The single photon emission microscope (SPEM) is an instrument developed to obtain high spatial resolution single photon emission computed tomography (SPECT) images of small structures inside the mouse brain. SPEM consists of two independent imaging devices, which combine a multipinhole collimator, a high-resolution, thallium-doped cesium iodide [CsI(Tl)] columnar scintillator, a demagnifying/intensifier tube, and an electron-multiplying charge-coupling device (CCD). Collimators have 300- and 450-µm diameter pinholes on tungsten slabs, in hexagonal arrays of 19 and 7 holes. Projection data are acquired in a photon-counting strategy, where CCD frames are stored at 50 frames per second, with a radius of rotation of 35 mm and magnification factor of one. The image reconstruction software tool is based on the maximum likelihood algorithm. Our aim was to evaluate the spatial resolution and sensitivity attainable with the seven-pinhole imaging device, together with the linearity for quantification on the tomographic images, and to test the instrument in obtaining tomographic images of different mouse organs. A spatial resolution better than 500 µm and a sensitivity of 21.6 counts·s-1·MBq-1 were reached, as well as a correlation coefficient between activity and intensity better than 0.99, when imaging 99mTc sources. Images of the thyroid, heart, lungs, and bones of mice were registered using 99mTc-labeled radiopharmaceuticals in times appropriate for routine preclinical experimentation of <1 h per projection data set. Detailed experimental protocols and images of the aforementioned organs are shown. We plan to extend the instrument's field of view to fix larger animals and to combine data from both detectors to reduce the acquisition time or applied activity.
The single photon emission microscope (SPEM) is an instrument developed to obtain high spatial resolution single photon emission computed tomography (SPECT) images of small structures inside the mouse brain. SPEM consists of two independent imaging devices, which combine a multipinhole collimator, a high-resolution, thallium-dopedcesium iodide [CsI(Tl)] columnar scintillator, a demagnifying/intensifier tube, and an electron-multiplying charge-coupling device (CCD). Collimators have 300- and 450-µm diameter pinholes on tungsten slabs, in hexagonal arrays of 19 and 7 holes. Projection data are acquired in a photon-counting strategy, where CCD frames are stored at 50 frames per second, with a radius of rotation of 35 mm and magnification factor of one. The image reconstruction software tool is based on the maximum likelihood algorithm. Our aim was to evaluate the spatial resolution and sensitivity attainable with the seven-pinhole imaging device, together with the linearity for quantification on the tomographic images, and to test the instrument in obtaining tomographic images of different mouse organs. A spatial resolution better than 500 µm and a sensitivity of 21.6 counts·s-1·MBq-1 were reached, as well as a correlation coefficient between activity and intensity better than 0.99, when imaging 99mTc sources. Images of the thyroid, heart, lungs, and bones of mice were registered using 99mTc-labeled radiopharmaceuticals in times appropriate for routine preclinical experimentation of <1 h per projection data set. Detailed experimental protocols and images of the aforementioned organs are shown. We plan to extend the instrument's field of view to fix larger animals and to combine data from both detectors to reduce the acquisition time or applied activity.
In vivo imaging techniques, such as single photon emission computed
tomography (SPECT) and positron emission tomography (PET), are becoming important tools
in the preclinical environment, as appropriate instrumentation is becoming more
accessible. Two main factors must be considered when imaging small animal organs: the
instrument's spatial resolution and its sensitivity, both of which are directly related
to the small size of the targets. Small animal organs are a factor of 300 (for rats) to
1000 (for mice) smaller than the equivalent organs of humans. Therefore, in order to
have the same image quality as in clinical investigations, preclinical instruments must
be built such that they can reach spatial resolutions better than clinical ones by
roughly a factor of 10. Sensitivity must also be improved, or, alternatively, larger
activities or acquisition times must be considered during experiments. Most of the
instruments recently developed for SPECT preclinical applications consider the use of
pinhole collimators, given the implicit magnification factor, to overcome the limitation
imposed by the detector's intrinsic spatial resolution, if the purpose is to upgrade
clinical instruments (1-5). To obtain greater sensitivity, multipinhole collimators have
been proposed, projecting individual images on independent areas of the detector, to
minimize the overlap while making efficient use of the detector's area (6). In the case of specially designed instruments
for preclinical applications, high intrinsic resolution scintillation detectors can be
used, e.g., pixellated thallium-dopedsodium iodide [NaI(Tl)] or thallium-doped cesium
iodide [CsI(Tl)] columnar scintillators. In this situation, large magnification factors
are no longer necessary, and it is even possible to consider demagnification to make
better use of the available sensitive area (7).
