Biomechanical effects of the implant material and implant-abutment interface in immediately loaded small-diameter implants.

Aaron Yu-Jen Wu1, Heng-Li Huang2, Jui-Ting Hsu2, Winston Chee3,4.   

Abstract

OBJECTIVE: Small-diameter implants have been available since the 1990s, but few studies have analyzed their mechanical properties. This study evaluated the effects of the implant material and the implant-abutment connection designs on the primary stability and the marginal bone strain of small-diameter implant subject to immediate loading.
MATERIALS AND METHODS: Insertion torque value (ITV), implant stability quotient (ISQ), and Periotest value (PTV) of three implant systems with four parameters (titanium, titanium alloy, internal and external hexagon connections) were measured after placing implants into artificial type 2 jaw-bone models. Specimens were tested under both vertical and oblique static loads at 190 N. Peak values of the principal bone strain were recorded and analyzed statistically by the Kruskal-Wallis test and multiple-comparisons Bonferroni test.
RESULTS: PTV and ISQ were higher for the NIOSM311 (internal-hex and Ti alloy) and FOSM311 (external-hex and pure Ti) implants, respectively, than for the NOSM311 (external-hex and Ti alloy) implant. Under vertical loading the peak value of peri-implant bone strains did not differ significantly among these three implant systems. However, the peak bone strains were at least 32 % lower for the NIOSM311 and FOSM311 implants than for the NOSM311 implant under lateral loading.
CONCLUSIONS: The implant material and the implant-abutment connection design significantly influence the peri-implant bone strain of immediately loaded small-diameter implants, but barely affect their primary stability. CLINICAL RELEVANCE: A commercially pure titanium implant with an internal connection has the potential to reduce the risk of implant failure of small-diameter implant related to biomechanical complications.

Entities:  

Mesh:

Year:  2013        PMID: 23999686     DOI: 10.1007/s00784-013-1099-8

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  41 in total

1.  Clinical uses of osteotomes.

Authors:  J Hahn
Journal:  J Oral Implantol       Date:  1999       Impact factor: 1.779

2.  A conical implant-abutment interface at the level of the marginal bone improves the distribution of stresses in the supporting bone. An axisymmetric finite element analysis.

Authors:  Stig Hansson
Journal:  Clin Oral Implants Res       Date:  2003-06       Impact factor: 5.977

3.  Use of 3 transitional implants as stabilization aid for predictable immediate loading of implants in the edentulous mandible.

Authors:  Aaron Yu-Jen Wu; Winston Chee
Journal:  Quintessence Int       Date:  2006-09       Impact factor: 1.677

4.  Load fatigue performance of four implant-abutment interface designs: effect of torque level and implant system.

Authors:  H C Quek; Keson B Tan; Jack I Nicholls
Journal:  Int J Oral Maxillofac Implants       Date:  2008 Mar-Apr       Impact factor: 2.804

5.  The impact of loads on standard diameter, small diameter and mini implants: a comparative laboratory study.

Authors:  Simon Rupert Allum; Rachel Anne Tomlinson; Rajendra Joshi
Journal:  Clin Oral Implants Res       Date:  2008-06       Impact factor: 5.977

Review 6.  Survival rate of immediately vs delayed loaded implants: analysis of the current literature.

Authors:  Georgios Romanos; Stuart Froum; Cyril Hery; Sang-Choon Cho; Dennis Tarnow
Journal:  J Oral Implantol       Date:  2010       Impact factor: 1.779

7.  Factors influencing resonance frequency analysis assessed by Osstell mentor during implant tissue integration: II. Implant surface modifications and implant diameter.

Authors:  Jie Han; Martina Lulic; Niklaus P Lang
Journal:  Clin Oral Implants Res       Date:  2010-06       Impact factor: 5.977

8.  Efficiency of immediately loaded mandibular full-arch implant restorations.

Authors:  Winston Chee; Sajid Jivraj
Journal:  Clin Implant Dent Relat Res       Date:  2003       Impact factor: 3.932

9.  Fracture resistance of the implant-abutment connection in implants with internal hex and internal conical connections under oblique compressive loading: an in vitro study.

Authors:  Abílio Ricciardi Coppedê; Edmilson Bersani; Maria da Gloria Chiarello de Mattos; Renata Cristina Silveira Rodrigues; Ivete Aparecida de Mattias Sartori; Ricardo Faria Ribeiro
Journal:  Int J Prosthodont       Date:  2009 May-Jun       Impact factor: 1.681

10.  Immediate versus one-stage restoration of small-diameter implants for a single missing maxillary lateral incisor: a 3-year randomized clinical trial.

Authors:  Marco Degidi; Diego Nardi; Adriano Piattelli
Journal:  J Periodontol       Date:  2009-09       Impact factor: 6.993

View more
  1 in total

1.  A Comparative 3D Finite Element Computational Study of Three Connections.

Authors:  Davide Farronato; Mattia Manfredini; Andrea Stevanello; Veronica Campana; Lorenzo Azzi; Marco Farronato
Journal:  Materials (Basel)       Date:  2019-09-26       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.