Nonenhanced ECG-gated quiescent-interval single shot MRA: image quality and stenosis assessment at 3 tesla compared with contrast-enhanced MRA and digital subtraction angiography.

Jan Hansmann1, John N Morelli, Henrik J Michaely, Thomas Riester, Johannes Budjan, Stefan O Schoenberg, Ulrike I Attenberger.   

Abstract

PURPOSE: To evaluate the diagnostic accuracy of a nonenhanced electrocardiograph-gated quiescent-interval single shot MR-angiography (QISS-MRA) at 3 Tesla with contrast-enhanced MRA (CE-MRA) and digital subtraction angiography (DSA) serving as reference standard.
MATERIALS AND METHODS: Following institutional review board approval, 16 consecutive patients with peripheral arterial disease underwent a combined peripheral MRA protocol consisting of a large field-of-view QISS-MRA, continuous table movement MRA, and an additional time-resolved MRA of the calves. DSA correlation was available in eight patients. Image quality and degree of stenosis was assessed. Sensitivity and specificity of QISS-MRA was evaluated with CE-MRA and DSA serving as the standards of reference and compared using the Fisher exact test.
RESULTS: With the exception of the calf station, image quality with QISS-MRA was rated statistically significantly less than that of CE-MRA (P < 0.05, P = 0.17, and P = 0.6, respectively). A greater percentage of segments were not accessible with QISS-MRA (19.5-20.1%) in comparison to CE-MRA (10.9%). Relative to DSA, sensitivity for QISS-MRA was high (100% versus 91.2% for CE-MRA, P = 0.24) in the evaluated segments; however, specificity (76.5%) was substantially less than that of CE-MRA (94.6%, P = 0.003).
CONCLUSION: Overall image quality and specificity of QISS-MRA at 3T are diminished relative to CE-MRA. However, when image quality is adequate, QISS-MRA has high sensitivity and, thus, has potential use in patients with contraindications to gadolinium.
Copyright © 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  3 Tesla; MR-Angiography; QISS; nonenhanced MR-angiography

Mesh:

Substances:

Year:  2013        PMID: 24338813     DOI: 10.1002/jmri.24324

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  13 in total

1.  Non-enhanced, ECG-gated MR angiography of the pedal vasculature: comparison with contrast-enhanced MR angiography and digital subtraction angiography in peripheral arterial occlusive disease.

Authors:  Tilman Schubert; Martin Takes; Markus Aschwanden; Markus Klarhoefer; Tanja Haas; Augustinus L Jacob; David Liu; Andreas Gutzeit; Sebastian Kos
Journal:  Eur Radiol       Date:  2015-10-29       Impact factor: 5.315

2.  Nonenhanced magnetic resonance angiography (MRA) of the calf arteries at 3 Tesla: intraindividual comparison of 3D flow-dependent subtractive MRA and 2D flow-independent non-subtractive MRA.

Authors:  Gesine Knobloch; Marie-Teres Lauff; Sebastian Hirsch; Carsten Schwenke; Bernd Hamm; Moritz Wagner
Journal:  Eur Radiol       Date:  2016-02-10       Impact factor: 5.315

Review 3.  Noncontrast MR angiography: An update.

Authors:  Robert R Edelman; Ioannis Koktzoglou
Journal:  J Magn Reson Imaging       Date:  2018-12-19       Impact factor: 4.813

4.  Free-Breathing Fast Low-Angle Shot Quiescent-Interval Slice-Selective Magnetic Resonance Angiography for Improved Detection of Vascular Stenoses in the Pelvis and Abdomen: Technical Development.

Authors:  Akos Varga-Szemes; Emily A Aherne; U Joseph Schoepf; Thomas M Todoran; Ioannis Koktzoglou; Robert R Edelman
Journal:  Invest Radiol       Date:  2019-12       Impact factor: 6.016

5.  Nonenhanced peripheral MR-angiography (MRA) at 3 Tesla: evaluation of quiescent-interval single-shot MRA in patients undergoing digital subtraction angiography.

Authors:  Moritz Wagner; Gesine Knobloch; Martin Gielen; Marie-Teres Lauff; Valentina Romano; Bernd Hamm; Thomas Kröncke
Journal:  Int J Cardiovasc Imaging       Date:  2015-02-20       Impact factor: 2.357

6.  A meta-analysis of the diagnostic performance of quiescent-interval-single-shot magnetic resonance angiography in peripheral arterial disease.

Authors:  Mansi Verma; Niraj Nirmal Pandey; Vishwajeet Singh; Priya Jagia
Journal:  Eur Radiol       Date:  2021-11-12       Impact factor: 7.034

7.  3D non-contrast-enhanced ECG-gated MR angiography of the lower extremities with dual-source radiofrequency transmission at 3.0 T: Intraindividual comparison with contrast-enhanced MR angiography in PAOD patients.

Authors:  Michael Rasper; Moritz Wildgruber; Marcus Settles; Hans-Henning Eckstein; Alexander Zimmermann; Christian Reeps; Ernst J Rummeny; Armin M Huber
Journal:  Eur Radiol       Date:  2015-11-11       Impact factor: 5.315

Review 8.  Advances in non-contrast quiescent-interval slice-selective (QISS) magnetic resonance angiography.

Authors:  R R Edelman; M Carr; I Koktzoglou
Journal:  Clin Radiol       Date:  2018-01-12       Impact factor: 2.350

9.  Noncontrast MR angiography (MRA) of infragenual arteries using flow-sensitive dephasing (FSD)-prepared steady-state free precession (SSFP) at 3.0 Tesla: Comparison with contrast-enhanced MRA.

Authors:  Nan Zhang; Zhaoyang Fan; Nan Luo; Xiaoming Bi; Yike Zhao; Jing An; Jiayi Liu; Zhong Chen; Zhanming Fan; Debiao Li
Journal:  J Magn Reson Imaging       Date:  2015-07-16       Impact factor: 4.813

10.  "Push-button" noncontrast MR angiography using balanced T1 relaxation-enhanced steady-state (bT1RESS).

Authors:  Robert R Edelman; Ioannis Koktzoglou
Journal:  Magn Reson Med       Date:  2020-09-16       Impact factor: 4.668

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.