Sets of detector heads, each one with a single pinhole collimator, and arranged
surrounding the target, are an alternative to improve sensitivity, as a large set of
projections is acquired simultaneously, while each detector head guarantees improved
spatial resolution (8).The single photon emission microscope (SPEM), an instrument developed at the University
of Illinois with the main purpose of making images of small structures inside the mouse
brain, combines some of those technologies to reach high spatial resolution and
appropriate sensitivity for preclinical research. The objective of this study was to
evaluate the imaging capabilities of the instrument with in vivo animal
targets. We have obtained images of different mouse organs, which allow us to conclude
that, in its current configuration, the instrument can be used as the base of a core
facility for molecular imaging preclinical applications. A preliminary description of
the instrument's characteristics has been presented elsewhere (9).
Material and Methods
SPEM
The SPEM was developed by the University of Illinois at Urbana-Champaign, with the
main objective of obtaining ultra-high spatial resolution SPECT images of small
volumes inside the brains of mice (10).
Currently, this equipment is installed at the Center for Preclinical Imaging of the
CETEC/IIEPAE (Centro de Experimentação e Treinamento em Cirurgia, Instituto Israelita
de Ensino e Pesquisa Albert Einstein) in São Paulo, Brazil. The SPEM consists of two
independent imaging heads, which simultaneously acquire data from the target organ.
Each imaging head is composed of a multipinhole collimator, a high intrinsic
resolution columnar scintillator, an image intensifier and demagnifier tube, and a
high-sensitivity electron-multiplying charge-coupling device (CCD). The main
components of this instrument are shown in Figure
1. To create an experimental record, the target is located inside a
transparent plastic tube or animal holder and maintained in the vertical position to
minimize oscillation of the internal organs during data acquisition. Oxygen can be
supplied to the animal during the experiment, by means of a ventilation tube, which
can also be used to provide inhaled anesthetics.
Figure 1
Single photon emission microscope (SPEM). The main components are
labeled.
Planar projections of radiopharmaceutical distribution inside the target organ are
obtained by means of multipinhole collimators: 19 and 7 pinholes, with 300- and
450-µm diameters for imaging heads 1 and 2, respectively. The distribution, size, and
quantity of pinholes are devised to minimize overlap of the individual projections
and to maximize sensitivity and use of the detector's area. The projections are
registered using a CsI(Tl) columnar scintillator detector attached to an image
intensifier and to a high-sensitivity CCD camera. Images are recorded following a
photon-counting strategy, which consists of registering individual CCD frames at high
speed, typically between 25 and 200 frames per second. In this way, individual frames
contain just a few independent events of photon-detector interaction, superimposed on
a background pattern. After recording the image, a preprocessing stage is carried
out, when individual frames are evaluated to identify photon-detector interaction
events. These events are assessed to determine subpixel interaction coordinates and
deposited energy, which are saved in list-mode files. In a second preprocessing
stage, the list-mode files are read, and events are selected if the deposited energy
is between appropriate limits corresponding to the radiopharmaceutical and are then
organized to produce the set of projections (11,12).To produce the volumetric images, the projections are combined using a software tool
specially developed for this application, based on the maximum likelihood expectation
maximization algorithm (13,14). As input, the software tool uses projection
files and a model of the system's response to a point source. This system's response
function is obtained by simulation, considering that a small size source is located
along a large set of different positions in the instrument's field of view. The
physical characteristics of the instrument, which are considered for calculating the
response functions, are determined by acquiring a set of images of a real small size
57Co source on a short set of positions in the field of view (16
projections, 3 different radial positions, and 3 different vertical positions, for a
total of 144 images).The instrument's spatial resolution in the current configuration was determined by
using a phantom composed of a set of 13 glass capillaries positioned next to each
other. Seven of the 13 capillaries were filled with small quantities or sodium
pertechnetate, interleaved with empty capillaries, resulting in variable distances
between them. This phantom was imaged following our standard protocol (16
projections, 2 min/projection) with the seven-pinhole collimator imaging head.
