Literature DB >> 23960692

Visualization of patent ductus arteriosus using real-time three-dimensional echocardiogram: Comparative study with 2D echocardiogram and angiography.

Alaa Roushdy1, Azza El Fiky, Dina Ezz El Din.   

Abstract

PURPOSE: To determine the feasibility and accuracy of real time 3D echocardiography (RT3DE) in determining the dimensions and anatomical type of the patent ductus arteriosus (PDA).
METHODS: The study included 42 pediatric patients with a mean age of 3.6 years (ranging from 2 months to 14 years) who were referred for elective percutaneous PDA closure. All patients underwent full 2D echocardiogram as well as RT3DE with off line analysis using Q lab software within 6 h from their angiograms. The PDA was studied as regard the anatomical type, length of the duct as well as the ampulla and the pulmonary end of the PDA. Data obtained by RT3DE was compared against 2D echocardiogram and the gold standard angiography.
RESULTS: Offline analysis of the PDA was feasible in 97.6% of the cases while determination of the anatomical type using gated color flow 3D acquisitions was achieved in 78.5% of the cases. The pulmonary end of the duct was rather elliptical using 3D echocardiogram. There was significant difference between the pulmonary end measured by 3D echocardiogram and angiography (P < 0.001). There was no significant difference between either the length or the ampulla of the PDA measured by 3D echocardiogram and that measured by angiography (P value = 0.325 and 0.611, respectively). There was a good agreement between both 2D or 3D echocardiogram and angiography in determining the anatomical type of the PDA (K = 0.744 and 0.773, respectively). However 3D echocardiogram could more accurately determine type A and type E ductus compared to 2D echocardiogram.
CONCLUSION: 3D echocardiogram was more accurate than 2D echocardiogram in determining the length and the ampulla of the PDA. The morphologic assessment of the PDA using gated 3D color flow was achieved in 78.5% of the patients. Nevertheless the use of 3D echocardiogram in assessment of small vascular structures like PDA in children with rapid heart rates is still of limited clinical value.

Entities:  

Keywords:  2D, two-dimensional; 3D, three-dimensional; CHD, congenital heart disease; Congenital heart disease; MPR, multiplane reformatted mode; PDA; PDA, patent ductus arteriosus; RAO, right anterior oblique; RT3DE; RVOT, right ventricular outflow tract; Three-dimensional echocardiography

Year:  2012        PMID: 23960692      PMCID: PMC3727490          DOI: 10.1016/j.jsha.2012.02.001

Source DB:  PubMed          Journal:  J Saudi Heart Assoc        ISSN: 1016-7315


  15 in total

1.  Live three-dimensional transthoracic echocardiographic delineation of patent ductus arteriosus.

Authors:  Ashish Sinha; Navin C Nanda; Deepak Khanna; Fikret Ilgenli; Maninder Sidhu; Steven G Lloyd; Himanshu Gupta; Benigno Soto; James K Kirklin
Journal:  Echocardiography       Date:  2004-07       Impact factor: 1.724

2.  Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology.

Authors:  Roberto M Lang; Michelle Bierig; Richard B Devereux; Frank A Flachskampf; Elyse Foster; Patricia A Pellikka; Michael H Picard; Mary J Roman; James Seward; Jack S Shanewise; Scott D Solomon; Kirk T Spencer; Martin St John Sutton; William J Stewart
Journal:  J Am Soc Echocardiogr       Date:  2005-12       Impact factor: 5.251

Review 3.  Patent ductus arteriosus.

Authors:  Douglas J Schneider; John W Moore
Journal:  Circulation       Date:  2006-10-24       Impact factor: 29.690

4.  Practical guide for three-dimensional transthoracic echocardiography using a fully sampled matrix array transducer.

Authors:  Hyun Suk Yang; Ramesh C Bansal; Farouk Mookadam; Bijoy K Khandheria; A Jamil Tajik; Krishnaswamy Chandrasekaran
Journal:  J Am Soc Echocardiogr       Date:  2008-09       Impact factor: 5.251

5.  Head-to-head comparison of real-time three-dimensional transthoracic echocardiography with transthoracic and transesophageal two-dimensional contrast echocardiography for the detection of patent foramen ovale.

Authors:  Ines Monte; Stefania Grasso; Salvatore Licciardi; Luigi P Badano
Journal:  Eur J Echocardiogr       Date:  2009-11-28

Review 6.  Three-dimensional echocardiography in mitral valve disease.

Authors:  Gabriel Valocik; Otto Kamp; Cees A Visser
Journal:  Eur J Echocardiogr       Date:  2005-12

7.  The role of surgical ligation of patent ductus arteriosus in the era of the Rashkind device.

Authors:  O Galal; R Nehgme; F al-Fadley; M de Moor; F I Abbag; S H al-Oufi; E Williams; M E Fawzy; Z al-Halees
Journal:  Ann Thorac Surg       Date:  1997-02       Impact factor: 4.330

8.  Real-time 3-dimensional echocardiographic imaging of congenital heart disease using matrix-array technology: freehand real-time scanning adds instant morphologic details not well delineated by conventional 2-dimensional imaging.

Authors:  Mohamed A Seliem; Anysia Fedec; Meryl S Cohen; Stanford Ewing; Paul E Farrell; Jack Rychik; Amy H Schultz; J William Gaynor; Thomas L Spray
Journal:  J Am Soc Echocardiogr       Date:  2006-02       Impact factor: 5.251

9.  Real-time 3-dimensional echocardiography in assessing atrial and ventricular septal defects: an echocardiographic-surgical correlative study.

Authors:  Tsung O Cheng; Ming-Xing Xie; Xin-Fang Wang; Ying Wang; Qing Lu
Journal:  Am Heart J       Date:  2004-12       Impact factor: 4.749

10.  Impact of three-dimensional echocardiography in complex congenital heart defect cases: the surgical view.

Authors:  Alessia Del Pasqua; Stephen P Sanders; Andrea de Zorzi; Alessandra Toscano; Roberta Iacobelli; Carlo Pierli; Luciano Pasquini; Roberto Di Donato; Gabriele Rinelli
Journal:  Pediatr Cardiol       Date:  2008-12-16       Impact factor: 1.655

View more
  2 in total

1.  Comparative study between real time three dimensional echocardiogram and angiography in evaluation of patent ductus arteriosus, single center experience.

Authors:  Raghda Ghonimy Elsheikh; Ahmed Z Darweish; Mohamed Elsetiha; Hazem Kamel
Journal:  J Saudi Heart Assoc       Date:  2014-03-14

2.  The Association between Patent Ductus Arteriosus and Perinatal Infection in A Group of Low Birth Weight Preterm Infants.

Authors:  Edmond Pistulli; Arjan Hamiti; Sokol Buba; Alketa Hoxha; Nita Kelmendi; Gentian Vyshka
Journal:  Iran J Pediatr       Date:  2013-11-22       Impact factor: 0.364

  2 in total