Literature DB >> 23982758

Biomechanical analysis of the tibial tray design in TKA: comparison between modular and offset tibial trays.

Ângela Chan1, João Gamelas, João Folgado, Paulo R Fernandes.   

Abstract

PURPOSE: The aim of this work was to develop a computational biomechanical study to compare the performance of tibial trays with different offsets for a total knee arthroplasty. The goal was to investigate whether the offset tibial tray shifts the bone stress distribution, influencing the clinical outcome.
METHODS: Three geometric models were developed for the intact tibia bone: one considering a standard tibia case and the other two reproducing tibias with a medial or a lateral offset of the metaphysis. Appropriate prosthetic components were assembled in the bone for the aforementioned cases. The finite element method was used to obtain the mechanical stress distribution for the models, and the stress shielding effect due to the prosthesis was analysed.
RESULTS: The obtained results revealed that the offset cases are subjected to higher stresses than the standard case. These values can be two times superior to the ones verified in a standard case. The stress shielding effect was confirmed along all the analysed paths, except near the stem's end in some areas.
CONCLUSION: The higher stresses registered can originate lower clinical outcomes in the offset cases. These findings can be an important beginning to understand whether better bone stress distribution could be achieved in deformity correction with associated osteotomies instead of offsetting.

Entities:  

Mesh:

Year:  2013        PMID: 23982758     DOI: 10.1007/s00167-013-2628-3

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  20 in total

1.  An improved method for the automatic mapping of computed tomography numbers onto finite element models.

Authors:  Fulvia Taddei; Alberto Pancanti; Marco Viceconti
Journal:  Med Eng Phys       Date:  2004-01       Impact factor: 2.242

2.  Contact stresses and fatigue life in a knee prosthesis: comparison between in vitro measurements and computational simulations.

Authors:  Tomaso Villa; Francesco Migliavacca; Dario Gastaldi; Maurizio Colombo; Riccardo Pietrabissa
Journal:  J Biomech       Date:  2004-01       Impact factor: 2.712

3.  Analysis of the shape of the tibial tray in total knee arthroplasty using a three dimension finite element model.

Authors:  Shinya Miyoshi; Toshiaki Takahashi; Masahiro Ohtani; Hiroshi Yamamoto; Kanji Kameyama
Journal:  Clin Biomech (Bristol, Avon)       Date:  2002-08       Impact factor: 2.063

4.  An anatomical study of tibial metaphyseal/diaphyseal mismatch during revision total knee arthroplasty.

Authors:  Roy Abraham; Arthur L Malkani; Jason Lewis; Dennis Beck
Journal:  J Arthroplasty       Date:  2007-02       Impact factor: 4.757

5.  Finite element and experimental cortex strains of the intact and implanted tibia.

Authors:  A Completo; F Fonseca; J A Simões
Journal:  J Biomech Eng       Date:  2007-10       Impact factor: 2.097

6.  ESB Clinical Biomechanics Award 2008: Complete data of total knee replacement loading for level walking and stair climbing measured in vivo with a follow-up of 6-10 months.

Authors:  Bernd Heinlein; Ines Kutzner; Friedmar Graichen; Alwina Bender; Antonius Rohlmann; Andreas M Halder; Alexander Beier; Georg Bergmann
Journal:  Clin Biomech (Bristol, Avon)       Date:  2009-03-13       Impact factor: 2.063

7.  Fixed and mobile bearing total knee arthroplasty--influence on wear generation, corresponding wear areas, knee kinematics and particle composition.

Authors:  T M Grupp; C Kaddick; J Schwiesau; A Maas; S D Stulberg
Journal:  Clin Biomech (Bristol, Avon)       Date:  2008-12-31       Impact factor: 2.063

8.  The influence of tibial component intramedullary stems and implant-cortex contact on the strain distribution of the proximal tibia following total knee arthroplasty. An in vitro study.

Authors:  R B Bourne; J B Finlay
Journal:  Clin Orthop Relat Res       Date:  1986-07       Impact factor: 4.176

9.  The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA.

Authors:  A Completo; J A Simões; F Fonseca; M Oliveira
Journal:  Knee       Date:  2008-06       Impact factor: 2.199

10.  Stress fractures after total knee arthroplasty.

Authors:  J A Rand; M B Coventry
Journal:  J Bone Joint Surg Am       Date:  1980-03       Impact factor: 5.284

View more
  2 in total

1.  The role of offset stems in revision knee arthroplasty.

Authors:  Andrea Baldini; Giovanni Balato; Vincenzo Franceschini
Journal:  Curr Rev Musculoskelet Med       Date:  2015-12

2.  Finite element analysis of the tibial bone graft in cementless total knee arthroplasty.

Authors:  Koji Totoribe; Etsuo Chosa; Go Yamako; Hiroaki Hamada; Koki Ouchi; Shutaro Yamashita; Gang Deng
Journal:  J Orthop Surg Res       Date:  2018-05-16       Impact factor: 2.359

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.