Development and validation of a measurement-based source model for kilovoltage cone-beam CT Monte Carlo dosimetry simulations.

Kyle McMillan1, Michael McNitt-Gray, Dan Ruan.   

Abstract

PURPOSE: The purpose of this study is to adapt an equivalent source model originally developed for conventional CT Monte Carlo dose quantification to the radiation oncology context and validate its application for evaluating concomitant dose incurred by a kilovoltage (kV) cone-beam CT (CBCT) system integrated into a linear accelerator.
METHODS: In order to properly characterize beams from the integrated kV CBCT system, the authors have adapted a previously developed equivalent source model consisting of an equivalent spectrum module that takes into account intrinsic filtration and an equivalent filter module characterizing the added bowtie filtration. An equivalent spectrum was generated for an 80, 100, and 125 kVp beam with beam energy characterized by half-value layer measurements. An equivalent filter description was generated from bowtie profile measurements for both the full- and half-bowtie. Equivalent source models for each combination of equivalent spectrum and filter were incorporated into the Monte Carlo software package MCNPX. Monte Carlo simulations were then validated against in-phantom measurements for both the radiographic and CBCT mode of operation of the kV CBCT system. Radiographic and CBCT imaging dose was measured for a variety of protocols at various locations within a body (32 cm in diameter) and head (16 cm in diameter) CTDI phantom. The in-phantom radiographic and CBCT dose was simulated at all measurement locations and converted to absolute dose using normalization factors calculated from air scan measurements and corresponding simulations. The simulated results were compared with the physical measurements and their discrepancies were assessed quantitatively.
RESULTS: Strong agreement was observed between in-phantom simulations and measurements. For the radiographic protocols, simulations uniformly underestimated measurements by 0.54%-5.14% (mean difference = -3.07%, SD = 1.60%). For the CBCT protocols, simulations uniformly underestimated measurements by 1.35%-5.31% (mean difference = -3.42%, SD = 1.09%).
CONCLUSIONS: This work demonstrates the feasibility of using a measurement-based kV CBCT source model to facilitate dose calculations with Monte Carlo methods for both the radiographic and CBCT mode of operation. While this initial work validates simulations against measurements for simple geometries, future work will involve utilizing the source model to investigate kV CBCT dosimetry with more complex anthropomorphic phantoms and patient specific models.

Entities:  

Mesh:

Year:  2013        PMID: 24320440      PMCID: PMC3808443          DOI: 10.1118/1.4823795

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  21 in total

1.  Monte Carlo assessment of computed tomography dose to tissue adjacent to the scanned volume.

Authors:  J M Boone; V N Cooper; W R Nemzek; J P McGahan; J A Seibert
Journal:  Med Phys       Date:  2000-10       Impact factor: 4.071

2.  Assessment of effective dose from concomitant exposures required in verification of the target volume in radiotherapy.

Authors:  S P Waddington; A L McKenzie
Journal:  Br J Radiol       Date:  2004-07       Impact factor: 3.039

3.  Second cancer risk, concomitant exposures and IRMER(2000).

Authors:  E G A Aird
Journal:  Br J Radiol       Date:  2004-12       Impact factor: 3.039

4.  Spektr: a computational tool for x-ray spectral analysis and imaging system optimization.

Authors:  J H Siewerdsen; A M Waese; D J Moseley; S Richard; D A Jaffray
Journal:  Med Phys       Date:  2004-11       Impact factor: 4.071

5.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

6.  A comparison of radiation dose measured in CT dosimetry phantoms with calculations using EGS4 and voxel-based computational models.

Authors:  M Caon; G Bibbo; J Pattison
Journal:  Phys Med Biol       Date:  1997-01       Impact factor: 3.609

7.  Equivalent spectra as a measure of beam quality.

Authors:  J M Boone
Journal:  Med Phys       Date:  1986 Nov-Dec       Impact factor: 4.071

8.  CT doses in cylindrical phantoms.

Authors:  J V Atherton; W Huda
Journal:  Phys Med Biol       Date:  1995-05       Impact factor: 3.609

9.  Monte Carlo modeling for dose assessment in cone beam CT for oral and maxillofacial applications.

Authors:  G Zhang; N Marshall; R Bogaerts; R Jacobs; H Bosmans
Journal:  Med Phys       Date:  2013-07       Impact factor: 4.071

10.  Monte Carlo simulations in CT for the study of the surface air kerma and energy imparted to phantoms of varying size and position.

Authors:  P Avilés Lucas; D R Dance; I A Castellano; E Vañó
Journal:  Phys Med Biol       Date:  2004-04-21       Impact factor: 3.609

View more
  7 in total

1.  A new technique to characterize CT scanner bow-tie filter attenuation and applications in human cadaver dosimetry simulations.

Authors:  Xinhua Li; Jim Q Shi; Da Zhang; Sarabjeet Singh; Atul Padole; Alexi Otrakji; Mannudeep K Kalra; X George Xu; Bob Liu
Journal:  Med Phys       Date:  2015-11       Impact factor: 4.071

2.  The potential impact of ultrathin filter design on dosimetry and relative biological effectiveness in modern image-guided small animal irradiators.

Authors:  Yannick Poirier; Christopher Daniel Johnstone; Charles Kirkby
Journal:  Br J Radiol       Date:  2018-11-15       Impact factor: 3.039

3.  A measurement-based generalized source model for Monte Carlo dose simulations of CT scans.

Authors:  Xin Ming; Yuanming Feng; Ransheng Liu; Chengwen Yang; Li Zhou; Hezheng Zhai; Jun Deng
Journal:  Phys Med Biol       Date:  2017-01-12       Impact factor: 3.609

4.  Characterization and comparison of imaging contrast enhancement with PEG-functionalized gold nanoparticles in kV cone beam computed tomography and computed tomography imaging.

Authors:  Tara Gray; Nema Bassiri; Shaquan David; Devanshi Yogeshkumar Patel; Mohammad Abdul-Moqueet; Neil Kirby; Kathryn M Mayer
Journal:  Biomed Phys Eng Express       Date:  2020-05-29

5.  Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition.

Authors:  Andrew P Santoso; Kwang H Song; Yujiao Qin; Stephen J Gardner; Chang Liu; Indrin J Chetty; Benjamin Movsas; Munther Ajlouni; Ning Wen
Journal:  Radiat Oncol       Date:  2016-07-29       Impact factor: 3.481

6.  Technical Note: spektr 3.0-A computational tool for x-ray spectrum modeling and analysis.

Authors:  J Punnoose; J Xu; A Sisniega; W Zbijewski; J H Siewerdsen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

7.  Modeling a superficial radiotherapy X-ray source for relative dose calculations.

Authors:  Christopher D Johnstone; Richard LaFontaine; Yannick Poirier; Mauro Tambasco
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.