Testing of a tri-instrumented-treadmill unit for kinetic analysis of locomotion tasks in static and dynamic loading conditions.

Gabriele Paolini1, Ugo Della Croce, Patrick O Riley, Frederic K Newton, D Casey Kerrigan.   

Abstract

In this study, we present a multi-treadmill system instrumented with three force platforms capable of measuring vertical and shear ground reaction forces and moments during both walking and running. Linearity, belts speed variations, repeatability of the measures, cross-talk, natural frequency, instrumental noise, moving part induced noise and drift were investigated. The noise due to vibrations and to moving parts was also investigated having a subject walking and running on the treadmill. The linearity test results showed a high linearity of all three treadmill force platforms, and vertical force natural frequency values of 219, 308, 307Hz, obtained for the three force platforms, were considered appropriate for the investigation of walking and running. The instrumental noise did not appear to be a significant source of error. The characteristics of the noise due to vibrations and moving parts changed when in the presence of a subject walking and running on the treadmill. For walking trials, averaging of gait cycles led to a systematic improvement of the signal to noise ratio, particularly for the medio-lateral component of the force. For running trials, even though averaging was not as beneficial as for walking trials, the greater force amplitude led to a better signal to noise ratio value. This instrumented treadmill demonstrated acceptable accuracy and signal to noise ratios for all ground reaction force components such that it can be useful for a variety of research and clinical gait analysis applications.

Mesh:

Year:  2006        PMID: 16759895     DOI: 10.1016/j.medengphy.2006.04.002

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  6 in total

1.  Dynamic assessment of center of pressure measurements from an instrumented AMTI treadmill with controlled precision.

Authors:  Emma Fortune; Jeremy Crenshaw; Vipul Lugade; Kenton R Kaufman
Journal:  Med Eng Phys       Date:  2017-02-01       Impact factor: 2.242

2.  Lower limb joint kinetics during moderately sloped running.

Authors:  Gaurav Telhan; Jason R Franz; Jay Dicharry; Robert P Wilder; Patrick O Riley; D Casey Kerrigan
Journal:  J Athl Train       Date:  2010 Jan-Feb       Impact factor: 2.860

3.  A simple method for calibrating force plates and force treadmills using an instrumented pole.

Authors:  Steven H Collins; Peter G Adamczyk; Daniel P Ferris; Arthur D Kuo
Journal:  Gait Posture       Date:  2008-08-27       Impact factor: 2.840

4.  Does an instrumented treadmill correctly measure the ground reaction forces?

Authors:  Patrick A Willems; Thierry P Gosseye
Journal:  Biol Open       Date:  2013-12-15       Impact factor: 2.422

5.  Developing a Low-Cost Force Treadmill via Dynamic Modeling.

Authors:  Chih-Yuan Hong; Lan-Yuen Guo; Rong Song; Mark L Nagurka; Jia-Li Sung; Chen-Wen Yen
Journal:  J Healthc Eng       Date:  2017-06-04       Impact factor: 2.682

6.  Design and evaluation of a new mechatronic platform for assessment and prevention of fall risks.

Authors:  Lorenzo Bassi Luciani; Vincenzo Genovese; Vito Monaco; Luca Odetti; Emanuele Cattin; Silvestro Micera
Journal:  J Neuroeng Rehabil       Date:  2012-07-28       Impact factor: 4.262

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.