Chengxin Liu, Guanzhong Gong, Chen Guo, Tonghai Liu, Jie Lu, Hong Zhao, Wei Dong, Yong Yin1.
Abstract
BACKGROUND: To investigate the feasibility of contouring the planning risk organ volume (PRV) for the heart, and to determine the probability of evaluating radiation dose to the heart using kilovoltage-cone beam computed tomography (kV-CBCT) in early-stage non-small-cell lung cancer (NSCLC) patients, who received stereotactic body radiotherapy (SBRT).Entities:
Mesh:
Year: 2013 PMID: 24369788 PMCID: PMC3909334 DOI: 10.1186/1748-717X-8-295
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Figure 1Mean heart volume of CT and CBCT images. The volume data of CBCT1 was obtained by averaging heart volumes of the first CBCT images from 17 patients. All other data were obtained similarly.
Figure 2Heart reproducibility in 40 series. The fifth layer shows the worst reproducibility, which reaches more than 0.98. The coinciding heart regions on the same anatomical section between the first series of CBCT images and other images reached up to 0.985 ± 0.020 from the first to the 40th section.
Figure 3Reproducibility of CBCT images. CBCT images of the heart from one representative patient show overlap of the anatomical region. The reproducibility of the heart between the first CBCT (red line) and other images (blue line) are shown on the sectional and coronal planes.
Figure 4External margins in different axial directions. The margin on the sectional direction is 0.91 cm and 1.29 cm on the coronal plane. Similar margins were obtained for all other patient presented.
Dose-volume indices between CBCT and CT
| V5Gy-Ra | 47.7 ± 16.4% | 49.0 ± 16.0% | 1.5 | 0.17 |
| V10Gy-R | 18.9 ± 11.7% | 17.4 ± 11.8% | 2.3 | 0.03 |
| V15Gy-R | 14.4 ± 10.2% | 12.8 ± 10.4% | 2.3 | 0.02 |
| V20Gy-R | 10.5 ± 9.0% | 9.0 ± 8.9% | 2.3 | 0.04 |
| V25Gy-R | 6.6 ± 2.8% | 4.5 ± 2.3% | 3.6 | 0.01 |
| V30Gy-R | 4.2 ± 2.2% | 2.5 ± 1.6% | 3.7 | 0.01 |
| V35Gy-R | 3.2 ± 1.8% | 1.7 ± 1.3% | 3.9 | 0.01 |
| V5Gy-Ab | 333.0 ± 78.4 cm3 | 282.0 ± 61.7 cm3 | 7.9 | 0.0001 |
| V10Gy-A | 133.0 ± 68.6 cm3 | 101.0 ± 59.0 cm3 | 6.8 | 0.0001 |
| V15Gy-A | 102.0 ± 62.6 cm3 | 74.6 ± 53.7 cm3 | 6.2 | 0.0001 |
| V20Gy-A | 74.0 ± 55.8 cm3 | 52.5 ± 46.4 cm3 | 4.8 | 0.0001 |
| V25Gy-A | 46.0 ± 16.9 cm3 | 27.0 ± 12.0 cm3 | 6.2 | 0.0001 |
| V30Gy-A | 30.0 ± 14.4 cm3 | 15.0 ± 8.7 cm3 | 5.5 | 0.0001 |
| V35Gy-A | 24.0 ± 12.0 cm3 | 10.0 ± 7.0 cm3 | 5.5 | 0.0001 |
| D50cm3c | 19.9 ± 7.7 Gy | 17.1 ± 6.3 Gy | 4.5 | 0.0001 |
| D100cm3 | 15.3 ± 6.2 Gy | 12.8 ± 5.7 Gy | 3.6 | 0.003 |
| D150cm3 | 11.8 ± 5.5 Gy | 9.2 ± 3.4 Gy | 3.6 | 0.003 |
| D200cm3 | 8.2 ± 2.0 Gy | 6.8 ± 1.5 Gy | 6.2 | 0.0001 |
| D250cm3 | 6.4 ± 1.7 Gy | 5.2 ± 2.0 Gy | 6.5 | 0.0001 |
| Dmean | 7.0 ± 2.8 Gy | 6.5 ± 2.4 Gy | 3.0 | 0.009 |
| Dmax | 45.7 ± 13.1 Gy | 45.1 ± 13.5 Gy | 2.9 | 0.009 |
a). V5Gy-R means the heart relative volume in the 5Gy region, V10Gy-R, V15Gy-R, V20Gy-R, V25Gy-R, V30Gy-R, and V35Gy-R are of the same mean;
b). V5Gy-A means the heart absolute volume in the 5Gy region, V10Gy-A, V15Gy-A, V20Gy-A, V25Gy-A, V30Gy-A, and V35Gy-A are of the same mean;
c). D50cm3 means the exposure dose in the 50 cm3, D100cm3, D150cm3, D200cm3, and D250cm3 are of the same mean.
Figure 5Mean heart DVH of CT and CBCT images. Relative dose-volume indices were calculated through the DVH and caves (red line for CT and blue line for CBCT) were obtained using MATLAB 2012a. The two caves are separated.