Mitchell W Woodberry1, Rohit Shankar, Anne Cent, Keith R Jerome, Jane Kuypers.
Abstract
The results of the Focus Simplexa FluA/B & RSV Direct assay were compared to those of laboratory-developed reverse transcription PCR tests for 498 nasopharyngeal swabs. Concordance rates were 96.6% (476/493; κ = 0.91), 97.6% (481/493; κ = 0.47), and 99.2% (488/492; κ = 0.94) for influenza A, influenza B, and respiratory syncytial virus, respectively.Entities:
Mesh:
Year: 2013 PMID: 24048529 PMCID: PMC3889790 DOI: 10.1128/JCM.02395-13