A novel technique to enable experimental validation of deformable dose accumulation.

Carolyn J Niu1, Warren D Foltz, Michael Velec, Joanne L Moseley, Adil Al-Mayah, Kristy K Brock.   

Abstract

PURPOSE: To propose a novel technique to experimentally validate deformable dose algorithms by measuring 3D dose distributions under the condition of deformation using deformable gel dosimeters produced by a novel gel fabrication method.
METHOD: Five gel dosimeters, two rigid control gels and three deformable gels, were manufactured and treated with the same conformal plan that prescribed 400 cGy to the isocenter. The control gels were treated statically; the deformable gels were treated while being compressed by an actuation device to simulate breathing motion (amplitude of compression  = 1, 1.5, and 2 cm, respectively; frequency  =  16 rpm). Comparison between the dose measured by the control gels and the corresponding static dose distribution calculated in the treatment planning system (TPS) has determined the intrinsic dose measurement uncertainty of the gel dosimeters. Doses accumulated using MORFEUS, a biomechanical model-based deformable registration and dose accumulation algorithm, were compared with the doses measured by the deformable gel dosimeters to verify the accuracy of MORFEUS using dose differences at each voxel as well as the gamma index test. Flexible plastic wraps were used to contain and protect the deformable gels from oxygen infiltration, which inhibits the gels' dose sensitizing ability. Since the wraps were imperfect oxygen barrier, dose comparison between MORFEUS and the deformable gels was performed only in the central region with a received dose of 200 cGy or above to exclude the peripheral region where oxygen penetration had likely affected dose measurements.
RESULTS: Dose measured with the control gels showed that the intrinsic dose measurement uncertainty of the gel dosimeters was 11.8 cGy or 4.7% compared to the TPS. The absolute mean voxel-by-voxel dose difference between the accumulated dose and the dose measured with the deformable gels was 4.7 cGy (SD = 36.0 cGy) or 1.5% (SD = 13.4%) for the three deformable gels. The absolute mean vector distance between the 250, 300, 350, and 400 cGy isodose surfaces on the accumulated and measured distributions was 1.2 mm (SD < 1.5 mm). The gamma index test that used the dose measurement precision of the control gels as the dose difference criterion and 2 mm as the distance criterion was performed, and the average pass rate of the accumulated dose distributions for all three deformable gels was 92.7%. When the distance criterion was relaxed to 3 mm, the average pass rate increased to 96.9%.
CONCLUSION: This study has proposed a novel technique to manufacture deformable volumetric gel dosimeters. By comparing the doses accumulated in MORFEUS and the doses measured with the dosimeters under the condition of deformation, the study has also demonstrated the potential of using deformable gel dosimetry to experimentally validate algorithms that include deformations into dose computation. Since dose less than 200 cGy was not evaluated in this study, future investigations will focus more on low dose regions by either using bigger gel dosimeters or prescribing a lower dose to provide a more complete experimental validation of MORFEUS across a wider dose range.

Entities:  

Mesh:

Year:  2012        PMID: 22320786      PMCID: PMC3272068          DOI: 10.1118/1.3676185

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  30 in total

1.  Potential dosimetric benefits of four-dimensional radiation treatment planning.

Authors:  George Starkschall; Keith Britton; Mary F McAleer; Melenda D Jeter; Michael R Kaus; Karl Bzdusek; Radhe Mohan; James D Cox
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-02-21       Impact factor: 7.038

2.  Clinical experience using hypofractionated radiation schedules in breast cancer.

Authors:  Timothy J Whelan; Do-Hoon Kim; Jonathan Sussman
Journal:  Semin Radiat Oncol       Date:  2008-10       Impact factor: 5.934

3.  Experimental measurements and Monte Carlo simulations for dosimetric evaluations of intrafraction motion for gated and ungated intensity modulated arc therapy deliveries.

Authors:  Mike Oliver; Adam Gladwish; Robert Staruch; Jeff Craig; Stewart Gaede; Jeff Chen; Eugene Wong
Journal:  Phys Med Biol       Date:  2008-10-21       Impact factor: 3.609

4.  Improving image-guided target localization through deformable registration.

Authors:  Kristy K Brock; Maria Hawkins; Cynthia Eccles; Joanne L Moseley; Douglas J Moseley; David A Jaffray; Laura A Dawson
Journal:  Acta Oncol       Date:  2008       Impact factor: 4.089

5.  Comparing the accuracy of four-dimensional photon dose calculations with three-dimensional calculations using moving and deforming phantoms.

