Jochen Bäuerle1, Florian Schuchardt, Laure Schroeder, Karl Egger, Matthias Weigel, Andreas Harloff.
Abstract
BACKGROUND: Quantification of the optic nerve sheath diameter (ONSD) by transbulbar sonography is a promising non-invasive technique for the detection of altered intracranial pressure. In order to establish this method as follow-up tool in diseases with intracranial hyper- or hypotension scan-rescan reproducibility and accuracy need to be systematically investigated.Entities:
Mesh:
Year: 2013 PMID: 24289136 PMCID: PMC4219451 DOI: 10.1186/1471-2377-13-187
Source DB: PubMed Journal: BMC Neurol ISSN: 1471-2377 Impact factor: 2.474
Figure 1Transbulbar sonography and MRI of the optic nerve sheath. (A) The eye bulb and the optic nerve are displayed by B-mode sonography. The optic nerve sheath diameter was measured 3 and 5 mm behind the papilla (dotted arrow) in an axial plane showing the optic nerve in its longitudinal course. Transverse (B) and sagital (C) T2-weighted TSE sequences were created to schedule coronal and orthogonal HASTE sequences of the optic nerve sheath 3 mm (D) and 5 mm (E) behind the papilla.
Bland-Altman and correlation analysis of ultrasound and MRI results
| | |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| | | | | | | | | | | | | ||
| ONSD 3 mm | 5.43 ± 0.49 | −0.07 | ±0.69 | 0.75 | 0.001 | 0.25 | ±0.51 | 0.82 | <0.001 | −0.01 | ±0.22 | 0.97 | <0.001 |
| (4.6 – 6.4) | |||||||||||||
| ONSD 5 mm | 5.53 ± 0.57 | −0.11 | ±0.66 | 0.82 | <0.001 | 0.52 | ±1.12 | 0.62 | 0.01 | −0.09 | ±0.47 | 0.90 | <0.001 |
| (4.8 – 6.8) | |||||||||||||
| | | | | | | | | | | | | ||
| ONSD 3 mm | 5.69 ± 0.77 | 0.00 | ±0.85 | 0.83 | <0.001 | −0.27 | ±0.36 | 0.98 | <0.001 | −0.01 | ±0.26 | 0.99 | <0.001 |
| (4.7 – 7.9) | |||||||||||||
| ONSD 5 mm | 5.09 ± 0.66 | −0.01 | ±0.87 | 0.76 | 0.001 | −0.14 | ±0.39 | 0.97 | <0.001 | 0.01 | ±0.25 | 0.98 | <0.001 |
| (4.2 – 7.1) | |||||||||||||
d, mean difference; σd, limits of agreement; r, correlation coefficient.
Figure 2Reproducibility and accuracy of ONSD assessment. Bland-Altman plots displaying the agreement of scan-rescan measurements of the optic nerve sheath diameter (ONSD) 3 mm behind the papilla by transorbital sonography (A) and MRI (B). Panel C demonstrates the agreement between sonographic and MRI-based ONSD quantification at the first visit. Continuous lines depict the mean of differences; dashed lines denote limits of agreement (mean ± 1.96 times of standard deviation).
Comparison of transorbital sonography and MR imaging
| | ||||
|---|---|---|---|---|
| ONSD 3 mm | −0.25 | ±1.06 | 0.72 | 0.002 |
| ONSD 5 mm | 0.43 | ±0.82 | 0.78 | <0.001 |
d, mean difference; σd, limits of agreement; r, correlation coefficient.