Sensitivity was assessed by imaging a small size source of sodium pertechnetate (46
MBq), positioned near the center of the field of view and on the rotation axis, for a
period of 10 min. From this record, the number of registered interaction events per
unit time was determined. To verify the linearity between activity in the target and
intensity in the reconstructed volumetric image, we used a set of five small-volume
syringes containing different activities of sodium pertechnetate (from 7.75 to 59.0
MBq) in the same volume (0.2 mL). After processing of the images, we defined a
cylindrical region of interest 3 mm in diameter and 10 mm in length to determine the
mean intensity value. Those data were compared to determine the Pearson correlation
coefficient.
Animals
In order to test the imaging capabilities of the SPEM, young and adult Swiss mice,
varying from 30 to 50 g, were included in the experiments. All animals were
maintained under controlled temperature and humidity conditions (22±2°C, 50% relative
humidity), with free access to food and water. Special care was taken to avoid
unnecessary suffering to the animals. All the experimental protocols were evaluated
and approved by the Institutional Animal Care and Use Committee.
SPECT imaging protocol
In this experiment, animals were injected with appropriate radiopharmaceuticals
targeted for specific organs. Even though SPEM development was aimed at imaging brain
substructures, we assessed the instrument's imaging capabilities focusing on
different organs: thyroid, lungs, heart, and bones, labeled with sodium
pertechnetate, 99mTc-MAA, 99mTc-MIBI, and 99mTc-MDP,
respectively. In all cases, different injected activities and times between injection
of the radiopharmaceutical and the beginning of the image acquisition were tested, in
order to determine those that produce the best quality tomographic images. Other
parameters, including number of projections, time per projection, and number and
distribution of pinholes on the collimator, were maintained constant for each target
organ. The experimental parameters corresponding to the results described in this
study are shown in Table 1.
To prepare for image acquisition, animals were anesthetized by inhalation with
isofluorane and the radiopharmaceutical was injected through the lateral tail vein.
After the appropriate uptake and background washout time, the animals were
anesthetized intraperitoneally with a combination of ketamine and xylazine (125:12.5
mg/kg) and placed inside the animal holder. The animals remained in the vertical
position during the data acquisition protocol, in order to minimize oscillation of
the internal organs because of the rotation. Neither heart rate nor body temperature
was monitored during the experiment. However, all the animals survived the imaging
experiments.Tomographic reconstruction was carried out using a specially developed software tool
based on the iterative maximum likelihood expectation maximization algorithm. After
each iteration, a Gaussian smoothing filter was applied, whose width varied from five
to one pixels, one pixel for every 20 iterations, for a total of 100 iterations.
Visual inspection of the resulting images allowed us to verify that, after iteration
80, the images became very noisy. For that reason, iteration 80 was considered the
final one for all the imaging tests, unless otherwise stated.
Results
Spatial resolution, sensitivity, and quantification linearity
verification
Figure 2 (upper panels) shows the results of
the spatial resolution determination test. At the left, a slice through the
volumetric reconstruction of the phantom's emission can be seen. All the filled
capillaries are well identified, even those that are next to one another, at the
center of the phantom. Considering that those capillaries have 1.5-mm external
diameter and 1.0-mm internal diameter, the glass space between two contiguous tubes
is 0.5 mm thick, so we can conclude that sources located at distances of roughly 500
µm can be individually identified. An intensity profile along the center of the
capillaries is shown in the right panel of the same figure. Again, clear
identification of the three central sources can be verified. Additionally, by fitting
a Gauss function to the individual line profiles, we found a mean full width at half
maximum (FWHM) of 1.06 mm. Removing the effect of the capillary's internal diameter
of 1 mm, that value implies a system's spatial resolution of 0.35 mm (FWHM). For the
sensitivity, we found that 591,000 events were registered in a period of 10 min,
which results in a total of 21.6 counts·s−1·MBq−1. Finally, in
Figure 2 (lower panels), we illustrate the
results of imaging the quantification phantom. By comparing the activities in the
phantom with the mean intensities measured from the reconstructed image, we found a
Pearson correlation coefficient greater than 0.99.
Figure 2
Spatial resolution and quantification. Upper panels,
capillary tube phantom. Left: characteristic slice through the
volumetric reconstruction of the phantom's emission. Right:
intensity profile along a line intersecting the seven capillaries. It can be
seen that all the capillary's emissions are well identified. Lower
panels, quantification phantom. Left:
characteristic slice through the volumetric reconstruction of the phantom's
emission. Right: plot of the activity in the phantom versus
mean intensity in the phantom.
