Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea.

Adrian Smedowski1, Beata Weglarz, Dorota Tarnawska, Kai Kaarniranta, Edward Wylegala.   

Abstract

PURPOSE: The aim of this study was to show the usefulness of three methods for measuring IOP: Goldmann applanation tonometry, rebound tonometry, and Ultra-High-Speed Scheimpflug technology.
METHODS: The examined group consisted of 96 patients (192 eyes), including 63 women and 33 men with a mean age of 59.3 ± 19.9 years. Together, 152 healthy eyes and 40 eyes with different pathologies were examined. Intraocular pressure was measured using the Goldmann applanation tonometer (GAT), the Icare Pro rebound tonometer (RT), and Ultra-High-Speed Scheimpflug technology (UHS ST; Corvis ST with pachymetry). Additionally, corneal pachymetry was conducted with a Scheimpflug camera (Pentacam) and an Ultrasound Pachymeter (A-scan Plus) as a comparison for Corvis ST pachymetry.
RESULTS: The mean IOPs were 15.6 ± 3.75 mm Hg, 15.6 ± 3.5 mm Hg, and 16.1 ± 4.0 mm Hg when measured with the GAT, the RT, and the UHS ST, respectively. The mean central corneal thickness (CCT) was 543.7 ± 52.7 μm, 547.9 ± 54.0 μm, and 556.25 ± 38.8 μm as measured with the UHS ST, the Pentacam, and the Ultrasound Pachymeter, respectively. In comparison between devices, there was a significant difference between IOP values measured with the GAT and the RT versus the UHS ST (P < 0.001), and there was no significant difference between GAT and RT (P = 0.5). No significant differences were observed in CCT measured with the UHS ST, Pentacam, and Ultrasound Pachymeter.
CONCLUSIONS: We showed that the RT Icare Pro ensures IOP measurements that are more comparable with the measurements obtained with the GAT than the measurements that are provided by UHS ST.

Entities:  

Keywords:  corneal biomechanics; intraocular pressure; tonometry

Mesh:

Year:  2014        PMID: 24425850     DOI: 10.1167/iovs.13-13172

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  25 in total

1.  Intraocular pressure measurement with Corvis ST in comparison with applanation tonometry and Tomey non-contact tonometry.

Authors:  Jan Luebke; L Bryniok; M Neuburger; J F Jordan; D Boehringer; T Reinhard; T Wecker; A Anton
Journal:  Int Ophthalmol       Date:  2019-04-09       Impact factor: 2.031

2.  Applanation optical coherence elastography: noncontact measurement of intraocular pressure, corneal biomechanical properties, and corneal geometry with a single instrument.

Authors:  Manmohan Singh; Zhaolong Han; Achuth Nair; Alexander Schill; Michael D Twa; Kirill V Larin
Journal:  J Biomed Opt       Date:  2017-02-01       Impact factor: 3.170

3.  IOP measurement in silicone oil tamponade eyes by Corvis ST tonometer, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Yang Zhang; Lin Zheng; Ailing Bian; Qi Zhou
Journal:  Int Ophthalmol       Date:  2017-04-25       Impact factor: 2.031

4.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

5.  Comparison of Diaton transpalpebral tonometer with applanation tonometry in keratoconus.

Authors:  Robert Pl Wisse; Natalie Peeters; Saskia M Imhof; Allegonda van der Lelij
Journal:  Int J Ophthalmol       Date:  2016-03-18       Impact factor: 1.779

Review 6.  A Review of Intraocular Pressure (IOP) and Axial Myopia.

Authors:  Dongyan Zhang; Liyin Wang; Le Jin; Yingying Wen; Xuhong Zhang; Liyue Zhang; Hong Zhu; Ziyu Wang; Xin Yu; Chen Xie; Jianping Tong; Ye Shen
Journal:  J Ophthalmol       Date:  2022-07-09       Impact factor: 1.974

7.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

8.  A new method for detecting the outer corneal contour in images from an ultra-fast Scheimpflug camera.

Authors:  Magdalena Jędzierowska; Robert Koprowski; Sławomir Wilczyński; Katarzyna Krysik
Journal:  Biomed Eng Online       Date:  2019-12-03       Impact factor: 2.819

9.  Short-term effects of caffeine intake on anterior chamber angle and intraocular pressure in low caffeine consumers.

Authors:  Beatríz Redondo; Jesús Vera; Rubén Molina; Raimundo Jiménez
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-12-10       Impact factor: 3.117

10.  [Pachymetry and intraocular pressure measurement by corneal visualization Scheimpflug technology (Corvis ST): A clinical comparison to the gold standard].

Authors:  J Steinberg; J Mehlan; A Frings; V Druchkiv; G Richard; T Katz; S J Linke
Journal:  Ophthalmologe       Date:  2015-09       Impact factor: 1.059

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.