Abstract
[Purpose] Balance is an integral part of human ability. The smart balance master system (SBM) is a balance test instrument with good reliability and validity, but it is expensive. Therefore, we modified a Wii Fit balance board, which is a convenient balance assessment tool, and analyzed its reliability and validity.Entities:
Keywords: Balance; Smart balance master system; Wii Fit balance board
Year: 2013 PMID: 24259769 PMCID: PMC3820194 DOI: 10.1589/jpts.25.1251
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
The inter-rater reliability of the three balance tests for the young adults and elderly people
| Young adults (n = 20) | Elderly (n = 20) | |||
| Average value | ICC3,1 (95% CI) | Average value | ICC3,1 (95% CI) | |
| Wii Fit balance board | ||||
| Standing with eyes open (cm) | 1.87 ± 0.59 | 0.19 (−0.57–0.64) | 2.37 ± 1.44 | 0.93 (0.86–0.97) |
| Standing with eyes closed (cm) | 2.29 ± 0.77* | 0.21 (−0.70–0.66) | 9.75 ± 9.56 | 0.99 (0.97–0.99) |
| One leg standing (cm) | 6.45 ± 1.62* | 0.28 (−0.55–0.70) | 13.56 ± 9.14 | 0.99 (0.98–0.99) |
| SBM | ||||
| Standing with eyes open (cm) | 1.05 ± 0.41* | 0.70 (0.39–0.87) | 2.89 ± 2.47 | 0.99 (0.98–0.99) |
| Standing with eyes closed (cm) | 0.97 ± 0.30* | 0.78 (0.36–0.85) | 4.73 ± 2.39 | 0.99 (0.98–0.99) |
| One leg standing (cm) | 3.92 ± 1.48* | 0.97 (0.93–0.97) | 10.96 ± 2.89 | 0.99 (0.98–0.99) |
SBM, smart balance master system. * Young adults vs. elderly, p < 0.05
Correlation analysis of the Wii Fit balance board and SBM
| Wii Fit balance board | |||
| Standing with eyes open (cm) | Standing with eyes closed (cm) | One leg standing (cm) | |
| SBM | |||
| Standing with eyes open (cm) | 0.58* | 0.12 | 0.01 |
| Standing with eyes closed (cm) | 0.27 | 0.86* | 0.08 |
| One leg standing (cm) | 0.06 | 0.52a | 0.61* |