Imaging of mouse organs with 99mTc-labeled
radiopharmaceuticals
In Figures 3 to 6, we illustrate the results of imaging different mouse organs with
specific radiopharmaceuticals, all labeled with 99mTc. In the upper panels
of Figure 3, we show 4 of the 16 projections
obtained while imaging the mouse thyroid gland, labeled with sodium pertechnetate.
Besides the thyroid, a larger structure corresponding to the salivary glands is
visualized. In the lower panels, we show characteristic slices through the volumetric
reconstruction of the emission from the thyroid region, where it is possible to
verify that both lobes of the thyroid are clearly identified, which allows us to
define a practical value for the spatial resolution when imaging in
vivo biological targets. Considering a lobe diameter and an inter-lobe
distance of 0.8 and 1.2 mm, respectively, we can verify that in vivo
sources separated by nearly 1 mm between them can be individually identified.
Similarly, in Figure 4, we show a set of three
slices perpendicular to the long axis of the volumetric reconstruction of the
emission from the mouse heart, labeled with 99mTc-MIBI. This is a
challenging organ to image, mainly because it is beating fast, but also because of
the emission contribution of neighboring organs, which also cumulate the same
radiopharmaceutical, e.g., the liver and kidneys.
Figure 3
Thyroid images. Upper panels, sequence of 4 of the 16
projections obtained with the seven-pinhole imaging device of the mouse neck
area, labeled with sodium pertechnetate. Lower panels,
characteristic slices through the reconstructed volumetric model of the
radioactive emission from the mouse thyroid and salivary glands along the three
main axes, after 80 iterations.
Figure 6
Bone images. Upper panels, sequence of 4 of the 16
projections obtained with the seven-pinhole imaging device of the mouse head
area, labeled with 99mTc-MDP. Lower panels,
characteristic transversal and coronal slices through the reconstructed
volumetric model of the radioactive emission from the mouse skull, after 80
iterations.
Figure 4
Cardiac images. Characteristic slices perpendicular to the main axis of the
reconstructed volumetric model of the radioactive emission from the mouse
heart, labeled with 99mTc-MIBI, after 80 iterations.
Extended structures can also be imaged with SPEM, as in Figure 5, which shows, in the upper panels, 4 of the 16
projections obtained while imaging the mouse lungs, labeled with
99mTc-MAA. In this case, the individual pinhole projections almost filled
the complete field of view, which imposes a limit on the size of the target. The
lower panels show two characteristic slices through the volumetric reconstruction.
Other examples are given in Figure 6, which
again shows a set of 4 of the 16 projections obtained while imaging the mouse skull,
labeled with 99mTc-MDP, together with two representative slices through
the volumetric reconstruction.
Figure 5
Pulmonary images. Upper panels, sequence of 4 of the 16
projections obtained with the seven-pinhole imaging device of the mouse chest
area, labeled with 99mTc-MAA. Lower panels,
characteristic transversal and coronal slices through the reconstructed
volumetric model of the radioactive emission from the mouse lungs, after 60
iterations.
Discussion
In the last two decades, a great effort has been made to transfer clinical imaging
techniques to the preclinical environment. There is almost unanimous agreement that the
use of imaging techniques allows for a significant increase in the statistical value of
the experimental data obtained while the animal remains alive. This is mainly because
the animal can be used as its own control, instead of comparing experimental groups
where all the experimental variables cannot always be controlled, and can be evaluated
at different time points during the experiment (15). At the same time, the use of these techniques reduces maintenance costs
because fewer animals are needed, which also has clear ethical benefits.With this objective, two main lines of development have been pursued: upgrading already
available clinical gamma-cameras to be used during the time they are available, or
developing specific-purpose instruments, dedicated to imaging small animals. Following
the first approach, mechanical adaptations include the use of single or multiple pinhole
collimators, most often in association with a computer-controlled system for positioning
and rotation of the animal in front of the collimator (3,4, 16-18). In order to avoid movement of
the target, which can imply oscillation of the internal organs, the use of a set of
detectors mounted on a ring, completely surrounding the target, has been proposed, which
results in special-purpose instruments based on large scintillation detectors, for
