| Literature DB >> 24285705 |
Patrick A Willems1, Thierry P Gosseye.
Abstract
Since the 1990s, treadmills have been equipped with multi-axis force transducers to measure the three components of the ground reaction forces during walking and running. These measurements are correctly performed if the whole treadmill (including the motor) is mounted on the transducers. In this case, the acceleration of the treadmill centre of mass relative to the reference frame of the laboratory is nil. The external forces exerted on one side of the treadmill are thus equal in magnitude and opposite in direction to the external forces exerted on the other side. However, uncertainty exists about the accuracy of these measures: due to friction between the belt and the tread-surface, due to the motor pulling the belt, some believe that it is not possible to correctly measure the horizontal components of the forces exerted by the feet on the belt. Here, we propose a simple model of an instrumented treadmill and we demonstrate (1) that the forces exerted by the subject moving on the upper part of the treadmill are accurately transmitted to the transducers placed under it and (2) that all internal forces - including friction - between the parts of the treadmill are cancelling each other.Entities:
Keywords: Force transducers; Ground reaction forces; Instrumented treadmill; Locomotion
Year: 2013 PMID: 24285705 PMCID: PMC3863428 DOI: 10.1242/bio.20136379
Source DB: PubMed Journal: Biol Open ISSN: 2046-6390 Impact factor: 2.422
Fig. 1.Schema of an instrumented treadmill.
The interrupted arrows represent the direction of the movement of the belt. For further explanations, see text.
List of symbols.
Fig. 2.Diagram of the fore–aft components of the external forces applied on an instrumented treadmill and of the internal forces between the different components of the treadmill.
For further explanations, see text.