David F Ten Cate, Jolanda J Luime, Myrthe van der Ven, Johanna M W Hazes, Klazina Kooiman, Nico de Jong, Johannes G Bosch.
Abstract
INTRODUCTION: In many patients with rheumatoid arthritis (RA) subclinical disease activity can be detected with ultrasound (US), especially using power Doppler US (PDUS). However, PDUS may be highly dependent on the type of machine. This could create problems both in clinical trials and in daily clinical practice. To clarify how the PDUS signal differs between machines we created a microvessel flow phantom.Entities:
Mesh:
Year: 2013 PMID: 24286540 PMCID: PMC3978434 DOI: 10.1186/ar4345
Source DB: PubMed Journal: Arthritis Res Ther ISSN: 1478-6354 Impact factor: 5.156
Figure 1Different performance of the power Doppler ultrasound modality of two machines in one patient. (A) Machine B, presence of a positive power Doppler signal (arrow) within the region of synovial proliferation. (B) Machine A, this signal is absent. Arrowhead, vessel; *noise on cortical surface.
Figure 2Flow phantom with a fixated probe. BMF, blood-mimicking fluid.
Figure 3Continuous PDUS signal in a 1 mm vessel on Machine B.
Machines tested (alphabetical order) and probes used
| Machine A | Aloka Ltd., Tokyo, Japan α7 (probe UST-5411) |
| Machine B | Esaote, Maastricht, the Netherlands MyLab60 (probe LA 435) |
| Machine C | Philips Healthcare, Amsterdam, the Netherlands iU22 (probe L9-3) |
| Machine D | Ultrasonix Medical Corporation, Vancouver, Canada (probe L14-5/38) |
| Machine E | VisualSonics, Toronto, Canada Vevo2100 (probe MS200) |
Measuring the reliability of the pump
| Set flow | 11.31 | 3.142 | 0.141 | 0.070 | 0.035 |
| Measured flow | 10.68 | 2.948 | 0.276 | 0.108 | 0.049 |
Flow (ml/hour).
Lowest detected flow velocity resulting in a continuous positive power Doppler ultrasound signal
| A | 4 | 2.22 | 11.1 |
| B | 0.005 | 0.06 | 0.11 |
| C | 1 | 0.56 | 1.68 |
| D | 1 | 0.56 | Not detected |
| E | 0.5 | 0.33 | Not detected |
Flow velocity (mm/second).
Settings for detection of lowest flow velocity in the 150 μm vessel
| A | 1.3 cm/second | Level 1 | 8 MHz |
| B | 125 Hz | Level 1 | 10 MHz |
| C | 150 Hz | 15 Hz | R1a |
| Db | 200 Hz | Level 1 | 10 MHz |
| Eb | 1,000 Hz | Low | 12.5 MHz |
aActual frequency not displayed on this machine. bNo flow detected in this vessel.