Authors:  Yevgeniy Y Vinogradskiy; Peter Balter; David S Followill; Paola E Alvarez; R Allen White; George Starkschall
Journal:  Med Phys       Date:  2009-11       Impact factor: 4.071

6.  Verification of dynamic radiotherapy: the potential for 3D dosimetry under respiratory-like motion using polymer gel.

Authors:  S Ceberg; A Karlsson; H Gustavsson; L Wittgren; S A J Bäck
Journal:  Phys Med Biol       Date:  2008-09-30       Impact factor: 3.609

7.  Cumulative lung dose for several motion management strategies as a function of pretreatment patient parameters.

Authors:  Geoffrey D Hugo; Jonathon Campbell; Tiezhi Zhang; Di Yan
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-03-26       Impact factor: 7.038

8.  Monte Carlo dose mapping on deforming anatomy.

Authors:  Hualiang Zhong; Jeffrey V Siebers
Journal:  Phys Med Biol       Date:  2009-09-09       Impact factor: 3.609

9.  Contact surface and material nonlinearity modeling of human lungs.

Authors:  A Al-Mayah; J Moseley; K K Brock
Journal:  Phys Med Biol       Date:  2007-12-19       Impact factor: 3.609

10.  Introducing gel dosimetry in a clinical environment: customization of polymer gel composition and magnetic resonance imaging parameters used for 3D dose verifications in radiosurgery and intensity modulated radiotherapy.

Authors:  Remo A Crescenti; Stefan G Scheib; Uwe Schneider; Stefano Gianolini
Journal:  Med Phys       Date:  2007-04       Impact factor: 4.071

View more
  15 in total

1.  A mass-conserving 4D XCAT phantom for dose calculation and accumulation.

Authors:  Christopher L Williams; Pankaj Mishra; Joao Seco; Sara St James; Raymond H Mak; Ross I Berbeco; John H Lewis
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

2.  An externally and internally deformable, programmable lung motion phantom.

Authors:  Yam Cheung; Amit Sawant
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

3.  Automatic large quantity landmark pairs detection in 4DCT lung images.

Authors:  Yabo Fu; Xue Wu; Allan M Thomas; Harold H Li; Deshan Yang
Journal:  Med Phys       Date:  2019-08-07       Impact factor: 4.071

4.  Utility and validation of biomechanical deformable image registration in low-contrast images.

Authors:  Michael Velec; Titania Juang; Joanne L Moseley; Mark Oldham; Kristy K Brock
Journal:  Pract Radiat Oncol       Date:  2015-03-29

5.  On the need for comprehensive validation of deformable image registration, investigated with a novel 3-dimensional deformable dosimeter.

Authors:  Titania Juang; Shiva Das; John Adamovics; Ron Benning; Mark Oldham
Journal:  Int J Radiat Oncol Biol Phys       Date:  2013-07-23       Impact factor: 7.038

6.  Validation of biomechanical deformable image registration in the abdomen, thorax, and pelvis in a commercial radiotherapy treatment planning system.

Authors:  Michael Velec; Joanne L Moseley; Stina Svensson; Björn Hårdemark; David A Jaffray; Kristy K Brock
Journal:  Med Phys       Date:  2017-06-01       Impact factor: 4.071

7.  A novel deformable lung phantom with programably variable external and internal correlation.

Authors:  Maida Ranjbar; Pouya Sabouri; Carlo Repetto; Amit Sawant
Journal:  Med Phys       Date:  2019-04-22       Impact factor: 4.071

8.  3D dosimetric validation of ultrasound-guided radiotherapy with a dynamically deformable abdominal phantom.

Authors:  Charles K Matrosic; Wesley Culberson; Andrew Shepard; Sydney Jupitz; Bryan Bednarz
Journal:  Phys Med       Date:  2021-04-23       Impact factor: 2.685

9.  Spatiotemporal Free-Form Registration Method Assisted by a Minimum Spanning Tree During Discontinuous Transformations.

Authors:  Jang Pyo Bae; Siyeop Yoon; Malinda Vania; Deukhee Lee
Journal:  J Digit Imaging       Date:  2021-01-22       Impact factor: 4.056

10.  MIRSIG position paper: the use of image registration and fusion algorithms in radiotherapy.

Authors:  Nicholas Lowther; Rob Louwe; Johnson Yuen; Nicholas Hardcastle; Adam Yeo; Michael Jameson
Journal:  Phys Eng Sci Med       Date:  2022-05-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.