Literature DB >> 16409240

Comparison of the rebound tonometer (ICare) to the applanation tonometer (Tonopen XL) in normotensive dogs.

M Leiva1, C Naranjo, M T Peña.   

Abstract

OBJECTIVE: To compare intraocular pressure (IOP) measurements obtained by recently introduced rebound tonometer (ICare) and the well-known applanation tonometer Tonopen XL in normal canine eyes.
METHODS: In a prospective, randomized, single-center study, IOP measurements by ICare and Tonopen XL tonometers were compared in 160 nonpathologic canine eyes (80 dogs). Complete slit-lamp biomicroscopy and indirect ophthalmoscopy were performed on each dog. Rebound tonometry was performed first and immediately after topical anesthetic drops were instilled in both eyes. One minute after the application of the topical anesthetic, applanation tonometry was performed in both eyes. The intraocular pressures obtained by use of both techniques were compared by statistical analysis.
RESULTS: The mean IOP readings were 9.158 mmHg (SD 3.471 mmHg) for the ICare tonometer (x) and 11.053 mmHg (SD 3.451 mmHg) for the Tonopen XL readings (y). The mean difference in intraocular pressures (-1.905 mmHg) was within clinically acceptable limits. The correlation coefficient (r2) of the relationship within both tonometers was r2=0.7477. The corresponding linear regression between the tonometers readings was y=0.6662x+4.942.
CONCLUSIONS: Intraocular pressures obtained with the ICare rebound tonometer were concordant with the IOP readings obtained by applanation Tonopen XL, but ICare values were significantly (P<0.0001) lower. Rebound tonometry could be an appropriate tonometry method for routine clinical use after its calibration for canine eyes.

Entities:  

Mesh:

Year:  2006        PMID: 16409240     DOI: 10.1111/j.1463-5224.2005.00429.x

Source DB:  PubMed          Journal:  Vet Ophthalmol        ISSN: 1463-5216            Impact factor:   1.644


  20 in total

Review 1.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

2.  Mesenchymal stem cells improve healing of the cornea after alkali injury.

Authors:  Diamantis Almaliotis; Georgios Koliakos; Eleni Papakonstantinou; Anastasia Komnenou; Angelos Thomas; Spiros Petrakis; Ilias Nakos; Eleni Gounari; Vasileios Karampatakis
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2015-05-23       Impact factor: 3.117

3.  Validation of the Icare® TONOVET plus rebound tonometer in normal rabbit eyes.

Authors:  Shawna Gloe; Abby Rothering; Julie A Kiland; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2019-06-12       Impact factor: 3.467

4.  Intraocular pressure measurements in cattle, sheep, and goats with 2 different types of tonometers.

Authors:  Nina Peche; Johanna Corinna Eule
Journal:  Can J Vet Res       Date:  2018-07       Impact factor: 1.310

5.  Evaluation of rebound tonometry in non-human primates.

Authors:  Elizabeth J Elsmo; Julie A Kiland; Paul L Kaufman; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2011-02-16       Impact factor: 3.467

6.  Evaluation of the Perkins handheld applanation tonometer in horses and cattle.

Authors:  Silvia Franco Andrade; Daniel Silva Kupper; Luiz Fernando Rodrigues Pinho; Elizabeth Cunha Franco; Marcus Vinícius Felix Fabri Prataviera; Rodrigo Rolim Duarte; José Ricardo Cecílio Junqueira
Journal:  J Vet Sci       Date:  2011-06       Impact factor: 1.672

7.  Validation of the TonoVet® rebound tonometer in normal and glaucomatous cats.

Authors:  Gillian J McLellan; Jeremy P Kemmerling; Julie A Kiland
Journal:  Vet Ophthalmol       Date:  2012-06-06       Impact factor: 1.644

8.  Comparison of IOPen rebound tonometer with Goldmann applanation tonometer at different IOP levels.

Authors:  Fereydoun Farrahi; Farideh Sharifipour; Mohammad Malekahmadi; Bahman Cheraghian
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

Review 9.  Functional Anatomy of the Outflow Facilities.

Authors:  Stefano Pizzirani; Haiyan Gong
Journal:  Vet Clin North Am Small Anim Pract       Date:  2015-08-31       Impact factor: 2.093

10.  Intraocular pressure measurements using the TONOVET® rebound tonometer: Influence of the probe-cornea distance.

Authors:  Blanche D Rodrigues; Fabiano Montiani-Ferreira; Mariza Bortolini; André T Somma; András M Komáromy; Peterson Triches Dornbusch
Journal:  Vet Ophthalmol       Date:  2020-10-17       Impact factor: 1.644

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.