Literature DB >> 24124953

Evaluation of point-of-care blood glucose measurements in patients with diabetic ketoacidosis or hyperglycemic hyperosmolar syndrome admitted to a critical care unit.

Dawn E Corl1, Tom S Yin, Michelle E Mills, Tina L Spencer, Lucy Greenfield, Erin Beauchemin, Jessica Cochran, Louise D Suhr, Rachel E Thompson, Brent E Wisse.   

Abstract

BACKGROUND: Point-of-care (POC) blood glucose (BG) measurement is currently not recommended in the treatment of patients presenting with diabetic ketoacidosis (DKA) or hyperglycemic hyperosmolar syndrome (HHS).
METHODS: We prospectively evaluated and compared capillary and venous POC BG values with laboratory venous glucose in patients with DKA or HHS admitted to one critical care unit over 8 months.
RESULTS: Venous laboratory glucose was strongly correlated with venous (r = 0.98) and capillary (r = 0.96) POC glucose values, though POC glucose values were higher than venous laboratory values (venous POC 21 ± 3 mg/dl, capillary POC 30 ± 4 mg/dl; both p < .001). Increased plasma osmolality had no effect on glucose meter error, while acidemia (pH < 7.3) was associated with greater glucose meter error (p = .04) independent of glucose levels. Comparing hypothetical insulin infusion rates based on laboratory venous glucose to actual infusion rates based on POC glucose values showed that 33/61 insulin infusion rates would have been unchanged, while 28 out of 61 rates were on average 7% ± 2% higher. There were no instances of hypoglycemia in any of the patients.
CONCLUSIONS: Overall, both venous and capillary POC BG values were safe for the purpose of titrating insulin infusions in patients with severe hyperglycemia. Acidemia, but not hyperosmolality, increased POC BG value errors.
© 2013 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24124953      PMCID: PMC3876370          DOI: 10.1177/193229681300700516

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  16 in total

1.  A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.

Authors:  J L Parkes; S L Slatin; S Pardo; B H Ginsberg
Journal:  Diabetes Care       Date:  2000-08       Impact factor: 19.112

Review 2.  Glucose measurement: confounding issues in setting targets for inpatient management.

Authors:  Kathleen Dungan; John Chapman; Susan S Braithwaite; John Buse
Journal:  Diabetes Care       Date:  2007-02       Impact factor: 19.112

Review 3.  Factors affecting blood glucose monitoring: sources of errors in measurement.

Authors:  Barry H Ginsberg
Journal:  J Diabetes Sci Technol       Date:  2009-07-01

4.  Effects of pH on glucose measurements with handheld glucose meters and a portable glucose analyzer for point-of-care testing.

Authors:  Z Tang; X Du; R F Louie; G J Kost
Journal:  Arch Pathol Lab Med       Date:  2000-04       Impact factor: 5.534

5.  Self-monitoring of blood glucose. American Diabetes Association.

Authors: 
Journal:  Diabetes Care       Date:  1994-01       Impact factor: 19.112

6.  Evaluation of a real-time blood glucose monitor in children with diabetic ketoacidosis.

Authors:  J A Coss-Bu; L S Jefferson; S Stone-McCord; C N Ou; C Watrin; R Sachdeva; K C Copeland
Journal:  Diabetes Res Clin Pract       Date:  1999-06       Impact factor: 5.602

Review 7.  Analytical and clinical performance of blood glucose monitors.

Authors:  Suzanne Austin Boren; William L Clarke
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

8.  Accuracy of bedside capillary blood glucose measurements in critically ill patients.

Authors:  C Dana Critchell; Vincent Savarese; Amy Callahan; Christine Aboud; Serge Jabbour; Paul Marik
Journal:  Intensive Care Med       Date:  2007-09-01       Impact factor: 17.440

9.  Diabetic ketoacidosis: clinical characteristics, precipitating factors and outcomes of care.

Authors:  Leonid Barski; Roman Nevzorov; Elena Rabaev; Alan Jotkowitz; Ilana Harman-Boehm; Miri Zektser; Lior Zeller; Elena Shleyfer; Yaniv Almog
Journal:  Isr Med Assoc J       Date:  2012-05       Impact factor: 0.892

10.  Predictors of intensive care unit and hospital length of stay in diabetic ketoacidosis.

Authors:  Amado X Freire; Guillermo E Umpierrez; Bekele Afessa; Kashif A Latif; Lisa Bridges; Abbas E Kitabchi
Journal:  J Crit Care       Date:  2002-12       Impact factor: 3.425

View more
  1 in total

1.  Comparative performance assessment of point-of-care testing devices for measuring glucose and ketones at the patient bedside.

Authors:  Ferruccio Ceriotti; Ewa Kaczmarek; Elena Guerra; Fabrizio Mastrantonio; Fausto Lucarelli; Francesco Valgimigli; Andrea Mosca
Journal:  J Diabetes Sci Technol       Date:  2014-12-17
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.