTs of the experimental dosimeter prior to and during compression were exported into VelocityAI and used to map an Eclipse dose distribution calculated on the compressed dosimeter to its original shape. A comparison between the VelocityAI dose distribution and the distribution from the dosimeter showed field displacements up to 7.3 mm and up to a 175% difference in field dimensions. These results highlight the need for validating deformable dose mapping algorithms to ensure patient safety and quality of care.Entities:
Year: 2013 PMID: 24454522 PMCID: PMC3894117 DOI: 10.1088/1742-6596/444/1/012080
Source DB: PubMed Journal: J Phys Conf Ser ISSN: 1742-6588
Figure 16 cm diameter cylindrical deformable dosimeters with attached plates keyed for directional registration in the optical-CT scanner.
Figure 3Geometric dose tracking in the deformed dosimeter is quantified with an overlaid grid. Edge artifacts from dosimeter optical-CT images (B-C) have been cropped. (A) Diagram illustrating the MLC checkerboard pattern. 5 mm×5 mm radiation fields are shown in blue. (B) Axial cross-section of an uncompressed dosimeter irradiated with the pattern in (A). (C) Axial cross-section of a dosimeter laterally compressed by 27% during irradiation with the same pattern.
Figure 2(A) CT of 6 cm diameter dosimeter without compression. (B) CT of the same dosimeter with compression up to 1.6 cm (27%). Arrows show direction of compression.
Figure 4(A) Measured deformed dose distribution - axial cross-section of dosimeter laterally compressed by 27% during irradiation with the pattern shown in Figure 3A. (B) Corresponding axial cross-section showing the deformed dose distribution as predicted by VelocityAI deformable algorithm. (C) Quiver plot displaying magnitude of centroid displacement errors for each field in the VelocityAI distribution. White pixels mark the true location of centroids of each field in the dosimeter (i.e. positions in A).
Differences between the dose distributions from the Velocity AI deformable dose tracking algorithm and the physical dose tracking data from the deformable dosimeter.
| Field | Centroid | Horizontal FWHM | Vertical FWHM | ||
|---|---|---|---|---|---|
| Difference | Percent Error | Difference | Percent Error | ||
| 9 | 5.8 | −4 | −40.0% | 0 | 0.0% |
| 13 | 2.2 | 14 | 175.0% | 2 | 50.0% |
| 17 | 2.2 | 1 | 12.5% | −1 | −20.0% |
| 18 | 7.3 | 10 | 125.0% | −2 | −33.3% |
| 21 | 7.0 | 2 | 22.2% | 1 | 33.3% |
Field numbers refer to enumerated irradiated MLC fields as labelled in Figure 3D.
Magnitude of displacement between the centroid of a given field in the Velocity distribution versus the corresponding field in the dosimeter.
FWHMVelocity – FWHMdosimeter
(FWHMVelocity – FWHMdosimeter)/FWHMdosimeter