example, FastSPECT II (8), nanoSPECT (19), and U-SPECT II (20). The second approach is based on the implementation of sets of
small-size, high-resolution imagers, focusing on small volume targets like mice or young
rats. Examples of this kind of instrument are FastSPECT III (21) and SPEM. In both cases, high-resolution columnar scintillators
are used as gamma-ray detectors, whose emission is amplified by image intensifiers and
detected by CCD cameras. Unlike SPEM, FastSPECT III consists of 20 single pinhole
collimator imagers, mounted on three rings surrounding the target space. As already
described, SPEM uses only two multipinhole collimator imagers, and the target must be
rotated for recording.While developing these instruments, special attention has been given to spatial
resolution. Considering the small size of the animals under scrutiny and, consequently,
of the organs and structures of interest, the ability to identify neighboring sources of
small size is crucial. Spatial resolution can be improved in several ways: 1) by
locating the target as near to the collimator as possible, which increases the
magnification, provided that the projections are not truncated as the animal is rotated;
2) by labeling the imaging pharmaceuticals with low-energy gamma emitters, e.g.,
125iodine; 3) by reducing the diameter of the pinhole collimator. However,
sensitivity is also a very important parameter. When considering the mean life of
99mTc, the most common gamma emitter in nuclear medicine, and the
biological mean life of the radiopharmaceutical in the animal, experimental time is
limited to no more than a couple of hours. Additionally, depending on the anesthetics,
another limit is imposed of not much more than roughly 1 h. And finally, if routine sets
of experiments are to be carried out, a value of around 30 to 60 min for data
acquisition, or even better, is optimal. As sensitivity is mostly limited by the
diameter of the collimator, alternatives to recover it include the use of large sets of
imagers with single pinhole collimators or the use of multipinhole collimators, taking
special care to avoid or limit the overlapping of individual projections on each
detector.In the case of SPEM, very small diameter pinholes, in combination with high intrinsic
spatial resolution detectors, are used to guarantee that small structures inside the
mouse body, roughly 350 µm from one to the other, can be individually identified.
Sensitivity is recovered by using large sets of pinholes (7 and 19), whose opening angle
limits the size of the field of view, thus minimizing the overlap of individual images.
However, acquisition times are of the order of 45 to 80 min, which is already large for
some experimental protocols. For this reason, strategies are being considered to combine
the acquired data from both detectors to reduce the acquisition time by a factor of 2,
and also to duplicate the system to include at least four detection cameras, again
reducing the acquisition time by an additional factor of 2. In any case, as illustrated
in this study, the instrument allows us to obtain images of several different mouse
organs, so it can be used in experimental protocols in many preclinical research areas
such as cardiology, nephrology, oncology, or the respiratory system. On the other hand,
with small modifications, larger animals can also be studied, such as hamsters, gerbils,
or young rats, to increase the range of experimental possibilities.
Authors: Lars R Furenlid; Donald W Wilson; Yi-Chun Chen; Hyunki Kim; Philip J Pietraski; Michael J Crawford; Harrison H Barrett Journal: IEEE Trans Nucl Sci Date: 2004-06 Impact factor: 1.679
Authors: Woutjan Branderhorst; Brendan Vastenhouw; Frans van der Have; Erwin L A Blezer; Wim K Bleeker; Freek J Beekman Journal: Eur J Nucl Med Mol Imaging Date: 2010-11-10 Impact factor: 9.236
Authors: Frans van der Have; Brendan Vastenhouw; Ruud M Ramakers; Woutjan Branderhorst; Jens O Krah; Changguo Ji; Steven G Staelens; Freek J Beekman Journal: J Nucl Med Date: 2009-03-16 Impact factor: 10.057
Authors: Max C Wu; Dong-Wei Gao; Richard E Sievers; Randall J Lee; Bruce H Hasegawa; Michael W Dae Journal: J Am Coll Cardiol Date: 2003-08-06 Impact factor: 24.094
Authors: Reza Golestani; Chao Wu; René A Tio; Clark J Zeebregts; Artiom D Petrov; Freek J Beekman; Rudi A J O Dierckx; Hendrikus H Boersma; Riemer H J A Slart Journal: Eur J Nucl Med Mol Imaging Date: 2010-01-13 Impact factor: 9.236
Authors: Philipp S Lienemann; Stéphanie Metzger; Anna-Sofia Kiveliö; Alain Blanc; Panagiota Papageorgiou; Alberto Astolfo; Bernd R Pinzer; Paolo Cinelli; Franz E Weber; Roger Schibli; Martin Béhé; Martin Ehrbar Journal: Sci Rep Date: 2015-05-19 Impact factor: 4.379